PULSAVAC® 00-5150-495-00Recalls & reported problems
Zimmer Orthopaedic Surgical Products, Inc. · Shower Spray Kit · Class 2
All models of PULSAVAC (18)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 00889024375215), received between Apr 16, 2024 and Jun 24, 2026 — classified as 12 malfunctions.
The most frequently reported specific device problem is “Battery Problem” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Battery Problem4
- Delivered as Unsterile Product4
- Overheating of Device4
- Explosion3
- Device Contaminated During Manufacture or Shipping2
- Powder Leak2
- Expulsion1
- Failure to Power Up1
Patient problems
- No Clinical Signs, Symptoms or Conditions11
- Insufficient Information1
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 24, 2026Overheating of Device
IT WAS REPORTED THAT DURING AN UNKNOWN TIMING, THE UNIT WAS GETTING VERY HOT AS WELL AS THE BATTERY PACK. BATTERY EXPULSION WAS CONFIRMED AFTER FINAL INVESTIGATION. IT IS UNKNOWN IF THERE WAS A PATIENT IMPACT OR IF A DELAY OCCURRED. DUE DILIGENCE IS COMPLETE.
Patient problems: Insufficient Information
MAUDE report 25608621 ↗MalfunctionApr 27, 2026Powder Leak
IT WAS REPORTED THAT DURING SURGERY BATTERY LOOKS TO HAVE MALFUNCTIONED WITHIN THE PACKAGING. THERE WAS BLACK DEBRIS WITHIN THE PACKAGING. THE BATTERY PACK MAY HAVE EXPLODED. THERE WAS NO PATIENT IMPACT, NO DELAY, AND NO MEDICAL INTERVENTION. DUE DILIGENCE IS COMPLETE AS NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25008350 ↗MalfunctionFeb 20, 2026Failure to Power Up
IT WAS REPORTED THAT DURING SURGERY THE DEVICE WOULD NOT POWER ON AND HAD NO POWER, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THERE WERE NO CONSEQUENCES OR IMPACT TO THE PATIENT. AT INVESTIGATION IT WAS FOUND THAT BATTERIES HAD LEAKED ELECTROLYTE. DILIGENCE IS COMPLETE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24401053 ↗MalfunctionMar 25, 2025Overheating of Device; Overheating of Device
IT WAS REPORTED THAT DURING SURGERY ON THE PULSAVAC BATTERY PACK QUICKLY BECAME HOT. THERE WERE NO SIGNS OF DAMAGE OR LEAKAGE FROM BATTERY REPORTED. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS A SHORT DELAY IN PROCEDURE TO PREPARE THE ALTERNATE DEVICE. NO PATIENT HARM/IMPACT WAS REPORTED. DUE DILIGENCE INFORMATION IS COMPLETE. DURING DEVICE EVALUATION VISUAL INSPECTION OF THE INTERIOR OF THE PACK FOUND THE BATTERIES HAD LEAKED ELECTROLYTE INSIDE OF THE PACK.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21687317 ↗MalfunctionOct 31, 2024Explosion; Powder Leak; Explosion; Powder Leak
THERE IS NO ADDITIONAL EVENT INFORMATION AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20575672 ↗MalfunctionOct 18, 2024Overheating of Device; Expulsion; Overheating of Device; Expulsion
DUE DILIGENCE IS COMPLETE AND NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20482221 ↗MalfunctionSep 9, 2024Delivered as Unsterile Product; Delivered as Unsterile Product
DUE DILIGENCE IS COMPLETE AND NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20171278 ↗MalfunctionAug 5, 2024Delivered as Unsterile Product; Battery Problem; Delivered as Unsterile Product; Battery Problem
NO ADDITIONAL EVENT INFORMATION AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19905653 ↗MalfunctionAug 5, 2024Delivered as Unsterile Product; Battery Problem; Explosion; Delivered as Unsterile Product; Battery Problem; Explosion
THERE IS NO ADDITIONAL INFORMATION AVAILABLE REGARDING THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19905618 ↗MalfunctionMay 20, 2024Overheating of Device; Battery Problem; Device Contaminated During Manufacture or Shipping; Overheating of Device; Battery Problem; Device Contaminated During Manufacture or Shipping
NO ADDITIONAL EVENT INFORMATION AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19350973 ↗MalfunctionMay 1, 2024Battery Problem; Device Contaminated During Manufacture or Shipping; Explosion; Battery Problem; Device Contaminated During Manufacture or Shipping; Explosion
THERE WAS NO ADDITIONAL INFORMATION AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19224180 ↗MalfunctionApr 16, 2024Delivered as Unsterile Product; Delivered as Unsterile Product
THERE IS NO ADDITIONAL EVENT INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19121507 ↗
Other LAVAGE, JET devices
- Elefant ASP185Coloplast A/S · LAVAGE, JET260 reports
- PULSAVAC® 00-5150-475-01Zimmer Orthopaedic Surgical Products, Inc. · LAVAGE, JET255 reports
- PULSAVAC® 00-5150-482-00Zimmer Orthopaedic Surgical Products, Inc. · LAVAGE, JET197 reports
- PULSAVAC® 00-5150-475-00Zimmer Orthopaedic Surgical Products, Inc. · LAVAGE, JET164 reports
- Versajet 66800039Smith & Nephew Medical Limited · Lavage, Jet160 reports
- PULSAVAC® 00-5150-482-01Zimmer Orthopaedic Surgical Products, Inc. · LAVAGE, JET154 reports
- Surgiphor 910110Carefusion 2200, Inc · LAVAGE, JET140 reports
- Axius Blower/Mister CB-1000Maquet Cardiovascular LLC · Lavage, Jet1 recall43 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page