Biolox® delta 00-8775-028-02Recalls & reported problems

Zimmer GmbH · BIOLOX® DELTA, CERAMIC FEMORAL HEAD, M, Ø 28/0, TAPER 12/14 · Class 2 · Implant

All models of Biolox delta (16)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Zimmer BIOLOX delta ceramic hip implant (femoral head, 28 mm) is used in hip replacement surgery. The 36 adverse event reports mostly describe the hip joint coming apart or dislocating after surgery, sometimes soon after implantation and sometimes years later, with several patients needing revision surgery to replace the stem, neck or bearing parts. There were 33 injuries and no deaths. Three malfunctions were also reported, including a head that did not seat properly in its bearing. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
37
Deaths reported
0
Injuries reported
34
Recalls naming it
0

Summary

The FDA's MAUDE database contains 37 adverse event reports that list this exact device (DI 00889024430297), received between Mar 5, 2019 and Sep 29, 2026 — classified as 34 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (19 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 2 reportsAug 2024: 2 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated19
  • Adverse Event Without Identified Device or Use Problem9
  • Detachment of Device or Device Component3
  • Fracture2
  • Insufficient Device Problem Information2
  • Activation, Positioning or Separation Problem1
  • Biocompatibility1
  • Mechanical Problem1

Patient problems

  • Joint Dislocation19
  • Failure of Implant8
  • Pain4
  • Unspecified Infection4
  • Fall3
  • Ambulation or Postural Difficulties2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
  • Bone Fracture(s)1

Recent adverse event reports

The 25 most recent of 37 reports, in the reporter's own words as published by the FDA.

  • InjurySep 29, 2026Naturally Worn

    Patient problems: Insufficient Information

    MAUDE report 50932266 ↗
  • InjuryMay 15, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A HIP REVISION APPROXIMATELY 19 MONTHS POST-IMPLANTATION DUE TO A DISASSOCIATED DUAL MOBILITY BEARING AND HEAD. NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: Joint Dislocation

    MAUDE report 25179916 ↗
  • InjuryFeb 19, 2026Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT UNDERWENT PRIMARY SURGERY, AND SUBSEQUENTLY FELL BACKWARD, RESULTING IN AN ANTERIOR DISLOCATION AND REQUIRING REVISION SURGERY APPROXIMATELY 2 YEARS, 8 MONTHS POST PRIMARY SURGERY. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: Joint Dislocation

    MAUDE report 24386196 ↗
  • InjuryJan 2, 2026Adverse Event Without Identified Device or Use Problem

    THERE IS NO UPDATE TO THE PRIOR EVENT DESCRIPTION PROVIDED.

    Patient problems: Unspecified Infection

    MAUDE report 23941631 ↗
  • InjuryOct 28, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Joint Dislocation; Metal Related Pathology

    MAUDE report 23405138 ↗
  • InjuryJul 22, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THE PATIENT UNDERWENT A HIP REVISION APPROXIMATELY 1 WEEK POST IMPLANTATION DUE TO DISLOCATION. THERE WAS EARLY POSTOPERATIVE DISLOCATION; STEM AND NECK WERE REPLACED. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.

    Patient problems: Joint Dislocation

    MAUDE report 22565394 ↗
  • InjuryJun 9, 2025Separation Failure; Detachment of Device or Device Component; Separation Failure; Detachment of Device or Device Component

    IT WAS REPORTED THAT PATIENT UNDERWENT A REVISION APPROXIMATELY 3 YEARS POST IMPLANTATION DUE TO IMPLANT DISASSEMBLY (BETWEEN BEARING AND HEAD). ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.

    Patient problems: Failure of Implant

    MAUDE report 22168862 ↗
  • InjuryApr 25, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED PATIENT UNDERWENT HIP REVISION APPROXIMATELY 1 YEAR AND 4 MONTHS POST IMPLANTATION DUE TO DISLOCATION. MODULAR NECK, HEAD, AND DM BEARINGS WERE REPLACED IN THIS SURGERY. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.

