Adverse event reports
16
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 00889024375123), received between Aug 22, 2023 and Jun 13, 2025 — classified as 16 malfunctions.

The most frequently reported specific device problem is “Battery Problem” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 4 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 3 reportsMar 2025: 1 reportsApr 2025: 3 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Battery Problem7
  • Explosion6
  • Failure to Power Up3
  • Powder Leak3
  • Material Deformation2
  • Activation Problem1
  • Corroded1
  • Device Contaminated During Manufacture or Shipping1

Patient problems

  • No Clinical Signs, Symptoms or Conditions16
  • Insufficient Information1

Recent adverse event reports

The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 13, 2025Explosion; Powder Leak; Explosion; Powder Leak

    No further event information is available at the time of this report.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22202672 ↗
  • MalfunctionMay 30, 2025Explosion; Explosion

    IT WAS REPORTED THAT PRIOR TO SURGERY WHILE OPENING THE DELIVERY BOXES, THE PULSAVAC BATTERY HAD EXPLODED INSIDE THE PACKAGING. THERE WAS NO PATIENT INVOLVEMENT. DILIGENCE IS COMPLETE, THERE IS NO ADDITIONAL INFORMATION AVAILABLE.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 22120013 ↗
  • MalfunctionApr 21, 2025Failure to Power Up; Material Deformation; Powder Leak; Failure to Power Up; Material Deformation; Powder Leak

    IT WAS REPORTED THAT DURING SURGERY, DID NOT TURN ON. WHEN THE BATTERY PACK WAS INSPECTED THERE WAS SOME DEFORMATION OF THE PLASTIC NOTED ON THE OUTSIDE. WHEN THE BATTERY PACK WAS OPENED FOR FURTHER INSPECTION, IT WAS DISCOVERED THAT 4 OF THE BATTERIES WERE CORRODED. THERE WAS NO PATIENT INJURY, OR DELAY. DUE DILIGENCE IS COMPLETE AND THERE IS NO ADDITIONAL INFORMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21874510 ↗
  • MalfunctionApr 18, 2025Failure to Power Up; Material Deformation; Powder Leak; Failure to Power Up; Material Deformation; Powder Leak

    NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21864394 ↗
  • MalfunctionApr 1, 2025Expulsion; Explosion; Expulsion; Explosion

    IT WAS REPORTED THAT THERE WAS AN EXPLOSION INSIDE THE PACKAGING. THE EVENT TIMING WAS OUTSIDE OF SURGERY. THERE IS NO HARM OR DELAY REPORTED. THE PACKAGE WAS NOT OPENED. DUE DILIGENCE IS COMPLETE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21745502 ↗
  • MalfunctionMar 17, 2025Corroded; Leak/Splash; Explosion; Corroded; Leak/Splash; Explosion

    IT WAS REPORTED THAT THE PULSAVAC DID NOT FUNCTION, OPENED THE BATTERY PACK TO FIND CORROSION ON BATTERIES. PER THE IMAGE THE FIFTH BATTERY FROM THE RIGHT LEAKED ELECTROLYTE. THERE WAS NO PATIENT INVOLVEMENT. THERE IS NO HARM OR DELAY REPORTED. DUE DILIGENCE IS COMPLETE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21619244 ↗
  • MalfunctionFeb 21, 2025Activation Problem; Activation Problem

    IT WAS REPORTED THAT DURING SURGERY THE PULSAVAC DEVICE WOULD POWER ON BUT WOULD NOT WORK. THE BATTERY WAS REPLACED BUT STILL WOULD NOT WORK. THERE WAS NO REPORT OF HARM OR DELAY. DILIGENCE IS COMPLETE. NO ADDITIONAL INFORMATION IS AVAILABLE. DURING DEVICE EVALUATION VISUAL INSPECTION OF THE INTERIOR OF THE PACK FOUND THE BATTERIES HAD LEAKED ELECTROLYTE INSIDE OF THE PACK. NO ADVERSE EVENTS WERE REPORTED AS A RESULT OF THIS MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21432345 ↗
  • MalfunctionFeb 10, 2025Explosion; Explosion

