Skip to content

PERSONA® Natural Tibia® 42-5320-061-02Recalls & reported problems

Zimmer, Inc. · Class 2 · Implant

All models of PERSONA Natural Tibia (25)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00889024246645), received between Apr 15, 2024 and Sep 23, 2026 — classified as 3 injuries.

The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (1 report).

39 other recalls cover devices cleared under the same FDA submission (K113369, K172524); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Improper or Incorrect Procedure or Method1
  • Inadequacy of Device Shape and/or Size1
  • Loosening of Implant Not Related to Bone-Ingrowth1

Patient problems

  • Failure of Implant1
  • No Clinical Signs, Symptoms or Conditions1
  • Pain1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjurySep 23, 2026Improper or Incorrect Procedure or Method

    It was reported that wrong-side implants were implanted. It was noticed after the case, and the patient had to be revised. There was a 1-2 hour delay.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50867505 ↗
  • InjuryOct 24, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A RIGHT KNEE ARTHROPLASTY REVISION TO ADDRESS TIBIAL COMPONENT LOOSENING APPROXIMATELY THREE (3) YEARS POST-OPERATIVELY. ATTEMPTS HAVE BEEN MADE; HOWEVER, NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Failure of Implant

    MAUDE report 20521830 ↗
  • InjuryApr 15, 2024Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size

    IT WAS REPORTED PATIENT UNDERWENT A REVISION PROCEDURE THREE YEARS POST IMPLANTATION DUE TO PAIN AND OVERSIZED IMPLANT. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION HAVE BEEN MADE; HOWEVER, NO MORE IS AVAILABLE AT THIS TIME.

    Patient problems: Pain

    MAUDE report 19106719 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K113369, K172524. One submission often covers many models, so check the product and lot details in each record.

  • Z-1808-2018 · Aug 7, 2017 · Terminated
    Persona Cemented Tibial SZ PLT Size H Right, Item Number 42539908302 For use during orthopaedic surgery for implantation of a prosthesis.
  • Z-1807-2018 · Aug 7, 2017 · Terminated
    Persona Cemented Tibial SZ PLT Size H Left, Item Number 42539908301 For use during orthopaedic surgery for implantation of a prosthesis.
  • Z-1805-2018 · Aug 7, 2017 · Terminated
    Persona Stemmed Tibial Provisional Size H Left, Item Number 42532108301 For use during orthopaedic surgery for implantation of a prosthesis.
  • Z-1806-2018 · Aug 7, 2017 · Terminated
    Persona Stemmed Tibial Provisional Size H Right, Item Number 42532108302 For use during orthopaedic surgery for implantation of a prosthesis.
  • Z-1341-2016 · Mar 17, 2016 · Terminated
    Persona The Personalized Knee System Articular Surface Posterior Stabilized (PS) Left Height 18mm Sterile For use in total knee arthroplasty. REF 42-5114-008-18
  • Z-1340-2016 · Mar 17, 2016 · Terminated
    Persona The Personalized Knee System Articular Surface Posterior Stabilized (PS) Left Height 14mm Sterile. For use in total knee arthroplasty. REF 42-5114-005-14
  • Z-1157-2016 · Jan 11, 2016 · Terminated
    Product 59 consists of all product under product code: JWH and same usage: Item no: 42540000026 PSN ALL POLY PAT PLY 26MM 42540000029 PSN ALL POLY PAT PLY 29MM 42540000032 PSN ALL
  • Z-1360-2015 · Mar 10, 2015 · Terminated
    Brand Name: Persona Anterior Referencing Sizer with Locking Boom The Persona Anterior Referencing Sizer is a sterilizable instrument intended for use in multiple total knee arthrop
  • Z-0853-2015 · Dec 4, 2014 · Terminated
    Persona Cemented Tibial Broach Inserter/Extractor. The Persona Cemented Tibial Broach Inserter/Extractor Handle is a Use sterilizable instrument intended for use In multiple TKA p
  • Z-1042-2015 · Nov 25, 2014 · Terminated
    Zimmer PERSONA Tibial Articular Surface Inserter; Item # 42-5299-001-00. Manual surgical instrument for knee arthroplasty (orthopedic).
  • Z-0198-2015 · Oct 9, 2014 · Terminated
    Persona Stemmed 5 Degree Cemented Tibia Product Usage: This device is indicated for patients with severe knee pain and disability due to: Rheumatoid arthritis, osteoarthritis, t
  • Z-0123-2015 · Oct 7, 2014 · Terminated
    The Persona Tibial Articular Surface Inserter is a sterilizable instrument intended for use in multiple total knee arthroplasty procedures. Per the Persona Surgical technique, the
  • Z-2576-2014 · Aug 7, 2014 · Terminated
    PERSONA The Personalized Knee System Constrained Posterior Stabilized Tibial Articular Surface Provisional TOP with the following: L 3-5 CD; L 6-9 CD; L 3-5 EF; L 6-9 EF; L 10-11 E
  • Z-2577-2014 · Aug 7, 2014 · Terminated
    PERSONA The Personalized Knee System CR Tibial Articular Surface Provisional with the following: Left Size 3-9 CD Top; Left Size 3-11 EF Top; Left Size 7-12 GH Top; Left Size 9-12
  • Z-2580-2014 · Aug 7, 2014 · Terminated
    PERSONA" The Personalized Knee System UC Tibial Articular Surface Provisional Left Size 3-7 CD Top; Left Size 4-11 EF Top; Left Size 7-12 GH Top; Left Size 9-12 J Top; Right Size 3

…and 24 more.

Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices

All PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page