Trabecular Metal® 00-4349-036-06 (UDI 00889024269071)Recalls & reported problems
Zimmer, Inc. · Class 2 · Implant
All models of Trabecular Metal (74)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00889024269071), received between Mar 6, 2020 and Apr 2, 2024 — classified as 3 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).
14 other recalls cover devices cleared under the same FDA submission (K052906, K130661); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device Dislodged or Dislocated2
- Adverse Event Without Identified Device or Use Problem1
- Difficult to Insert1
- Positioning Failure1
- Tear, Rip or Hole in Device Packaging1
Patient problems
- Joint Dislocation2
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
- No Patient Involvement1
- Pain1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjuryApr 2, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO DISLOCATION. THE BEARING WAS REMOVED AND REPLACED APPROXIMATELY FOUR YEARS POST IMPLANTATION. NO ADDITIONAL PATIENT CONSEQUENCES WERE REPORTED. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.
Patient problems: Joint Dislocation
MAUDE report 19018495 ↗MalfunctionNov 29, 2023Difficult to Insert; Difficult to Insert
IT WAS REPORTED THAT A POLY LINER WOULD MATE NOT WITH A TM STEM SPACER UPON IMPACTION. ANOTHER SPACER AND LINER WERE USED TO COMPLETE THE PROCEDURE. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18227589 ↗InjuryNov 27, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO THE NERVE PAIN CAUSED BY THE SUPERIOR SCREW BEING TOO LONG. NO ADDITIONAL PATIENT CONSEQUENCES WERE REPORTED.
Patient problems: Pain
MAUDE report 18212878 ↗InjurySep 9, 2020Device Dislodged or Dislocated; Device Dislodged or Dislocated
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Joint Dislocation
MAUDE report 10509012 ↗MalfunctionAug 3, 2020Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Patient Involvement
MAUDE report 10355822 ↗MalfunctionMar 6, 2020Positioning Failure; Positioning Failure
IT HAS BEEN REPORTED AFTER MULTIPLE UNSUCCESSFUL IMPACTIONS OF THE POLY INTO THE REVERSE STEM USING DIFFERENT IMPACTORS, IT WAS DECIDED TO FINISH THE CASE USING A DIFFERENT IMPLANT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9800040 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K052906, K130661. One submission often covers many models, so check the product and lot details in each record.
- Z-0299-2020 · Oct 10, 2019 · TerminatedTrabecular Metal Reverse Shoulder System, Item Nos. 00434210613 00430900103 00434210613 00434210813 00434210817 00434210913 00434211013 00434211017 00434211113 00434211213 00434211
- Z-3144-2017 · May 22, 2017 · TerminatedZimmer Trabecular Metal Reverse Shoulder System Instrumentation, Drill 2.5 mm Diameter, Item Number/EDI 47430904601, Sterile
- Z-0609-2014 · Dec 10, 2013 · TerminatedTrabecular Metal (TM) Reverse Glenosphere 36mm Part 00-4349-036-11 The reverse shoulder system is used in the treatment of the following: severe pain or significant disability in
- Z-1079-2014 · Oct 17, 2013 · TerminatedTrabecular Metal" Reverse Shoulder System
- Z-2055-2013 · Aug 1, 2013 · Terminated00 4309-049-00 Trabecular Metal Reverse Shoulder Glenosphere Distractor Rx, Non sterile (sterilize before use) The Glenosphere Distractor must be able to attach, connect, or mate
- Z-0571-2011 · Sep 10, 2010 · TerminatedZIMMER TRABECULAR METAL" SHOULDER INSTRUMENTATION, DISTAL PILOT used for implantation of the Zimmer Trabecular Metal Humeral and/or Zimmer Trabecular Metal Reverse Shoulder Systems
- Z-0573-2011 · Sep 10, 2010 · TerminatedTRABECULAR METAL" SHOULDER REAMER CASE used for implantation of the Zimmer Trabecular Metal Humeral and/or Zimmer Trabecular Metal Reverse Shoulder Systems: Catalog Number 00-4312-
- Z-0575-2011 · Sep 10, 2010 · TerminatedZimmer Trabecular Metal Humeral Stem 4-Part Fracture Surgical Technique. Catalog Number 97-4309-004-00, Rev. 3. The Zimmer Trabecular Metal Shoulder System consists of convention
- Z-0576-2011 · Sep 10, 2010 · TerminatedZimmer Trabecular Metal Reverse Shoulder System Surgical Technique. Catalog Number 97-4309-003-00, Rev. 4. The Zimmer Trabecular Metal Reverse Shoulder System is a treatment option
- Z-0570-2011 · Sep 10, 2010 · TerminatedZIMMER TRABECULAR METAL SHOULDER INSTRUMENTATION, DISTAL PILOT used for implantation of the Zimmer Trabecular Metal Humeral and/or Zimmer Trabecular Metal Reverse Shoulder Systems:
- Z-0574-2011 · Sep 10, 2010 · TerminatedZimmer Trabecular Metal Humeral Stem Surgical Technique. Catalog Number: 97-4309-002-00, Rev. 3 The Zimmer Trabecular Metal Shoulder System consists of conventional and reverse, s
- Z-1416-06 · Jul 20, 2006 · TerminatedZimmer Trabecular Metal Reverse Shoulder System Glenosphere, 36 mm diameter; Catalog/Ref No. 00-4349-036-01.
- Z-1417-06 · Jul 20, 2006 · TerminatedZimmer Trabecular Metal Reverse Shoulder System Glenosphere, 40 mm diameter; Catalog/Ref No. 00-4349-040-01.
- Z-1418-06 · Jul 20, 2006 · TerminatedZimmer Trabecular Metal Reverse Shoulder System Base Plate; Catalog/Ref No. 00-4349-038-00.
Other PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED devices
- Equinoxe 320-42-00Exactech, Inc. · Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented3 recalls107 reports
- Equinoxe 320-38-00Exactech, Inc. · Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented3 recalls107 reports
- Equinoxe 320-42-03Exactech, Inc. · Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented2 recalls87 reports
- Equinoxe 320-38-03Exactech, Inc. · Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented2 recalls43 reports
- Equinoxe 320-42-13Exactech, Inc. · Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented2 recalls41 reports
- Equinoxe 320-10-00Exactech, Inc. · Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented132 reports
- Equinoxe 320-36-00Exactech, Inc. · Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented2 recalls27 reports
- Equinoxe 320-42-10Exactech, Inc. · Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented2 recalls14 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page