Skip to content

PULSAVAC® 00-5150-495-01Recalls & reported problems

Zimmer Orthopaedic Surgical Products, Inc. · Class 2

All models of PULSAVAC (18)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00889024375222), received between Oct 23, 2024 and Jan 16, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Failure to Power Up” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Appropriate Device Problem Term/Code Not Available1
  • Failure to Power Up1
  • Inability to Irrigate1
  • Powder Leak1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 16, 2025Powder Leak; Powder Leak

    IT WAS REPORTED THAT DURING INSPECTION THERE WAS BATTERY CORROSION. THERE WAS NO PATIENT INVOLVEMENT. DUE DILIGENCE IS COMPLETE AND THERE IS NO ADDITIONAL INFORMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21172080 ↗
  • MalfunctionOct 24, 2024Inability to Irrigate; Failure to Power Up; Inability to Irrigate; Failure to Power Up

    IT WAS REPORTED THAT DURING SURGERY, THE PULSE WASHING GUN WAS OPENED LN WHICH DID NOT WORK WHEN TRYING TO WASH THE JOINT. THE PULSATILE WASHING DIDN'T EVEN TURN ON. THERE WAS NO PATIENT HARM OR INJURY. THERE WAS NO MEDICAL INTERVENTION. THERE WAS A 15-MINUTE SURGICAL DELAY. DURING DEVICE EVALUATION AND VISUAL INSPECTION, IT WAS DISCOVERED THAT THE INTERIOR OF THE PACK FOUND THE BATTERIES HAD LEAKED ELECTROLYTE INSIDE OF THE PACK. DUE DILIGENCE IS COMPLETE, THERE IS NO ADDITIONAL INFORMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20521902 ↗
  • MalfunctionOct 23, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED DURING SURGERY THAT THE PRODUCT DIDN'T WORK AT ALL FROM THE BEGINNING. THERE IS NO HARM AND A 0-15 MINUTE DELAY REPORTED. DUE DILIGENCE IS COMPLETE AS THERE IS NO ADDITIONAL INFORMATION AVAILABLE. UPON INVESTIGATION, THE BATTERIES HAD LEAKED ELECTROLYTE INSIDE OF THE PACK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20517625 ↗

Other LAVAGE, JET devices

All LAVAGE, JET devices →

Sources and updates