Skip to content

NA 00-7050-034-00Recalls & reported problems

Zimmer Trabecular Metal Technology Inc. · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00889024311411), received between Mar 11, 2025 and Apr 24, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Fracture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture2
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 24, 2026Material Deformation

    IT WAS REPORTED THAT THE HANDLE WAS STRIPPED WHEN ATTACHING THE TRIAL, AND THE INSTRUMENT WAS NOT USED IN THE CASE. THERE WAS NO PATIENT INVOLVEMENT. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24982149 ↗
  • MalfunctionApr 22, 2026Fracture

    IT WAS REPORTED THAT THE TIP OF BLACK HANDLE CUP IMPACTOR HAS BROKEN OFF INTO THE THREADED HOLE OF THE 62/64 PIE IMPACTOR. THERE IS NO FURTHER INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24956510 ↗
  • MalfunctionMar 11, 2025Fracture; Fracture

    IT WAS REPORTED THAT THE TIP OF BLACK HANDLE CUP IMPACTOR HAS BROKEN OFF INTO THE THREADED HOLE OF THE 62/64 PIE IMPACTOR. THERE IS NO FURTHER INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21576721 ↗

Other PROBE devices

All PROBE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page