Skip to content

PERSONA® Vivacit-E® 42-5124-008-10Recalls & reported problems

Zimmer, Inc. · Class 2 · Implant

All models of PERSONA Vivacit-E (285)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00889024237322), received between Apr 22, 2025 and Aug 24, 2026 — classified as 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Naturally Worn” (2 reports).

2 other recalls cover devices cleared under the same FDA submission (K121771); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Naturally Worn2
  • Difficult to Insert1
  • Fracture1
  • Positioning Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Pain2
  • Failure of Implant1
  • Joint Laxity1
  • Metal Related Pathology1
  • Scar Tissue1
  • Swelling/ Edema1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 24, 2026Difficult to Insert

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50411087 ↗
  • InjuryDec 8, 2025Fracture; Naturally Worn; Fracture; Naturally Worn

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Joint Laxity; Pain

    MAUDE report 23747563 ↗
  • MalfunctionSep 6, 2025Positioning Failure; Positioning Failure

    IT WAS REPORTED THAT THE SURGEON COULD NOT LOCK DOWN THE ARTICULAR SURFACE DESPITE IT BEING IN THE RIGHT POSITION. SURGEON TRIED A FEW TIMES BUT STILL COULDN¿T HEAR THE LOCK, VISIBLY NOT LOCKED IN AS WELL. THEY REQUESTED TO OPEN A NEW ONE AND MANAGE TO LOCK IT IN. ATTEMPTS HAVE BEEN MADE, AND ALL AVAILABLE INFORMATION HAS BEEN PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22988387 ↗
  • InjuryApr 22, 2025Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL LEFT KNEE ARTHROPLASTY. SUBSEQUENTLY, THE PATIENT DEVELOPED PAIN, SWELLING, AND THERE WAS CONCERN FOR PROSTHESIS FAILURE ON RADIOGRAPH. THE PATIENT WAS REVISED WITH PLACEMENT OF COMPETITOR PRODUCT APPROXIMATELY 5 YEARS AND 3 MONTHS POST IMPLANTATION. THE INITIAL PATELLA WAS RETAINED. INTRA-OPERATIVE FINDINGS INCLUDED SCAR TISSUE, SIGNIFICANT POLY WEAR, AND METALLOSIS. NO FURTHER DETAILS HAVE BEEN PROVIDED AT THIS TIME.

    Patient problems: Failure of Implant; Metal Related Pathology; Pain; Scar Tissue; Swelling/ Edema

    MAUDE report 21882284 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K121771. One submission often covers many models, so check the product and lot details in each record.

  • Z-1917-2016 · May 11, 2016 · Terminated
    Persona Trabecular Metal Tibia; Persona Trabecular Metal Tibial Plate Instruments and Modular Brackets; Used to facilitate the implantation of joint replacement implants.
  • Z-1266-2015 · Jan 28, 2015 · Terminated
    Persona Trabecular Metal Tibial Plate / Persona TM Tibia-Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uncemented, Porous, Coated, Polymer/Metal/Polymer For use in eit