Persona® The Personalized Knee® 42-5279-003-00Recalls & reported problems
Zimmer, Inc. · Class 2
All models of Persona The Personalized Knee (77)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Zimmer Persona knee implant uses small metal shim instruments to help size and position the implant during surgery. Most of the 31 reports describe instruments coming apart or being found with missing ball bearings and beads, usually spotted during setup, sterilization or cleaning before any patient was involved. There are no deaths and no injuries, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 31
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 31 adverse event reports that list this exact device (DI 00889024245747), received between Mar 24, 2020 and May 11, 2026 — classified as 31 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (31 reports).
39 other recalls cover devices cleared under the same FDA submission (K113369, K172524); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component31
Patient problems
- No Clinical Signs, Symptoms or Conditions30
- No Patient Involvement1
Recent adverse event reports
The 25 most recent of 31 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 11, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT BEFORE THE START OF SURGERY, TWO BEADS FROM A SHIM WERE NOTED TO BE MISSING. THE ISSUE WAS IDENTIFIED DURING SETUP PRIOR TO THE PROCEDURE COMMENCING. THERE WAS NO PATIENT INVOLVEMENT. IT WAS REPORTED THAT NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25136352 ↗MalfunctionNov 11, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23523024 ↗MalfunctionNov 11, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23523005 ↗MalfunctionJun 11, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT THE INSTRUMENTS WERE RETURNED MISSING BEARINGS ON A WORN INSTRUMENT CLAIM FORM. IT WAS NOTED THEY WERE RETURNED TO THE WAREHOUSE IN THIS CONDITION AFTER WASHING. THERE WAS NO REPORTED PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22183452 ↗MalfunctionMay 22, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22071367 ↗MalfunctionJan 7, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
NO FURTHER EVENT INFORMATION AT TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21091716 ↗MalfunctionJan 7, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
NO FURTHER EVENT INFORMATION AT TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21091711 ↗MalfunctionJan 7, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT A TIBIAL ARTICULAR SURFACE PROVISIONAL SHIM INSTRUMENT WAS FOUND TO BE DISASSEMBLED AND MISSING BALL BEARINGS DURING THE STERILIZATION PROCESS ON AN UNKNOWN DATE. THERE WAS NO PATIENT INVOLVEMENT AND NO ADVERSE EVENTS HAVE BEEN REPORTED AS A RESULT OF THE MALFUNCTION. ALL AVAILABLE INFORMATION HAS BEEN REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21091698 ↗MalfunctionDec 9, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20878770 ↗MalfunctionNov 19, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED DEVICE WAS RETURNED TO THE WAREHOUSE MISSING BEARINGS. ALL PIECES HAVE NOT BEEN RETURNED. ATTEMPTS HAVE BEEN MADE AND ALL AVAILABLE INFORMATION HAS BEEN PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20725776 ↗MalfunctionMay 7, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT A TIBIAL ARTICULAR SURFACE PROVISIONAL SHIM INSTRUMENT DISASSEMBLED DURING REPROCESSING. THERE WAS NO PATIENT INVOLVEMENT AND NO ADVERSE EVENTS HAVE BEEN REPORTED AS A RESULT OF THE MALFUNCTION. IT WAS REPORTED THAT NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19263071 ↗MalfunctionApr 11, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT STERILIZATION STAFF NOTICED THAT TWO BALLS WERE MISSING FROM THE LOCKING SYSTEM OF A TASP SHIM. THERE IS NO ADDITIONAL INFORMATION AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19088704 ↗MalfunctionMar 26, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18977963 ↗MalfunctionSep 25, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17809163 ↗MalfunctionAug 29, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT THE DEVICE WAS DISCOVERED TO HAVE MISSING COMPONENTS. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION HAVE BEEN MADE; HOWEVER, NO MORE IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17644347 ↗MalfunctionAug 15, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17552756 ↗MalfunctionAug 11, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17513201 ↗MalfunctionJun 5, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT THE BALL BEARING WAS MISSING FROM AN INSTRUMENT AND ANOTHER INSTRUMENT HAD A CHIP OUT OF IT. IT WAS NOTICED AFTER CLEANING. THERE WAS NO KNOWN PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17067384 ↗MalfunctionApr 21, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT THE DEVICE WAS DISCOVERED TO HAVE MISSING COMPONENTS. NO MORE INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16786163 ↗MalfunctionMar 29, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16639516 ↗MalfunctionMar 15, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16547728 ↗MalfunctionMar 15, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT THE DEVICE WAS DISCOVERED TO HAVE MISSING COMPONENTS. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION HAVE BEEN MADE; HOWEVER, NO MORE IS AVAILABLE AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16547727 ↗MalfunctionMar 15, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16547700 ↗MalfunctionMar 15, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16547627 ↗MalfunctionMar 15, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16547611 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K113369, K172524. One submission often covers many models, so check the product and lot details in each record.
