Skip to content

PERSONA® Vivacit-E® 42-5122-006-11Recalls & reported problems

Zimmer, Inc. · Class 2 · Implant

All models of PERSONA Vivacit-E (285)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00889024236615), received between Jan 11, 2021 and Mar 20, 2024 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Difficult to Insert” (1 report).

2 other recalls cover devices cleared under the same FDA submission (K121771, K172524); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 2 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Difficult to Insert1

Patient problems

  • Adhesion(s)2
  • Loss of Range of Motion2
  • Ambulation or Postural Difficulties1
  • Joint Impingement1
  • Joint effusion1
  • No Consequences Or Impact To Patient1
  • Osteolysis1
  • Pain1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 20, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THE PATIENT COMPLETED MULTIPLE PT SESSIONS WITHOUT IMPROVEMENT IN ROM AND COMPLAINED OF STIFFNESS AND DIFFICULTY AMBULATING. OPEN ARTHROLYSIS WITH REVISION OF THE FEMORAL COMPONENT AND REPLACEMENT OF THE ARTICULAR SURFACE OCCURRED NINE MONTHS POST IMPLANTATION WITHOUT ANY KNOWN COMPLICATIONS. NO FURTHER DETAILS ARE AVAILABLE AT THIS TIME.

    Patient problems: Adhesion(s); Ambulation or Postural Difficulties; Loss of Range of Motion

    MAUDE report 18940151 ↗
  • InjuryMar 12, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Adhesion(s); Joint Impingement; Joint effusion; Loss of Range of Motion; Osteolysis; Pain; Scar Tissue; Swelling/ Edema

    MAUDE report 18890616 ↗
  • MalfunctionJan 11, 2021Difficult to Insert; Difficult to Insert

    IT WAS REPORTED THAT THE TIBIAL INSERT WOULD NOT SEAT INTO THE TIBIAL TRAY DURING A LEFT TOTAL KNEE ARTHROPLASTY. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION HAVE BEEN MADE; HOWEVER, NO MORE IS AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 11152753 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K121771, K172524. One submission often covers many models, so check the product and lot details in each record.

  • Z-1917-2016 · May 11, 2016 · Terminated
    Persona Trabecular Metal Tibia; Persona Trabecular Metal Tibial Plate Instruments and Modular Brackets; Used to facilitate the implantation of joint replacement implants.
  • Z-1266-2015 · Jan 28, 2015 · Terminated
    Persona Trabecular Metal Tibial Plate / Persona TM Tibia-Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uncemented, Porous, Coated, Polymer/Metal/Polymer For use in eit

Other Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer + additive/metal/polymer + additive devices

All Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer + additive/metal/polymer + additive devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page