Skip to content

NexGen® Legacy® 5994-17-91Recalls & reported problems

Zimmer, Inc. · Class 2 · Implant

All models of NexGen Legacy (119)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00889024224841), received between Apr 18, 2022 and Jun 18, 2025 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Tear, Rip or Hole in Device Packaging” (5 reports).

23 other recalls cover devices cleared under the same FDA submission (K173057, K960279); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Tear, Rip or Hole in Device Packaging5
  • Delivered as Unsterile Product1
  • Packaging Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 18, 2025Delivered as Unsterile Product; Delivered as Unsterile Product

    IT WAS REPORTED THAT THE CARDBOARD PACKAGING WAS DAMAGED AND UPON OPENING, THERE WAS A HOLE IN THE FIRST BLISTER. THE SECOND BLISTER ALSO HAD TWO NOTCHES. THE IMPLANT WAS UNUSABLE BECAUSE THE PACKAGING WAS SEVERELY DAMAGED. THE PATIENT REMAINED UNDER ANESTHESIA FOR ANOTHER 2 HOURS WHILE AN IMPLANT WAS RETRIEVED FROM ANOTHER FACILITY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22284578 ↗
  • MalfunctionOct 8, 2024Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging

    IT WAS REPORTED THAT THE PACKAGING OF THE IMPLANT WAS DAMAGED WITH POSSIBLE LOOSENING OF THE PAPER/PLASTIC SEAM. THE LAST PLASTIC PACKAGE WAS PUNCTURED AND THE IMPLANT WAS DEEMED UNUSABLE. THERE WAS A 20-30 MINUTE DELAY AND A SECOND IMPLANT WAS USED TO COMPLETE THE PROCEDURE. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20400172 ↗
  • MalfunctionJun 25, 2024Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19607300 ↗
  • MalfunctionJul 14, 2023Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17322576 ↗
  • MalfunctionJul 12, 2023Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17309960 ↗
  • MalfunctionApr 18, 2022Tear, Rip or Hole in Device Packaging; Packaging Problem; Tear, Rip or Hole in Device Packaging; Packaging Problem

    IT WAS REPORTED THAT WHILE PLACING IRRIGATION IN THE INNER PACKAGE AFTER REMOVING A FEMUR IMPLANT, A LEAK WAS NOTICED. THE OUTER PACKAGE WAS THEN EXAMINED AND WAS FOUND TO HAVE A SLICE IN IT. THE SURGEON DOWNSIZED TO A SIZE F FEMUR. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION HAVE BEEN MADE; HOWEVER, NO MORE INFORMATION IS AVAILABLE AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14135094 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K173057, K960279. One submission often covers many models, so check the product and lot details in each record.

  • Z-1038-2025 · Jan 6, 2025 · Open, Classified
    NexGen LCCK Legacy Articular Surface with Locking Screw - Striped Purple/C,D - Height 14 mm, Item Number 00-5994-022-14
  • Z-1039-2025 · Jan 6, 2025 · Open, Classified
    NexGen LCCK Legacy Articular Surface with Locking Screw - Striped Yellow/E,F - Height 10 mm, Item Number 00-5994-032-10
  • Z-2092-2023 · May 19, 2023 · Terminated
    NexGen Precoat Stemmed Tibial Plate Size 5
  • Z-0321-2020 · Oct 10, 2019 · Terminated
    Various Knee Systems for knee joint arthroplasty, knee prosthesis, Item Nos. 00522000100 00522000200 00522000300 00522000400 00522006304 00522006504 00522006600 00522006800
  • Z-1037-2018 · Nov 29, 2017 · Terminated
    LCCK FEM IMPLANT SZ C-L LCCK FEM IMPLANT SZ C-R LCCK FEM IMPLANT SZ D-R LCCK FEM IMPLANT SZ E-L LCCK FEM IMPLANT SZ E-R LCCK FEM IMPLANT SZ F-L LCCK FEM IMPLANT SZ F-R This device
  • Z-1034-2018 · Nov 29, 2017 · Terminated
    LPS FEM IMPLANT SZ H-R This device is indicated for patients with severe knee pain and disability
  • Z-1036-2018 · Nov 29, 2017 · Terminated
    LPS FEM COMP SZ B-L LPS FEM COMP SZ B-R LPS FEM COMP SZ C-L LPS FEM COMP SZ C-R LPS FEM COMP SZ D-L LPS FEM COMP SZ D-R LPS FEM COMP SZ E-L LPS FEM COMP SZ E-R LPS FEM COMP SZ F-L
  • Z-1035-2018 · Nov 29, 2017 · Terminated
    LPS OPTION FEMORAL C-L LPS OPTION FEMORAL C-R LPS OPTION FEMORAL D-L LPS OPTION FEMORAL D-R LPS OPTION FEMORAL E-L LPS OPTION FEMORAL E-R LPS OPTION FEMORAL F-L LPS OPTION FEMORAL
  • Z-1040-2018 · Nov 29, 2017 · Terminated
    LPS PRECOAT FEM SIZE D-RT LPS PRECOAT FEM SIZE E-RT LPS POROUS FEM SIZE F-RT This device is indicated for patients with severe knee pain and disability
  • Z-0834-2016 · Jan 11, 2016 · Terminated
    NexGen Knee LPS Porous FEM NexGen Knee LPS Option FEM NexGen Knee Lock FEM prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer various sizes Pr
  • Z-1146-2016 · Jan 11, 2016 · Terminated
    Product 48 consists of all product under product code: JWH and same usage: Item no: 597206526 NEXGEN ALL-POLY PATELLA, for use in total knee arthroplasty
  • Z-1575-2015 · Apr 2, 2015 · Terminated
    NexGen Augment; 5mm Distal Femoral Augment;Part number 00-5990-036-10. Intended to be used in total knee arthroplasty when there is insufficient bone in the distal femur.
  • Z-1023-2014 · Oct 11, 2013 · Terminated
    MODULAR CUP NEUTRAL LINER MODULAR CUP 10 DEGREE LIN MODULAR CUP 20 DEGREE LIN MODULAR CUP 7MM OFFSET LI
  • Z-0532-2011 · Sep 15, 2010 · Terminated
    Zimmer, NexGen Complete Knee Solution LPS Femoral Component, Size E Right, sterile, REF 00-5996-015-02.
  • Z-0533-2011 · Sep 15, 2010 · Terminated
    Zimmer, NexGen Complete Knee Solution LPS Flex Gender Femoral Component, Size G Left, sterile, REF 00-5764-017-51.

…and 8 more.

Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices

All PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page