Continuum® Trilogy® Allofit® Longevity® 00-8751-012-36Recalls & reported problems
Zimmer, Inc. · Class 2 · Implant
All models of Continuum Trilogy Allofit Longevity (39)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00889024149199), received between Nov 25, 2020 and May 16, 2022 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Naturally Worn” (1 report).
3 other recalls cover devices cleared under the same FDA submission (K091508, K123019, K151448); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Naturally Worn1
- Positioning Failure1
Patient problems
- Failure of Implant1
- No Known Impact Or Consequence To Patient1
- Osteolysis1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryMay 16, 2022Naturally Worn; Naturally Worn
IT WAS REPORTED THAT PATIENT DEVELOPED ELEVATED METAL IONS AND UNDERWENT A LEFT HIP REVISION APPROXIMATELY 3 YEARS POST IMPLANTATION. DURING THE PROCEDURE, ALTR, TRUNNION CORROSION AND LINER WEAR WERE FOUND. THE STEM AND CUP REMAINED WELL FIXED. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.
Patient problems: Failure of Implant; Osteolysis
MAUDE report 14410017 ↗MalfunctionNov 25, 2020Positioning Failure; Positioning Failure
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10904845 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K091508, K123019, K151448. One submission often covers many models, so check the product and lot details in each record.
- Z-0094-2015 · Oct 1, 2014 · TerminatedContinuum Acetabular System Trabecular Metal Shell with Multi Holes, Porous 68mm, Prosthesis, Hip, Semi-Constrained, Metal/Polymer
- Z-0071-2011 · Mar 19, 2010 · TerminatedContinuum Acetabular Systems Shell Inserter Adapter with Rotational Control, 00-8790-003-00, Zimmer Inc., Warsaw, IN. Used to mate the inserter with the shell implant.
- Z-0072-2011 · Mar 19, 2010 · TerminatedContinuum Acetabular Systems Shell Inserter Adapter without Rotational Control, 00-8790-003-10, Zimmer Inc., Warsaw, IN. Used to mate the inserter with the shell implant.
Other PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices
- PINNACLE 1217-25-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED735 reports
- NA 1365-52-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED683 reports
- APEX 1246-03-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED3 deaths613 reports
- PINNACLE 1217-20-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED2 deaths590 reports
- PINNACLE 1217-30-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 death484 reports
- NA 1365-53-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED404 reports
- PINNACLE ALTRX 1221-36-054Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED388 reports
- PINNACLE ALTRX 1221-32-050Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED386 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page