Skip to content

Persona® The Personalized Knee® 42-5270-005-05Recalls & reported problems

Zimmer, Inc. · Class 2

All models of Persona The Personalized Knee (77)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Zimmer Persona knee prosthesis has 234 adverse event reports, with 1 injury and no deaths. The reports mostly describe trial and instrument parts fracturing, often during setup or routine checks, and most say no patient was involved. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
234
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 234 adverse event reports that list this exact device (DI 00889024245006), received between Dec 17, 2019 and May 15, 2024 — classified as 1 injury, 233 malfunctions.

The most frequently reported specific device problem is “Fracture” (234 reports).

39 other recalls cover devices cleared under the same FDA submission (K113369, K172524); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 5 reportsJan 2024: 8 reportsFeb 2024: 5 reportsMar 2024: 9 reportsApr 2024: 4 reportsMay 2024: 5 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520269
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture234
  • Naturally Worn5
  • Mechanical Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions201
  • No Patient Involvement22
  • No Known Impact Or Consequence To Patient7
  • No Consequences Or Impact To Patient4
  • No Information3

Recent adverse event reports

The 25 most recent of 234 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 15, 2024Fracture; Fracture

    IT WAS REPORTED THAT A FRACTURED PROVISIONAL WAS RECEIVED VIA THE WORN INSTRUMENT RETURN PROGRAM. THERE WAS NO REPORTED PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19320776 ↗
  • MalfunctionMay 9, 2024Fracture; Fracture

    IT WAS REPORTED THAT A FRACTURED PROVISIONAL WAS RECEIVED FROM A HOSPITAL. NO ADVERSE EVENTS HAVE BEEN REPORTED AS A RESULT OF THE MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19280706 ↗
  • MalfunctionMay 7, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE DEVICE WAS FRACTURED. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION HAVE BEEN MADE; HOWEVER, NO MORE IS AVAILABLE AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19258412 ↗
  • MalfunctionMay 7, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE DEVICE WAS FRACTURED. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION HAVE BEEN MADE; HOWEVER, NO MORE IS AVAILABLE AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19258405 ↗
  • MalfunctionMay 2, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE INSTRUMENT WAS FRACTURED. ALL PIECES WERE RETURNED. THIS ISSUE WAS DISCOVERED DURING A ROUTINE CHECK. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19229516 ↗
  • MalfunctionApr 15, 2024Fracture; Fracture

    IT WAS REPORTED THAT WHEN THE TECHNICIAN OPENED THEL TRAY, THE PROVISIONAL WAS FRACTURED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19107328 ↗
  • MalfunctionApr 8, 2024Fracture; Fracture

    IT WAS REPORTED THAT A FRACTURED PROVISIONAL WAS RECEIVED VIA THE WORN INSTRUMENT RETURN PROGRAM. THERE WAS NO REPORTED PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19059231 ↗
  • MalfunctionApr 3, 2024Fracture; Fracture

    IT WAS REPORTED THAT WHILE INSERTING THE TRIAL ARTICULAR SURFACE, THE BASE FRACTURED OFF. ALSO, WHEN CEMENTING IN THE FEMORAL COMPONENT, THE IMPACTOR PAD FRACTURED IN HALF. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION HAVE BEEN MADE; HOWEVER, NO MORE INFORMATION IS AVAILABLE AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19035693 ↗
  • MalfunctionApr 2, 2024Fracture; Fracture

    IT WAS FURTHER REPORTED THAT THE DEVICE WAS FOUND BROKEN OUTSIDE OF SURGERY. THE REP NOTICED THE DEVICE WAS CRACKED IN THE TRAY. ATTEMPTS HAVE BEEN MADE AND ALL AVAILABLE INFORMATION HAS BEEN PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19020604 ↗
  • MalfunctionMar 26, 2024Fracture; Fracture

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18982083 ↗
  • MalfunctionMar 22, 2024Fracture; Fracture

    IT WAS REPORTED DURING KNEE ARTHROPLASTY THAT THE TIBIAL ARTICULAR SURFACE PROVISIONAL FRACTURED UPON INSERTION. NO ADVERSE EVENTS WERE REPORTED AS A RESULT OF THIS MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18959597 ↗
  • MalfunctionMar 21, 2024Fracture; Fracture

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18951741 ↗
  • MalfunctionMar 7, 2024Fracture; Fracture

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18858207 ↗
  • MalfunctionMar 7, 2024Fracture; Fracture

    IT WAS REPORTED A TASPS WERE RETURNED FRACTURED ON A WORN INSTRUMENT CLAIM FORM. ALL PIECES HAVE NOT BEEN RETURNED. THERE IS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18858202 ↗
  • MalfunctionMar 7, 2024Fracture; Fracture

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18858196 ↗
  • MalfunctionMar 5, 2024Fracture; Fracture

    NO FURTHER EVENT INFORMATION AT TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18836407 ↗
  • MalfunctionMar 5, 2024Fracture; Fracture

    IT WAS REPORTED THAT A PROVISIONAL FRACTURED DURING A KNEE PROCEDURE. NO ADVERSE EVENTS HAVE BEEN REPORTED AS A RESULT OF THE MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18836002 ↗
  • MalfunctionMar 5, 2024Fracture; Fracture

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18835735 ↗
  • MalfunctionFeb 19, 2024Fracture; Fracture

