This Zimmer Persona Vivacit-E knee implant is used in knee replacement surgery. Reports mainly describe difficulty fitting or positioning the implant, along with limited movement, pain, swelling, and occasional damaged components requiring another implant or revision surgery. There are 26 reports, including 15 injuries and 11 malfunctions; no deaths were reported. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
29
Deaths reported
0
Injuries reported
17
Recalls naming it
0

Summary

The FDA's MAUDE database contains 29 adverse event reports that list this exact device (DI 00889024468498), received between Mar 23, 2020 and Sep 25, 2026 — classified as 17 injuries, 12 malfunctions.

The most frequently reported specific device problem is “Difficult to Insert” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 2 reportsAug 2025: 1 reportsSep 2025: 1 reportsOct 2025: 1 reportsNov 2025: 1 reportsDec 2025: 3 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 2 reportsSep 2026: 3 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem15
  • Difficult to Insert6
  • Device Contaminated During Manufacture or Shipping2
  • Positioning Failure2
  • Defective Device1
  • Fracture1
  • Material Integrity Problem1
  • Peeled/Delaminated1

Patient problems

  • No Clinical Signs, Symptoms or Conditions11
  • Loss of Range of Motion7
  • Pain5
  • Joint effusion3
  • Swelling/ Edema3
  • Adhesion(s)2
  • Arthrofibrosis2
  • Joint Laxity2

Recent adverse event reports

The 25 most recent of 29 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 25, 2026Device Contaminated During Manufacture or Shipping

    It was reported that a hair was observed inside the sterile package. There was no patient involvement. Attempts have been made and no further information has been provided.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50896920 ↗
  • InjurySep 22, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Arthrofibrosis

    MAUDE report 50858448 ↗
  • InjurySep 22, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that a patient underwent an initial right surgery. Subsequently, 2.5 years post-implantation a revision was performed due to instability. The articular surface was exchanged. Attempts have been made and no further information has been provided.

    Patient problems: Joint Laxity

    MAUDE report 50853366 ↗
  • InjuryAug 10, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the patient underwent a right knee arthroplasty revision to address tibial fibrous ingrowth approximately two (2) years post-operatively.Attempts have been made and no further information has been provided.

    Patient problems: Arthrofibrosis

    MAUDE report 50192562 ↗
  • InjuryAug 6, 2026Fracture; Peeled/Delaminated; Pitted

    It was reported that a patient has had ongoing stiffness, pain, and decreased range of motion. The patient is now twelve months post total knee replacement. An arthroscopic procedure was done, and the surgeon noted synovitis, poly pitting, edge cracking and delamination. The articular surface has not been removed. There is no additional information available.

    Patient problems: Joint effusion; Loss of Range of Motion; Pain; Swelling/ Edema; Synovitis

    MAUDE report 50157154 ↗
  • MalfunctionJun 12, 2026Positioning Failure

    IT WAS REPORTED THAT DURING AN INITIAL TOTAL KNEE ARTHROPLASTY, THE ARTICULAR SURFACE WAS UNABLE TO FIT INTO THE TIBIAL TRAY. A SECOND IMPLANT WAS USED TO COMPLETE THE PROCEDURE WITHOUT FURTHER COMPLICATION. ALL AVAILABLE INFORMATION HAS BEEN PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25491702 ↗
  • InjuryMar 16, 2026Adverse Event Without Identified Device or Use Problem

    NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Loss of Range of Motion

    MAUDE report 24613411 ↗
  • InjuryDec 18, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Pain; Swelling/ Edema

    MAUDE report 23844009 ↗
  • InjuryDec 16, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Joint effusion; Pain; Synovitis

    MAUDE report 23817032 ↗
  • MalfunctionDec 14, 2025Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging

    IT WAS REPORTED THAT THE PACKAGING OF THE IMPLANT WAS FOUND TO BE SEVERELY DAMAGED UPON RECEIPT. DUE TO POTENTIAL BREACH OF STERILITY, THE CUSTOMER OPTED TO NOT USE THE DEVICE. THERE WAS NO PATIENT INVOLVEMENT AND NO ADVERSE EVENTS HAVE BEEN REPORTED AS A RESULT OF THE MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23799586 ↗
  • MalfunctionNov 11, 2025Product Quality Problem; Defective Device; Product Quality Problem; Defective Device

    IT WAS REPORTED THAT THE POLY WAS DAMAGED, AND ASSUMED MANUFACTURING DEFICIENCY BY THE REPORTER. SURGERY WAS COMPLETED WITH A SECOND DEVICE. THERE WAS NO IMPACT TO THE PATIENT. ATTEMPTS HAVE BEEN MADE AND ALL AVAILABLE INFORMATION HAS BEEN PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23523296 ↗
  • MalfunctionOct 15, 2025Difficult to Insert; Difficult to Insert

    NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23280451 ↗
  • MalfunctionSep 8, 2025Difficult to Insert; Difficult to Insert