    Patient problems: Joint Dislocation

    MAUDE report 21909738 ↗
  • MalfunctionAug 8, 2024Positioning Problem; Positioning Problem

    IT WAS REPORTED IN HIP REPLACEMENT THAT THE HEAD DO NOT ASSEMBLY AS INTENDED WITHIN THE BEARING DISPLAYED SIGNIFICANT PLAY. THERE WAS NO SURGICAL DELAY OR COMPLICATIONS. FOREIGN BODY WAS NOT RETAINED. ANOTHER HEAD WAS USED TO COMPLETE THE SURGERY. DILIGENCE IS COMPLETED AND NO FURTHER INFORMATION ON THE REPORTED EVENT HAS BEEN PROVIDED.

    Patient problems: Failure of Implant

    MAUDE report 19944024 ↗
  • InjuryAug 8, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT APPROXIMATELY TWO YEARS POST-IMPLANTATION, THE PATIENT UNDERWENT A RIGHT HIP REVISION DUE TO INSTABILITY. THE BEARING AND HEAD WERE REVISED. DUE DILIGENCE HAS BEEN COMPLETED FOR THIS COMPLAINT; TO DATE ALL AVAILABLE ADDITIONAL INFORMATION HAS BEEN RECEIVED.

    Patient problems: Joint Laxity

    MAUDE report 19938325 ↗
  • InjuryJul 8, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT APPROXIMATELY ONE DAY POST-IMPLANTATION, THE PATIENT UNDERWENT A HIP REVISION DUE TO A FALL AND HIP DISLOCATION. DURING SURGERY, THE SURGEON FOUND AN INTRA-PROSTHETIC DISLOCATION OF THE HEAD OUT OF THE BEARING. THE HEAD AND BEARING WERE REVISED. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: Failure of Implant; Fall; Joint Dislocation; Pain

    MAUDE report 19691718 ↗
  • InjuryJul 3, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A HIP PROCEDURE AND SUBSEQUENTLY WAS REVISED APPROXIMATELY 5 YEARS POST IMPLANTATION DUE TO DISLOCATION. THE CAUSE OF THE DISLOCATION IS UNKNOWN. THE STEM, HEAD, AND BEARING WERE REPLACED. DILIGENCE IS COMPLETED AND NO FURTHER INFORMATION ON THE REPORTED EVENT HAS BEEN PROVIDED.

    Patient problems: Joint Dislocation

    MAUDE report 19665405 ↗
  • InjuryApr 11, 2024Fracture; Fracture

    IT WAS REPORTED IN A PRIMARY HIP SURGERY THE CERAMIC HEAD WAS CHIPPED FRACTURED/CHIPPED DURING IMPACTION. THERE WAS NO IMPACT TO PATIENT AND THE SURGERY WAS COMPLETED WITH ANOTHER HEAD. DILIGENCE IS COMPLETED AND ADDITIONAL INFORMATION IS UNAVAILABLE AT THIS TIME.

    Patient problems: Failure of Implant

    MAUDE report 19088892 ↗
  • InjuryAug 29, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT A PATIENT WITH A HISTORY OF RHEUMATOID ARTHRITIS AND SCOLIOSIS UNDERWENT INITIAL RIGHT TOTAL HIP ARTHROPLASTY. SUBSEQUENTLY EXPERIENCED RECURRENT DISLOCATIONS AND UNDERWENT FIRST REVISION SURGERY. DURING WHICH, THE INITIAL SHELL WAS RETAINED, AND A NEW STEM, HEAD, AND DUAL MOBILITY BEARING WERE PLACED ALONG WITH A COMPETITOR METAL LINER. A FEW MONTHS LATER, THE PATIENT BENT OVER AND AGAIN DISLOCATED HER HIP WHICH WAS REDUCED IN THE EMERGENCY DEPARTMENT. X-RAYS SHOWED DISSOCIATION OF THE DUAL MOBILITY HEAD FROM THE BEARING. PATIENT UNDERWENT SECOND REVISION SURGERY DURING WHICH, THE COMPETITOR METAL LINER AND DUAL MOBILITY COMPONENTS WERE REPLACED WITH A NEW CERAMIC HEAD AND COMPETITOR CONSTRAINED LINER . THE INITIAL SHELL AND REVISION STEM REMAINED INTACT. DUE DILIGENCE IS IN PROGRESS FOR THIS COMPLAINT; TO DATE NO ADDITIONAL INFORMATION OR PRODUCT HAS BEEN RECEIVED.

