TRABECULAR METAL® 00-5450-021-35Recalls & reported problems
Zimmer Trabecular Metal Technology Inc. · Class 2 · Implant
All models of TRABECULAR METAL (31)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00889024308466), received between Sep 14, 2023 and Dec 31, 2025 — classified as 2 injuries.
The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Loosening of Implant Not Related to Bone-Ingrowth2
Patient problems
- Pain2
- Ambulation or Postural Difficulties1
- Aseptic loosening1
- Failure of Implant1
- Joint Dislocation1
- Joint effusion1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryDec 31, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
IT WAS REPORTED THAT A PATIENT UNDERWENT AN INITIAL TOTAL KNEE ARTHROPLASTY. APPROXIMATELY NINE (9) MONTHS POST-IMPLANTATION, THE PATIENT BEGAN TO EXPERIENCE PAIN IN THE IMPLANTED JOINT. SUBSEQUENTLY, A REVISION SURGERY WAS PERFORMED DUE TO LOOSENING OF THE FEMORAL COMPONENTS. ALL AVAILABLE INFORMATION HAS BEEN PROVIDED.
Patient problems: Aseptic loosening; Pain
MAUDE report 23933422 ↗InjurySep 14, 2023Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
IT WAS REPORTED THAT A PATIENT UNDERWENT A REVISION TOTAL KNEE ARTHROPLASTY. APPROXIMATELY ONE YEAR AND NINE MONTHS POST-IMPLANTATION, THE PATIENT BEGAN TO EXPERIENCE PAIN, INSTABILITY, IMMOBILITY AND MODERATE JOINT EFFUSION. DIAGNOSTIC IMAGES REVEALED THAT THE LOCKING BAR COMPONENT OF THE ARTICULAR SURFACE HAD FRACTURED. SUBSEQUENTLY, THE PATIENT UNDERWENT REVISION SURGERY. DURING THE SURGERY, THE KNEE JOINT WAS FOUND TO BE UNCOUPLED FROM THE CONNECTOR DUE TO BREAKAGE OF THE LOCKING BOLT AS WELL AS DORSAL DISLOCATION OF THE ARTICULAR SURFACE AND LOOSENING OF THE CEMENTED FEMORAL COMPONENT. THE FEMORAL COMPONENTS AND ARTICULAR SURFACE WERE REVISED WITHOUT COMPLICATION. DUE DILIGENCE IS COMPLETE AS MULTIPLE ATTEMPTS HAVE BEEN MADE; ALL AVAILABLE INFORMATION HAS BEEN REPORTED.
Patient problems: Ambulation or Postural Difficulties; Failure of Implant; Joint Dislocation; Joint effusion; Pain
MAUDE report 17745668 ↗
Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.