    IT WAS REPORTED THAT A PROVISIONAL FRACTURED AFTER BEING STRUCK WITH A MALLET DURING INSERTION INTO THE JOINT SPACE AS PART OF A KNEE PROCEDURE. NO ADVERSE EVENTS HAVE BEEN REPORTED AS A RESULT OF THE MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18731176 ↗
  • MalfunctionFeb 9, 2024Fracture; Fracture

    IT WAS REPORTED THE INSTRUMENT BROKE WHILE TRAILING. THERE WAS NO KNOWN IMPACT OR CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18676281 ↗
  • MalfunctionFeb 7, 2024Fracture; Naturally Worn; Fracture; Naturally Worn

    IT WAS REPORTED THAT A TASP WAS WORN TO A POINT IT WAS UNABLE TO BE USED BUT WAS LATER FOUND TO BE FRACTURED. NO DEBRIS FELL OR WAS LEFT IN THE PATIENT. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18661458 ↗
  • MalfunctionFeb 7, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE DEVICE WAS FRACTURED. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION HAVE BEEN MADE; HOWEVER, NO MORE IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18661231 ↗
  • InjuryFeb 1, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE DEVICE WAS FRACTURED. NO MORE INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18623539 ↗
  • MalfunctionJan 31, 2024Fracture; Fracture

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18614876 ↗
  • MalfunctionJan 23, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE INSTRUMENT BROKE WHEN INSERTING. THE SURGICAL TECHNIQUE WAS UTILIZED. THERE WAS NO SURGICAL DELAY. NO FOREIGN BODIES WERE RETAINED. ANOTHER DEVICE WAS NOT NEEDED TO COMPLETE THE SURGERY. ATTEMPTS HAVE BEEN MADE AND ALL AVAILABLE INFORMATION HAS BEEN PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18563580 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K113369, K172524. One submission often covers many models, so check the product and lot details in each record.

  • Z-1808-2018 · Aug 7, 2017 · Terminated
    Persona Cemented Tibial SZ PLT Size H Right, Item Number 42539908302 For use during orthopaedic surgery for implantation of a prosthesis.
  • Z-1806-2018 · Aug 7, 2017 · Terminated
    Persona Stemmed Tibial Provisional Size H Right, Item Number 42532108302 For use during orthopaedic surgery for implantation of a prosthesis.
  • Z-1805-2018 · Aug 7, 2017 · Terminated
    Persona Stemmed Tibial Provisional Size H Left, Item Number 42532108301 For use during orthopaedic surgery for implantation of a prosthesis.
  • Z-1807-2018 · Aug 7, 2017 · Terminated
    Persona Cemented Tibial SZ PLT Size H Left, Item Number 42539908301 For use during orthopaedic surgery for implantation of a prosthesis.
  • Z-1341-2016 · Mar 17, 2016 · Terminated
    Persona The Personalized Knee System Articular Surface Posterior Stabilized (PS) Left Height 18mm Sterile For use in total knee arthroplasty. REF 42-5114-008-18
  • Z-1340-2016 · Mar 17, 2016 · Terminated
    Persona The Personalized Knee System Articular Surface Posterior Stabilized (PS) Left Height 14mm Sterile. For use in total knee arthroplasty. REF 42-5114-005-14
  • Z-1157-2016 · Jan 11, 2016 · Terminated
    Product 59 consists of all product under product code: JWH and same usage: Item no: 42540000026 PSN ALL POLY PAT PLY 26MM 42540000029 PSN ALL POLY PAT PLY 29MM 42540000032 PSN ALL
  • Z-1360-2015 · Mar 10, 2015 · Terminated
    Brand Name: Persona Anterior Referencing Sizer with Locking Boom The Persona Anterior Referencing Sizer is a sterilizable instrument intended for use in multiple total knee arthrop
  • Z-0853-2015 · Dec 4, 2014 · Terminated
    Persona Cemented Tibial Broach Inserter/Extractor. The Persona Cemented Tibial Broach Inserter/Extractor Handle is a Use sterilizable instrument intended for use In multiple TKA p
  • Z-1042-2015 · Nov 25, 2014 · Terminated
    Zimmer PERSONA Tibial Articular Surface Inserter; Item # 42-5299-001-00. Manual surgical instrument for knee arthroplasty (orthopedic).
  • Z-0198-2015 · Oct 9, 2014 · Terminated
    Persona Stemmed 5 Degree Cemented Tibia Product Usage: This device is indicated for patients with severe knee pain and disability due to: Rheumatoid arthritis, osteoarthritis, t
  • Z-0123-2015 · Oct 7, 2014 · Terminated
    The Persona Tibial Articular Surface Inserter is a sterilizable instrument intended for use in multiple total knee arthroplasty procedures. Per the Persona Surgical technique, the
  • Z-2578-2014 · Aug 7, 2014 · Terminated
    PERSONA The Personalized Knee System Tibial Articular Surface Provisional Left with the following: Size CD +0 mm Thickness Bottom(TB); Size CD +6 mm Thickness Bottom; Size EF +0 m
  • Z-2575-2014 · Aug 7, 2014 · Terminated
    PERSONA The Personalized Knee System Constrained Tibial Articular Surface Provisional BOTTOM with the following: L CD +0MM; L CD +6MM; L EF +0MM;L EF +6MM; L GH +0MM; L GH +6MM;
  • Z-2577-2014 · Aug 7, 2014 · Terminated
    PERSONA The Personalized Knee System CR Tibial Articular Surface Provisional with the following: Left Size 3-9 CD Top; Left Size 3-11 EF Top; Left Size 7-12 GH Top; Left Size 9-12

…and 24 more.

Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices

All PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page