    Patient problems: Ambulation or Postural Difficulties; Failure of Implant; Joint Dislocation

    MAUDE report 17641151 ↗
  • InjuryAug 23, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Joint Dislocation

    MAUDE report 17601590 ↗
  • InjuryAug 18, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED A PATIENT UNDERWENT RIGHT TOTAL HIP ARTHROPLASTY. SUBSEQUENTLY, THE PATIENT DISLOCATED WHICH WAS NOT REDUCIBLE. THE PATIENT WAS SUBSEQUENTLY REVISED AS IT WAS FOUND THERE WAS DISASSOCIATION OF THE IMPLANTS. ATTEMPTS HAVE BEEN MADE AND ALL ADDITIONAL INFORMATION RECEIVED HAS BEEN INCLUDED IN THIS REPORT.

    Patient problems: Joint Dislocation

    MAUDE report 17571839 ↗
  • InjuryJul 17, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Joint Dislocation

    MAUDE report 17332383 ↗
  • InjuryMay 12, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Failure of Implant; High metal ion levels; Pain

    MAUDE report 16917627 ↗
  • MalfunctionSep 5, 2022Activation, Positioning or Separation Problem; Activation, Positioning or Separation Problem

    IT WAS REPORTED THAT THE CERAMIC HEAD DID NOT ARTICULATE CORRECTLY WITHIN THE E1 BIPOLAR LINER. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT HEAD AND LINER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15354428 ↗
  • InjuryAug 30, 2022Unintended Movement; Unintended Movement

    IT WAS REPORTED PATIENT UNDERWENT A LEFT HIP REVISION APPROXIMATELY 2 MONTHS¿ POST IMPLANTATION DUE TO A DISLOCATION AFTER A FALL. BEFORE THE REVISION, THE ER FAILED TO COMPLETE A CLOSED REDUCTION. WHILE TRYING TO REDUCE THE HIP, THE CUP THEN DISASSOCIATED FROM THE ACETABULUM, DUE DILIGENCE IS IN PROGRESS FOR THIS COMPLAINT; TO DATE NO ADDITIONAL INFORMATION OR PRODUCT HAS BEEN RECEIVED.

    Patient problems: Joint Dislocation

    MAUDE report 15318170 ↗
  • InjuryAug 25, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL HIP ARTHROPLASTY ON AN UNKNOWN DATE. SUBSEQUENTLY, THE PATIENT WAS REVISED DUE TO DISLOCATION AFTER A FALL. DUE DILIGENCE IS IN PROGRESS FOR THIS COMPLAINT; TO DATE, NO ADDITIONAL INFORMATION OR PRODUCT HAS BEEN RECEIVED.

    Patient problems: Fall; Joint Dislocation

    MAUDE report 15291247 ↗
  • InjuryJun 28, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO DISLOCATION OF THE HIP TWO DAYS AFTER IMPLANTATION. NO FURTHER PATIENT OUTCOME PROVIDED.

    Patient problems: Joint Dislocation

    MAUDE report 14840847 ↗
  • InjuryApr 28, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT: ACTIVE ARTICULATION BEARING WAS DISLOCATED DUE TO DISASSOCIATION. THE SURGEON DID NOT COMMENT ON THIS PATIENT'S FACTORS FOR THE DISLOCATION. THE MODULAR HEAD AND NECK WERE ALSO REPLACED IN THIS REVISION SURGERY. PATIENT OUTCOME - REVISION DUE TO DISLOCATION.

    Patient problems: Joint Dislocation

    MAUDE report 14239274 ↗
  • InjurySep 3, 2021Adverse Event Without Identified Device or Use Problem

    INVESTIGATION HAS BEEN COMPLETED.

    Patient problems: Unspecified Infection

    MAUDE report 12421650 ↗
  • InjuryAug 25, 2021Fracture; Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    NO EVENT UPDATE. INVESTIGATION RESULTS ARE NOW AVAILABLE.

    Patient problems: Bone Fracture(s)

    MAUDE report 12364198 ↗

Other Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented devices

All Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented devices →