Persona® The Personalized Knee® 42-5170-003-03Recalls & reported problems
Zimmer, Inc. · Class 2
All models of Persona The Personalized Knee (77)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Zimmer Persona The Personalized Knee is a knee replacement system. Nearly all of the 135 reports describe a tibial trial instrument fracturing during knee replacement surgery or after it was returned. There are no deaths or injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 135
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 135 adverse event reports that list this exact device (DI 00889024238459), received between Feb 13, 2020 and Jun 26, 2024 — classified as 135 malfunctions.
The most frequently reported specific device problem is “Fracture” (134 reports).
39 other recalls cover devices cleared under the same FDA submission (K113369, K172524); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture134
- Detachment of Device or Device Component1
- Difficult to Insert1
Patient problems
- No Clinical Signs, Symptoms or Conditions124
- No Patient Involvement7
- No Known Impact Or Consequence To Patient4
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 25 most recent of 135 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 26, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19616172 ↗MalfunctionJun 5, 2024Fracture; Fracture
IT WAS REPORTED THAT DURING A PROCEDURE, THE TASP BROKE. NO PATIENT CONSEQUENCES WERE REPORTED AS A RESULT OF THE EVENT. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19468933 ↗MalfunctionMay 15, 2024Fracture; Fracture
IT WAS REPORTED THAT A TASP WAS RETURNED FRACTURED. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION HAVE BEEN MADE; HOWEVER, NO MORE INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19320941 ↗MalfunctionMay 7, 2024Fracture; Fracture
IT WAS REPORTED THAT THE DEVICE WAS FRACTURED. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION HAVE BEEN MADE; HOWEVER, NO MORE IS AVAILABLE AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19258415 ↗MalfunctionApr 19, 2024Fracture; Fracture
IT WAS REPORTED THAT A TIBIAL ARTICULAR SURFACE PROVISIONAL INSTRUMENT FRACTURED. THERE WAS NO PATIENT IMPACT AND NO ADVERSE EVENTS HAVE BEEN REPORTED AS A RESULT OF THE MALFUNCTION. IT WAS REPORTED THAT NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19143227 ↗MalfunctionApr 19, 2024Fracture; Fracture
IT WAS REPORTED THAT A TIBIAL ARTICULAR SURFACE PROVISIONAL INSTRUMENT FRACTURED. THERE WAS NO PATIENT IMPACT AND NO ADVERSE EVENTS HAVE BEEN REPORTED AS A RESULT OF THE MALFUNCTION. IT WAS REPORTED THAT NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19143191 ↗MalfunctionApr 4, 2024Fracture; Fracture
IT WAS REPORTED THAT FOLLOWING AN INITIAL TOTAL KNEE ARTHROPLASTY, THE TIBIAL ARTICULAR SURFACE PROVISIONAL INSTRUMENT WAS FOUND TO BE FRACTURED UPON DISASSEMBLY. THERE WAS NO PATIENT IMPACT AND NO ADVERSE EVENTS HAVE BEEN REPORTED AS A RESULT OF THE MALFUNCTION. DUE DILIGENCE IS IN PROGRESS FOR THIS EVENT; NO FURTHER INFORMATION HAS BEEN REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19041884 ↗MalfunctionApr 3, 2024Fracture; Fracture
IT WAS REPORTED THAT DURING AN INITIAL TOTAL KNEE ARTHROPLASTY, THE TIBIAL ARTICULAR SURFACE PROVISIONAL INSTRUMENT FRACTURED DURING THE TRIAL PROCESS. NO PATIENT IMPACT OR ADVERSE EVENTS HAVE BEEN REPORTED AS A RESULT OF THE MALFUNCTION. DUE DILIGENCE IS IN PROGRESS FOR THIS EVENT; TO DATE NO FURTHER INFORMATION HAS BEEN REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19038038 ↗MalfunctionApr 1, 2024Fracture; Fracture
IT WAS REPORTED THAT AN ARTICULAR SURFACE PROVISIONAL FRACTURED DURING TRIALING AS PART OF A KNEE PROCEDURE. NO ADVERSE EVENTS HAVE BEEN REPORTED AS A RESULT OF THE MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19014974 ↗MalfunctionMar 29, 2024Fracture; Fracture
IT WAS REPORTED THAT DURING CLEANING OF INSTRUMENTS AFTER SURGERY, THE INSTRUMENT BROKE. ALL AVAILABLE INFORMATION HAS BEEN PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19003175 ↗MalfunctionMar 28, 2024Fracture; Fracture
IT WAS REPORTED THAT A PROVISIONAL FRACTURED AS PART OF A KNEE PROCEDURE. NO ADVERSE EVENTS HAVE BEEN REPORTED AS A RESULT OF THE MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18993735 ↗MalfunctionMar 27, 2024Fracture; Fracture
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18987603 ↗MalfunctionMar 27, 2024Fracture; Fracture
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18986555 ↗MalfunctionMar 12, 2024Fracture; Fracture
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18887677 ↗MalfunctionMar 7, 2024Fracture; Fracture
IT WAS REPORTED A TASPS WERE RETURNED FRACTURED ON A WORN INSTRUMENT CLAIM FORM. ALL PIECES HAVE NOT BEEN RETURNED. THERE IS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18858168 ↗MalfunctionFeb 26, 2024Fracture; Fracture
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18777016 ↗MalfunctionFeb 21, 2024Fracture; Fracture
