Tapered Screw-Vent® TSVM4H10Recalls & reported problems
ZimVie US Corp LLC · Class 2 · Implant
All models of Tapered Screw-Vent (131)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 18
- Deaths reported
- 0
- Injuries reported
- 12
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 18 adverse event reports that list this exact device (DI 00889024343993), received between Jun 12, 2020 and Mar 9, 2026 — classified as 12 injuries, 6 malfunctions.
The most frequently reported specific device problem is “Separation Failure” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem9
- Separation Failure4
- Fracture3
- Detachment of Device or Device Component1
- Mechanical Problem1
Patient problems
- Unspecified Infection8
- No Clinical Signs, Symptoms or Conditions7
- Inflammation4
- Osteolysis2
- Erosion1
- No Information1
Recent adverse event reports
The 18 most recent of 18 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 9, 2026Mechanical Problem
NO ADDITIONAL EVENT INFORMATION RECEIVED AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24550287 ↗InjuryNov 7, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #6 WAS REMOVED DUE TO INFECTION. SYMPTOMS AS A RESULT OF THE EVENT: INFLAMMATION.
Patient problems: Inflammation; Unspecified Infection
MAUDE report 23495772 ↗InjuryJul 24, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #11 WAS REMOVED DUE TO INFECTION. SYMPTOMS AS A RESULT OF THE EVENT: ABSCESS.
Patient problems: Erosion; Unspecified Infection
MAUDE report 22587420 ↗MalfunctionJul 15, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT THE IMPLANT FELL OUT OF THE MOUNT. WHEN REMOVING THE IMPLANT FROM THE CASE, IT FELL OUT OF THE MOUNT. THE DOCTOR COMMENTED THAT IT WAS LIKELY LOOSE FROM THE BEGINNING. THE DOCTOR SAID HE HAD SEEN SIMILAR CASES TWICE BEFORE. THE PROCEDURE WAS COMPLETED USING ANOTHER IMPLANT OR ANOTHER DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22506436 ↗InjuryMar 27, 2025Fracture; Fracture
IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #19 WAS REMOVED AS IT WAS FRACTURED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21712706 ↗InjuryMar 10, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE IMPLANT AT TOOTH LOCATION #22 WAS INFECTED AND WAS REMOVED.
Patient problems: Unspecified Infection
MAUDE report 21559960 ↗MalfunctionDec 2, 2024Separation Failure; Separation Failure
NO ADDITIONAL EVENT INFORMATION RECEIVED AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20819169 ↗InjuryOct 16, 2024Fracture; Fracture
IT WAS REPORTED THAT THE PATIENT PRESENTED WITH THE CROWN OFF OF THE IMPLANT AT TOOTH LOCATION #19. THE OUTSIDE COLLAR OF THE IMPLANT WAS NOTED TO BE FRACTURED. THE PATIENT WILL BE SEEN BY A PERIODONTIST FOR REMOVAL AND TO PLACE A NEW IMPLANT. SYMPTOMS AS A RESULT OF THE EVENT: INFLAMMATION.
Patient problems: Inflammation
MAUDE report 20464368 ↗InjuryApr 30, 2024Fracture; Fracture
IT WAS REPORTED THAT DURING CLEANING, THE IMPLANT AT TOOTH LOCATION #4 WAS FOUND TO BE FRACTURED. THE IMPLANT WAS REMOVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19213827 ↗InjuryApr 25, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE IMPLANT BECOME MOBILE AND WAS REMOVED DUE TO BONE LOSS. TOOTH #22 (UNIVERSAL).
Patient problems: Osteolysis
MAUDE report 19180372 ↗InjuryFeb 15, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT WENT TO THE CLINIC TO CHECK IMPLANT MOBILITY AND PAIN AROUND THE IMPLANT AT TOOTH SITE #6. THE IMPLANT WAS REMOVED DUE TO PERI-IMPLANTITIS. SYMPTOMS AS A RESULT OF THE EVENT: PAIN AND INFLAMMATION.
Patient problems: Inflammation; Osteolysis; Unspecified Infection
MAUDE report 18711034 ↗InjuryOct 13, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE IMPLANT IN SITE 13 WAS REMOVED DUE TO INFECTION. SITE WAS NOT GRAFTED, NO IMPACT TO PATIENT AND NO MEDICAL HISTORY REPORTED. SYMPTOMS: INFLAMMATION.
Patient problems: Inflammation; Unspecified Infection
MAUDE report 15597594 ↗InjuryJul 11, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE IMPLANT AT TOOTH LOCATION #2 WAS REMOVED DUE TO INFECTION.
Patient problems: Unspecified Infection
MAUDE report 14981661 ↗InjuryJul 11, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE IMPLANT AT TOOTH LOCATIONS #14 AND #15 WERE REMOVED DUE TO INFECTION.
Patient problems: Unspecified Infection
MAUDE report 14978486 ↗InjuryJul 11, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE IMPLANT AT TOOTH LOCATION #29 WAS REMOVED DUE TO INFECTION.
Patient problems: Unspecified Infection
MAUDE report 14978122 ↗MalfunctionFeb 24, 2022Separation Failure; Separation Failure
IT WAS REPORTED THAT DR. INDICATED DEVICE MALFUNCTION. THE IMPLANT BODY CANNOT BE REMOVED FROM THE FIXTURE MOUNT. THE IMPLANT WAS REMOVED AND ANOTHER IMPLANT WAS PLACED. TOOTH # 30.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13593876 ↗MalfunctionAug 19, 2021Separation Failure; Separation Failure
IT WAS REPORTED THAT THE IMPLANT BODY CANNOT BE REMOVED FROM THE FIXTURE MOUNT. THE IMPLANT WAS REMOVED AND ANOTHER IMPLANT WAS PLACED. TOOTH SITE #20.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12344615 ↗MalfunctionJun 12, 2020Separation Failure; Separation Failure
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Information
MAUDE report 10148646 ↗
Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM43,616 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM40,404 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM27,400 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,621 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLA Titanium BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,298 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM23,448 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,320 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,267 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page