    IT WAS REPORTED THAT BEFORE SURGERY, THE BATTERIES EXPLODED INSIDE PULSAVAC BOX. THERE WAS NO PATIENT INVOLVEMENT. DUE DILIGENCE IS COMPLETE, NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21344494 ↗
  • MalfunctionFeb 10, 2025Explosion; Explosion

    IT WAS REPORTED THAT BEFORE SURGERY, THE BATTERIES EXPLODED INSIDE PULSAVAC BOX. THERE WAS NO PATIENT INVOLVEMENT. DUE DILIGENCE IS COMPLETE, NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21344410 ↗
  • MalfunctionFeb 20, 2024Battery Problem; Device Contaminated During Manufacture or Shipping; Battery Problem; Device Contaminated During Manufacture or Shipping

    IT WAS REPORTED THAT THE BATTERY EXPLODED SO HE COULDN'T USE IT. THE EVENT OCCURRED DURING SURGERY .THERE WAS NO REPORTED PATIENT HARM, OR DELAY. DUE DILIGENCE IS COMPLETE, NO FURTHER INFORMATION IS AVAILABLE AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18740020 ↗
  • MalfunctionDec 6, 2023Failure to Power Up; Battery Problem; Failure to Power Up; Battery Problem

    IT WAS REPORTED THAT AT THE BEGINNING OF SURGERY THE BATTERIES IN SYSTEM WERE CORRODED, AND THE DEVICE DID NOT WORK. THE DEVICE WOULD NOT TURN ON. NO IMPACT, HARM, OR DELAY WERE REPORTED. DUE DILIGENCE INFORMATION IS COMPLETE. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18277957 ↗
  • MalfunctionNov 6, 2023Battery Problem; Battery Problem

    IT WAS REPORTED THAT DURING SURGERY THE BATTERY WAS LEAKING ACID UPON OPENING. THERE WAS NO PATIENT HARM AND NO DELAY NOTED. AN ALTERNATE DEVICE WAS UTILIZED TO COMPLETE THE PROCEDURE. DUE DILIGENCE IS IN PROCESS AND THERE IS NO ADDITIONAL INFORMATION AVAILABLE. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18074151 ↗
  • MalfunctionNov 6, 2023Battery Problem; Battery Problem

    IT WAS REPORTED THAT DURING SURGERY THE BATTERY WAS LEAKING ACID UPON OPENING. THERE WAS NO PATIENT HARM AND NO DELAY NOTED. AN ALTERNATE DEVICE WAS UTILIZED TO COMPLETE THE PROCEDURE. DUE DILIGENCE IS IN PROCESS AND THERE IS NO ADDITIONAL INFORMATION AVAILABLE. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18074144 ↗
  • MalfunctionNov 6, 2023Battery Problem; Battery Problem

    IT WAS REPORTED THAT DURING SURGERY THE BATTERY WAS LEAKING ACID UPON OPENING. THERE WAS NO PATIENT HARM AND NO DELAY NOTED. AN ALTERNATE DEVICE WAS UTILIZED TO COMPLETE THE PROCEDURE. DUE DILIGENCE IS IN PROCESS AND THERE IS NO ADDITIONAL INFORMATION AVAILABLE. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THE MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18074136 ↗
  • MalfunctionNov 6, 2023Battery Problem; Battery Problem

    IT WAS REPORTED THAT DURING SURGERY THE BATTERY WAS LEAKING ACID UPON OPENING. THERE WAS NO PATIENT HARM AND NO DELAY NOTED. AN ALTERNATE DEVICE WAS UTILIZED TO COMPLETE THE PROCEDURE. DUE DILIGENCE IS IN PROCESS AND THERE IS NO ADDITIONAL INFORMATION AVAILABLE. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18074114 ↗
  • MalfunctionAug 22, 2023Battery Problem; Manufacturing, Packaging or Shipping Problem; Battery Problem; Manufacturing, Packaging or Shipping Problem

    THERE IS NO ADDITIONAL INFORMATION AVAILABLE REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17597192 ↗