NexGen® Legacy® 5964-40-10 (REF 00-5964-040-10)Recalls & reported problems

Zimmer, Inc. · Class 2 · Implant

All models of NexGen Legacy (119)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
12
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 00889024211940), received between Apr 27, 2020 and Apr 21, 2025 — classified as 3 injuries, 9 malfunctions.

The most frequently reported specific device problem is “Difficult to Insert” (8 reports).

9 other recalls cover devices cleared under the same FDA submission (K173057, K991581); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Insert8
  • Insufficient Device Problem Information2
  • Material Deformation2
  • Positioning Failure2
  • Adverse Event Without Identified Device or Use Problem1
  • Noise, Audible1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • No Known Impact Or Consequence To Patient3
  • Insufficient Information2
  • Ambulation or Postural Difficulties1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 21, 2025Difficult to Insert; Difficult to Insert

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21872467 ↗
  • InjuryFeb 4, 2025Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information

    IT WAS REPORTED PATIENT IS EXPERIENCING INSTABILITY POST IMPLANTATION. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION HAVE BEEN MADE; HOWEVER, NO MORE IS AVAILABLE.

    Patient problems: Insufficient Information

    MAUDE report 21305488 ↗
  • InjurySep 3, 2024Noise, Audible; Noise, Audible

    IT WAS REPORTED THAT PATIENT UNDERWENT A RIGHT TOTAL KNEE ARTHROPLASTY. SUBSEQUENTLY, PATIENT REPORTS AUDIBLE NOISE AND AMBULATION DIFFICULTIES. NO REVISION PROCEDURE OR INTERVENTION HAS BEEN REPORTED.

    Patient problems: Ambulation or Postural Difficulties

    MAUDE report 20130429 ↗
  • InjuryApr 29, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT PATIENT UNDERWENT A RIGHT TOTAL KNEE ARTHROPLASTY. SUBSEQUENTLY, THE PATIENT CLAIMS THEY REQUIRE A REVISION PROCEDURE FOR AN UNKNOWN REASON. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: Insufficient Information

    MAUDE report 19202027 ↗
  • MalfunctionOct 11, 2021Difficult to Insert; Difficult to Insert

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12605717 ↗
  • MalfunctionAug 19, 2021Difficult to Insert; Difficult to Insert

    IT WAS REPORTED THAT THE LINER WOULD NOT SEAT INTO THE TIBIAL TRAY. THE DOCTOR USED A SPARE LINER TO COMPLETE THE OPERATION. THE SURGERY WAS DELAYED FOR 25 MINUTES. ATTEMPTS HAVE BEEN MADE, BUT THERE IS NO ADDITIONAL INFORMATION AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12345117 ↗
  • MalfunctionJun 15, 2021Difficult to Insert; Difficult to Insert

    IT WAS REPORTED THAT DURING A KNEE PROCEDURE, THE TIBIAL INSERT WOULD NOT SEAT IN THE TIBIAL TRAY. NO FURTHER INFORMATION IS AVAILABLE AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11999029 ↗
  • MalfunctionApr 2, 2021Difficult to Insert; Difficult to Insert

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11607920 ↗
  • MalfunctionFeb 4, 2021Difficult to Insert; Material Deformation; Difficult to Insert; Material Deformation

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 11278810 ↗
  • MalfunctionDec 3, 2020Positioning Failure; Difficult to Insert; Positioning Failure; Difficult to Insert

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient

    MAUDE report 10938049 ↗
  • MalfunctionOct 22, 2020Positioning Failure; Difficult to Insert; Positioning Failure; Difficult to Insert

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10718052 ↗
  • MalfunctionApr 27, 2020Material Deformation; Material Deformation

    IT WAS REPORTED THAT THE ARTICULAR SURFACE COULD NOT BE IMPLANTED NORMALLY DURING KNEE ARTHROPLASTY DUE TO DORSAL DEFORMATION RESULTING IN A THIRTY-FIVE (35) MINUTE DELAY. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. ATTEMPTS HAVE BEEN MADE FOR ADDITIONAL INFORMATION AND NO ADDITIONAL INFORMATION IS CURRENTLY AVAILABLE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10000148 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K173057, K991581. One submission often covers many models, so check the product and lot details in each record.

  • Z-2092-2023 · May 19, 2023 · Terminated
    NexGen Precoat Stemmed Tibial Plate Size 5
  • Z-0321-2020 · Oct 10, 2019 · Terminated
    Various Knee Systems for knee joint arthroplasty, knee prosthesis, Item Nos. 00522000100 00522000200 00522000300 00522000400 00522006304 00522006504 00522006600 00522006800
  • Z-1043-2018 · Nov 29, 2017 · Terminated
    LPS-FLEX OPTION FEMORAL C-L LPS-FLEX OPTION FEMORAL C-L LPS-FLEX OPTION FEMORAL C-R LPS-FLEX OPTION FEMORAL D-L LPS-FLEX OPTION FEMORAL D-R LPS-FLEX OPTION FEMORAL E-L LPS-FLEX OPT
  • Z-1044-2018 · Nov 29, 2017 · Terminated
    LPS-FLEX PRECOAT FEMORAL B-R LPS-FLEX PRECOAT FEMORAL C-L LPS-FLEX PRECOAT FEMORAL C-R LPS-FLEX PRECOAT FEMORAL D-L LPS-FLEX PRECOAT FEMORAL D-R LPS-FLEX PRECOAT FEMORAL E-L LPS-FL
  • Z-1045-2018 · Nov 29, 2017 · Terminated
    NEXGEN LPS-FLEX PRECOAT FEMORAL SIZE E-LT This device is indicated for patients with severe knee pain and disability
  • Z-0832-2016 · Jan 11, 2016 · Terminated
    NexGen LPS-FLEX Option Femoral NexGen LPS-FLEX Precoat Femoral prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer various sizes Product Usage
  • Z-1152-2016 · Jan 11, 2016 · Terminated
    Product 54 consists of all product under product code: JWH and same usage: Item no: 596009900 NEXGEN TAPER PLUG for use in total knee arthroplasty
  • Z-1023-2014 · Oct 11, 2013 · Terminated
    MODULAR CUP NEUTRAL LINER MODULAR CUP 10 DEGREE LIN MODULAR CUP 20 DEGREE LIN MODULAR CUP 7MM OFFSET LI
  • Z-0609-2011 · Apr 16, 2010 · Terminated
    NexGen Complete Knee Solution Stemmed tibial broach impactor, Catalog # 00-5977-011-00, Zimmer Inc., Warsaw, IN.

Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices

All PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices →