Skip to content

NA 00-5850-032-14Recalls & reported problems

Zimmer, Inc. · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00889024195813), received between Aug 15, 2024 and Aug 15, 2024 — classified as 1 injury.

The most frequently reported specific device problem is “Fracture” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture1

Patient problems

  • Failure of Implant1
  • Foreign Body In Patient1
  • Pain1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 15, 2024Fracture; Fracture

    IT WAS FURTHER REPORTED THAT THE PATIENT UNDERWENT AN INITIAL SURGERY. APPROXIMATELY 5 YEARS POST-OP, THE PATIENT PRESENTED WITH LEFT LATERAL HIP PAIN AND COULD FEEL THE FRACTURED IMPLANTS. THE PATIENT WAS SUBSEQUENTLY REVISED DUE TO 2 IMPLANTED BOLTS FRACTURING. THE THREADED PART OF THE BOLT SCREWS WERE UNABLE TO BE REMOVED AND LEFT IN PLACE AS THERE WERE NO PROPER INSTRUMENTS TO UNSCREW IT FROM THE FEMUR. IN ADDITION, THE SURGEON DID NOT ATTEMPT TO PLACE A NEW WASHER AND BOLT SCREW AS THE CHANCES OF DAMAGING THE BOLT SCREW THREAD WERE HIGH AND IT COULD LEAD TO LOOSENING. ATTEMPTS HAVE BEEN MADE AND ALL AVAILABLE INFORMATION HAS BEEN PROVIDED.

    Patient problems: Failure of Implant; Foreign Body In Patient; Pain

    MAUDE report 19990964 ↗

Other Prosthesis, hip, semi-constrained, metal/polymer, cemented devices

All Prosthesis, hip, semi-constrained, metal/polymer, cemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page