Skip to content

NexGen® Legacy® Prolong® 5962-40-12Recalls & reported problems

Zimmer, Inc. · Class 2 · Implant

All models of NexGen Legacy Prolong (26)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00889024211254), received between Sep 6, 2022 and Jan 28, 2026 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Difficult to Insert” (1 report).

4 other recalls cover devices cleared under the same FDA submission (K042271, K173057); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Difficult to Insert1
  • Positioning Failure1
  • Unstable1

Patient problems

  • Loss of Range of Motion1
  • No Clinical Signs, Symptoms or Conditions1
  • Pain1
  • Scar Tissue1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 28, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS PATIENT WAS REVISED APPROXIMATELY SIX YEARS POST IMPLANTATION DUE TO ARTHROLYSIS AROUND THE NATIVE PATELLA. THE SURGEON REVISED THE ARTICULAR SURFACE WHILE REMOVING SCAR TISSUE PREDOMINANTLY AROUND THE PATELLA TO IMPROVE PATELLA TRACKING AND INCREASE RANGE OF MOTION. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION HAVE BEEN MADE; HOWEVER, NO MORE IS AVAILABLE.

    Patient problems: Loss of Range of Motion; Scar Tissue

    MAUDE report 24199845 ↗
  • MalfunctionJun 6, 2023Positioning Failure; Difficult to Insert; Positioning Failure; Difficult to Insert

    IT WAS REPORTED THAT THE POLYETHYLENE IMPLANT DID NOT FIT INTO THE METALLIC TIBIAL BASE. SEVERAL ATTEMPTS WERE MADE WITH NO SUCCESS. ANOTHER ARTICULAR SURFACE WAS USED TO COMPLETE THE SURGERY. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION HAVE BEEN MADE, HOWEVER, NO MORE INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17073375 ↗
  • InjurySep 6, 2022Unstable; Unstable

    IT WAS REPORTED PATIENT UNDERWENT A REVISION PROCEDURE EIGHT YEARS POST IMPLANTATION DUE TO PAIN, INSTABILITY AND IMPLANT FRACTURE. NO MORE INFORMATION IS AVAILABLE.

    Patient problems: Pain

    MAUDE report 15364231 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K042271, K173057. One submission often covers many models, so check the product and lot details in each record.

  • Z-1433-2025 · Feb 27, 2025 · Open, Classified
    NexGen LPS Flex, Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis, Model Number 00-5962-042-10
  • Z-2092-2023 · May 19, 2023 · Terminated
    NexGen Precoat Stemmed Tibial Plate Size 5
  • Z-0321-2020 · Oct 10, 2019 · Terminated
    Various Knee Systems for knee joint arthroplasty, knee prosthesis, Item Nos. 00522000100 00522000200 00522000300 00522000400 00522006304 00522006504 00522006600 00522006800
  • Z-0609-2011 · Apr 16, 2010 · Terminated
    NexGen Complete Knee Solution Stemmed tibial broach impactor, Catalog # 00-5977-011-00, Zimmer Inc., Warsaw, IN.

Other PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER devices

All PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page