Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00889024278004), received between Jan 28, 2021 and Jun 19, 2023 — classified as 2 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Fracture” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Fracture3
- Break1
- Material Deformation1
- Material Integrity Problem1
- Mechanical Jam1
- Mechanical Problem1
Patient problems
- Insufficient Information3
- No Clinical Signs, Symptoms or Conditions2
- Foreign Body In Patient1
- Limb Fracture1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 19, 2023Fracture; Fracture
IT WAS REPORTED THAT DURING A TRIAL OF A STEM, THE SET SCREW BROKE WHILE USING THE SLAPHAMMER. THE EXACT CAUSE OF THE BREAKAGE IS UNCERTAIN, BUT IT IS POSSIBLE THAT THE STEM WAS OVERSIZED OR NOT REAMED SUFFICIENTLY DISTALLY. AT THE TIME OF THE BREAK, THE DEVICE WAS LOCATED INSIDE THE FEMORAL CANAL. TO REMOVE THE DEVICE, AN OSTEOTOMY WAS REQUIRED. IT IS LIKELY THAT DEVIATING FROM THE RECOMMENDED TECHNIQUE, SPECIFICALLY OVER-REAMING OR INADEQUATE REAMING DISTALLY, CONTRIBUTED TO THIS INCIDENT. THE OVERALL DELAY CAUSED BY THIS EVENT WAS APPROXIMATELY ONE HOUR. DUE DILIGENCE IS IN PROGRESS FOR THIS EVENT; TO DATE NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: Insufficient Information
MAUDE report 17152026 ↗InjuryFeb 25, 2022Break; Fracture; Fracture
IT WAS REPORTED THAT WHILE SURGEON WAS REMOVING WAGNER SL TRIAL STEM 16X190MM, THE SCREW BROKE, LEAVING IN THE FEMORAL CANAL THE DISTAL 16MM TRIAL. AFTER ADEPTS AT EXTRACTING DISTAL TRIAL STEM, SURGEON PERFORMED OSTEOTOMY TO RETRIEVE DISTAL TRIAL STEM. PATIENT INVOLVEMENT - PROLONGED SURGERY 60 MINUTE DELAY.
Patient problems: Insufficient Information
MAUDE report 13599223 ↗MalfunctionDec 2, 2021Mechanical Problem; Mechanical Jam; Mechanical Problem; Mechanical Jam
IT WAS REPORTED THAT WHEN TAKING STEM OFF OF THE NECK, THE TRIAL SCREW BECAME STUCK IN THE NECK. WHILE TRYING TO GET THE SCREW OUT, IT BECAME STRIPPED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12916640 ↗MalfunctionDec 2, 2021Material Deformation; Material Integrity Problem; Material Deformation; Material Integrity Problem
IT WAS REPORTED THAT A SCREW HEAD STRIPPED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 12915043 ↗InjuryJan 28, 2021Fracture; Fracture
NO CHANGE TO PREVIOUSLY REPORTED EVENT.
Patient problems: Foreign Body In Patient; Limb Fracture
MAUDE report 11242714 ↗
Other Prosthesis, hip, semi-constrained, metal/polymer, cemented devices
- NA 1365-51-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented753 reports
- NA 1365-50-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented412 reports
- ARTICUL/EZE 1365-21-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 recall280 reports
- ARTICUL/EZE 1365-11-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 death262 reports
- LFIT 6260-9-236Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, cemented256 reports
- PINNACLE GRIPTION 1217-32-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented2 deaths239 reports
- ELEOS™ 25002111EOnkos Surgical, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented3 deaths230 reports
- Novation 130-36-52Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented2 recalls124 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page