- Z-1808-2018 · Aug 7, 2017 · TerminatedPersona Cemented Tibial SZ PLT Size H Right, Item Number 42539908302 For use during orthopaedic surgery for implantation of a prosthesis.
- Z-1807-2018 · Aug 7, 2017 · TerminatedPersona Cemented Tibial SZ PLT Size H Left, Item Number 42539908301 For use during orthopaedic surgery for implantation of a prosthesis.
- Z-1805-2018 · Aug 7, 2017 · TerminatedPersona Stemmed Tibial Provisional Size H Left, Item Number 42532108301 For use during orthopaedic surgery for implantation of a prosthesis.
- Z-1806-2018 · Aug 7, 2017 · TerminatedPersona Stemmed Tibial Provisional Size H Right, Item Number 42532108302 For use during orthopaedic surgery for implantation of a prosthesis.
- Z-1341-2016 · Mar 17, 2016 · TerminatedPersona The Personalized Knee System Articular Surface Posterior Stabilized (PS) Left Height 18mm Sterile For use in total knee arthroplasty. REF 42-5114-008-18
- Z-1340-2016 · Mar 17, 2016 · TerminatedPersona The Personalized Knee System Articular Surface Posterior Stabilized (PS) Left Height 14mm Sterile. For use in total knee arthroplasty. REF 42-5114-005-14
- Z-1157-2016 · Jan 11, 2016 · TerminatedProduct 59 consists of all product under product code: JWH and same usage: Item no: 42540000026 PSN ALL POLY PAT PLY 26MM 42540000029 PSN ALL POLY PAT PLY 29MM 42540000032 PSN ALL
- Z-1360-2015 · Mar 10, 2015 · TerminatedBrand Name: Persona Anterior Referencing Sizer with Locking Boom The Persona Anterior Referencing Sizer is a sterilizable instrument intended for use in multiple total knee arthrop
- Z-0853-2015 · Dec 4, 2014 · TerminatedPersona Cemented Tibial Broach Inserter/Extractor. The Persona Cemented Tibial Broach Inserter/Extractor Handle is a Use sterilizable instrument intended for use In multiple TKA p
- Z-1042-2015 · Nov 25, 2014 · TerminatedZimmer PERSONA Tibial Articular Surface Inserter; Item # 42-5299-001-00. Manual surgical instrument for knee arthroplasty (orthopedic).
- Z-0198-2015 · Oct 9, 2014 · TerminatedPersona Stemmed 5 Degree Cemented Tibia Product Usage: This device is indicated for patients with severe knee pain and disability due to: Rheumatoid arthritis, osteoarthritis, t
- Z-0123-2015 · Oct 7, 2014 · TerminatedThe Persona Tibial Articular Surface Inserter is a sterilizable instrument intended for use in multiple total knee arthroplasty procedures. Per the Persona Surgical technique, the
- Z-2580-2014 · Aug 7, 2014 · TerminatedPERSONA" The Personalized Knee System UC Tibial Articular Surface Provisional Left Size 3-7 CD Top; Left Size 4-11 EF Top; Left Size 7-12 GH Top; Left Size 9-12 J Top; Right Size 3
- Z-2577-2014 · Aug 7, 2014 · TerminatedPERSONA The Personalized Knee System CR Tibial Articular Surface Provisional with the following: Left Size 3-9 CD Top; Left Size 3-11 EF Top; Left Size 7-12 GH Top; Left Size 9-12
- Z-2575-2014 · Aug 7, 2014 · TerminatedPERSONA The Personalized Knee System Constrained Tibial Articular Surface Provisional BOTTOM with the following: L CD +0MM; L CD +6MM; L EF +0MM;L EF +6MM; L GH +0MM; L GH +6MM;
…and 24 more.
Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page