NexGen® Legacy® Prolong® 5962-40-10Recalls & reported problems

Zimmer, Inc. · Class 2 · Implant

All models of NexGen Legacy Prolong (26)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00889024211247), received between Jul 21, 2020 and Dec 9, 2024 — classified as 5 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Positioning Failure” (3 reports).

4 other recalls cover devices cleared under the same FDA submission (K042271, K173057); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem4
  • Positioning Failure3
  • Device Damaged Prior to Use1
  • Difficult to Insert1
  • Migration1
  • Unstable1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Pain2
  • Ambulation or Postural Difficulties1
  • Damage to Ligament(s)1
  • Joint Laxity1
  • Loss of Range of Motion1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 9, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT PATIENT UNDERWENT A LEFT KNEE REVISION APPROXIMATELY FORTY-SIX MONTHS POST-IMPLANTATION DUE TO LIMITED FLEXION. NO ADDITIONAL PATIENT CONSEQUENCES WERE REPORTED.

    Patient problems: Loss of Range of Motion

    MAUDE report 20878679 ↗
  • InjuryApr 16, 2024Unstable; Unstable

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Ambulation or Postural Difficulties

    MAUDE report 19117687 ↗
  • InjuryAug 30, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT A RIGHT KNEE REVISION APPROXIMATELY 4 YEARS POST IMPLANTATION DUE TO INSTABILITY. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: Joint Laxity

    MAUDE report 17653025 ↗
  • MalfunctionAug 4, 2022Positioning Failure; Migration; Positioning Failure; Migration

    IT WAS REPORTED THAT A PATIENT WAS REVISED DUE TO THE LOCKING PIN OF THE ARTICULAR SURFACE BACKING OUT. DURING THE PRIMARY CASE, PROPER SEATING OF THE ARTICULAR SURFACE WAS ASCERTAINED. AFTER SURGERY THE PATIENT WAS FEELING PRICKING PAIN OVER MEDIAL ASPECT OF LEFT KNEE AND A POST OPERATIVE X-RAY SHOWED THAT THE LOCKING PIN OF ARTICULAR SURFACE WAS BACKING OUT. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION HAVE BEEN MADE; HOWEVER, NO MORE INFORMATION IS AVAILABLE AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Pain

    MAUDE report 15164364 ↗
  • MalfunctionFeb 3, 2022Positioning Failure; Positioning Failure

    IT WAS REPORTED THAT THE INSERT WOULD NOT LOCK AND FIT PROPERLY WHEN INSTALLED DURING THE PROCEDURE FOLLOWING THE PROPER SURGICAL TECHNIQUE AND TOOLS. THERE WAS NO SURGICAL DELAY. NO KNOWN IMPACT OR CONSEQUENCE TO THE PATIENT. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13444117 ↗
  • MalfunctionDec 6, 2021Positioning Failure; Positioning Failure

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12933552 ↗
  • InjurySep 21, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Pain

    MAUDE report 12505314 ↗
  • InjuryJun 18, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    UPON REASSESSMENT OF THE REPORTED EVENT, IT WAS IDENTIFIED THAT THIS DEVICE DID NOT CAUSE OR CONTRIBUTE TO THE REPORTED EVENT. THIS COMPLAINT HAS BEEN REPORTED UNDER 0002648920-2021-00160.

    Patient problems: Damage to Ligament(s)

    MAUDE report 12024632 ↗
  • MalfunctionJul 21, 2020Difficult to Insert; Device Damaged Prior to Use; Difficult to Insert; Device Damaged Prior to Use

    IT WAS REPORTED THAT THE BEARING DID NOT LOCKING PROPERLY. UPON FURTHER INSPECTION, IT WAS NOTED THAT THE LOCKING MECHANISM OF THE BEARING WAS DEFORMED. THE SURGERY WAS COMPLETED WITH ANOTHER DEVICE. THERE WAS NO HARM TO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10302799 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K042271, K173057. One submission often covers many models, so check the product and lot details in each record.

  • Z-1433-2025 · Feb 27, 2025 · Open, Classified
    NexGen LPS Flex, Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis, Model Number 00-5962-042-10
  • Z-2092-2023 · May 19, 2023 · Terminated
    NexGen Precoat Stemmed Tibial Plate Size 5
  • Z-0321-2020 · Oct 10, 2019 · Terminated
    Various Knee Systems for knee joint arthroplasty, knee prosthesis, Item Nos. 00522000100 00522000200 00522000300 00522000400 00522006304 00522006504 00522006600 00522006800
  • Z-0609-2011 · Apr 16, 2010 · Terminated
    NexGen Complete Knee Solution Stemmed tibial broach impactor, Catalog # 00-5977-011-00, Zimmer Inc., Warsaw, IN.

Other PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER devices

All PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page