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18746797 ↗MalfunctionFeb 1, 2024Fracture; Fracture
IT WAS REPORTED THAT THE TIBIAL ARTICULAR SURFACE PROVISIONAL WAS RETURNED TO ZIMMER BIOMET IN FRACTURED CONDITION. NO ADVERSE EVENTS WERE REPORTED AS A RESULT OF THIS MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18623551 ↗MalfunctionJan 5, 2024Fracture; Fracture
IT WAS REPORTED THAT THE INSTRUMENTS WERE DISCOVERED FRACTURED DURING AN INSTRUMENT CHECK IN PROCESS. ATTEMPTS HAVE BEEN MADE AND ALL AVAILABLE INFORMATION HAS BEEN PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18455355 ↗MalfunctionDec 19, 2023Fracture; Fracture
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18354838 ↗MalfunctionDec 14, 2023Fracture; Fracture
IT WAS REPORTED A FRACTURED INSTRUMENT WAS RECEIVED VIA THE WORN INSTRUMENT RETURN PROGRAM. THERE WAS NO REPORTED PATIENT INVOLVEMENT.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; No Clinical Signs, Symptoms or Conditions
MAUDE report 18325372 ↗MalfunctionDec 4, 2023Fracture; Difficult to Insert; Fracture; Difficult to Insert
NO ADDITIONAL INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18256014 ↗MalfunctionDec 4, 2023Fracture; Fracture
IT WAS REPORTED THAT THE DEVICE BROKE ON INSERTION. THE SURGICAL TECHNIQUE WAS UTILIZED. THERE WAS NO SURGICAL DELAY. NO FOREIGN BODIES WERE RETAINED. ATTEMPTS HAVE BEEN MADE AND ALL AVAILABLE INFORMATION HAS BEEN PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18254965 ↗MalfunctionNov 21, 2023Fracture; Fracture
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18183571 ↗MalfunctionNov 20, 2023Fracture; Fracture
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18171946 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K113369, K172524. One submission often covers many models, so check the product and lot details in each record.
- Z-1805-2018 · Aug 7, 2017 · TerminatedPersona Stemmed Tibial Provisional Size H Left, Item Number 42532108301 For use during orthopaedic surgery for implantation of a prosthesis.
- Z-1808-2018 · Aug 7, 2017 · TerminatedPersona Cemented Tibial SZ PLT Size H Right, Item Number 42539908302 For use during orthopaedic surgery for implantation of a prosthesis.
- Z-1807-2018 · Aug 7, 2017 · TerminatedPersona Cemented Tibial SZ PLT Size H Left, Item Number 42539908301 For use during orthopaedic surgery for implantation of a prosthesis.
- Z-1806-2018 · Aug 7, 2017 · TerminatedPersona Stemmed Tibial Provisional Size H Right, Item Number 42532108302 For use during orthopaedic surgery for implantation of a prosthesis.
- Z-1340-2016 · Mar 17, 2016 · TerminatedPersona The Personalized Knee System Articular Surface Posterior Stabilized (PS) Left Height 14mm Sterile. For use in total knee arthroplasty. REF 42-5114-005-14
- Z-1341-2016 · Mar 17, 2016 · TerminatedPersona The Personalized Knee System Articular Surface Posterior Stabilized (PS) Left Height 18mm Sterile For use in total knee arthroplasty. REF 42-5114-008-18
- Z-1157-2016 · Jan 11, 2016 · TerminatedProduct 59 consists of all product under product code: JWH and same usage: Item no: 42540000026 PSN ALL POLY PAT PLY 26MM 42540000029 PSN ALL POLY PAT PLY 29MM 42540000032 PSN ALL
- Z-1360-2015 · Mar 10, 2015 · TerminatedBrand Name: Persona Anterior Referencing Sizer with Locking Boom The Persona Anterior Referencing Sizer is a sterilizable instrument intended for use in multiple total knee arthrop
- Z-0853-2015 · Dec 4, 2014 · TerminatedPersona Cemented Tibial Broach Inserter/Extractor. The Persona Cemented Tibial Broach Inserter/Extractor Handle is a Use sterilizable instrument intended for use In multiple TKA p
- Z-1042-2015 · Nov 25, 2014 · TerminatedZimmer PERSONA Tibial Articular Surface Inserter; Item # 42-5299-001-00. Manual surgical instrument for knee arthroplasty (orthopedic).
- Z-0198-2015 · Oct 9, 2014 · TerminatedPersona Stemmed 5 Degree Cemented Tibia Product Usage: This device is indicated for patients with severe knee pain and disability due to: Rheumatoid arthritis, osteoarthritis, t
- Z-0123-2015 · Oct 7, 2014 · TerminatedThe Persona Tibial Articular Surface Inserter is a sterilizable instrument intended for use in multiple total knee arthroplasty procedures. Per the Persona Surgical technique, the
- Z-2576-2014 · Aug 7, 2014 · TerminatedPERSONA The Personalized Knee System Constrained Posterior Stabilized Tibial Articular Surface Provisional TOP with the following: L 3-5 CD; L 6-9 CD; L 3-5 EF; L 6-9 EF; L 10-11 E
- Z-2578-2014 · Aug 7, 2014 · TerminatedPERSONA The Personalized Knee System Tibial Articular Surface Provisional Left with the following: Size CD +0 mm Thickness Bottom(TB); Size CD +6 mm Thickness Bottom; Size EF +0 m
- Z-2579-2014 · Aug 7, 2014 · TerminatedPERSONA The Personalized Knee System PS Tibial Articular Surface Provisional Left with Size 3-5 CD Top; Size 6-9 CD Top; Size 3-5 EF Top; Size 6-9 EF Top; Size 10-11 EF Top; Size 6
…and 24 more.
Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page