    IT WAS REPORTED THAT AN ARTICULAR SURFACE WOULD NOT ASSEMBLE WITH A TIBIAL TRAY. THE SURGEON ATTEMPTED TO PLACE THE MC BEARING ON THE BASE PLATE AND IT DIDN'T FIT WELL; ANOTHER ARTICULAR SURFACE WAS USED. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION HAVE BEEN MADE; HOWEVER, NO MORE INFORMATION IS AVAILABLE AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22991456 ↗
  • InjuryAug 14, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED RIGHT KNEE SWELLING AND LIMITED RANGE OF MOTION APPROXIMATELY ONE (1) MONTH FOLLOWING RIGHT KNEE ARTHROPLASTY. THE PATIENT WAS TREATED WITH A STEROID AND COMPRESSION SLEEVE TO ADDRESS INFLAMMATION. INITIAL OPERATIVE NOTES NOTED NO INTRAOPERATIVE COMPLICATIONS. NO ADDITIONAL PATIENT CONSEQUENCES WERE REPORTED.

    Patient problems: Loss of Range of Motion; Swelling/ Edema

    MAUDE report 22806297 ↗
  • InjuryJul 30, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Loss of Range of Motion

    MAUDE report 22668645 ↗
  • InjuryJul 24, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL KNEE SURGERY. SUBSEQUENTLY, THE PATIENT UNDERWENT THE FIRST STAGE OF A TWO STAGE REVISION DUE TO A MSSA INFECTION ON AN UNKNOWN DATE. DURING STAGE I, THE PATIENT WAS IMPLANTED WITH AN UNKNOWN CEMENT SPACER WHICH WAS REVISED DURING STAGE 2 TO A ROTATING HINGE CONSTRUCT. ATTEMPTS HAVE BEEN MADE AND ALL AVAILABLE INFORMATION HAS BEEN PROVIDED.

    Patient problems: Unspecified Infection

    MAUDE report 22588575 ↗
  • MalfunctionJun 2, 2025Difficult to Insert; Difficult to Insert

    IT WAS REPORTED DURING KNEE ARTHROPLASTY THAT THE ARTICULAR SURFACE COULD NOT BE SEATED ON THE TIBIAL TRAY. ANOTHER ARTICULAR SURFACE WAS USED TO COMPLETE THE PROCEDURE. NO ADVERSE EVENTS WERE REPORTED AS A RESULT OF THIS MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22123869 ↗
  • MalfunctionMay 22, 2025Difficult to Insert; Difficult to Insert

    IT WAS REPORTED THE SURGEON STRUGGLED TO GET THE BEARING SEATED ONTO THE TIBIAL TRAY. AFTER MANY ATTEMPTS HE TRIED A DIFFERENT BEARING AND IT FITTED WITHOUT A PROBLEM. NO PATIENT HARM OR IMPACT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22073176 ↗
  • InjuryDec 30, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL RIGHT TOTAL KNEE ARTHROPLASTY. APPROXIMATELY 16 DAYS POST-OP, THE PATIENT WAS PRESCRIBED AN ANTIBIOTIC FOR MILD REDNESS TO PROXIMAL INCISION AS A PROPHYLAXIS TREATMENT AND IT RESOLVED WITH NO FURTHER ALLEGATIONS. ATTEMPTS HAVE BEEN MADE AND ALL AVAILABLE INFORMATION HAS BEEN PROVIDED.

    Patient problems: Erythema; Post Operative Wound Infection

    MAUDE report 21036277 ↗
  • InjuryNov 18, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Joint Laxity

    MAUDE report 20716180 ↗
  • MalfunctionJun 11, 2024Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping

    IT WAS REPORTED THAT DURING A PROCEDURE, A HAIR WAS FOUND IN THE STERILE PACKAGING. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. NO PATIENT CONSEQUENCES WERE REPORTED AS A RESULT OF THE EVENT. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19510019 ↗
  • MalfunctionApr 3, 2024Positioning Failure; Material Integrity Problem; Positioning Failure; Material Integrity Problem

    IT WAS REPORTED THAT ON AN MC POLY, THERE WAS A PIECE OF PLASTIC THAT SEEMED OUT OF PLACE ON THE UNDERSIDE OF THE POLY. IT CAUSED THE POLY TO NOT LOCK INTO THE TIBIA. THE ISSUE WAS DISCOVERED AFTER IMPLANTING THE DEVICE INITIALLY. THE SURGEON HAD TO USE A DIFFERENT POLY WHICH TIGHTENED UP THE KNEE MORE THAN THEY WANTED TO INITIALLY. ATTEMPTS HAVE BEEN MADE AND ALL AVAILABLE INFORMATION HAS BEEN PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19030732 ↗
  • MalfunctionDec 20, 2023Difficult to Insert; Difficult to Insert

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18368138 ↗
  • InjuryJul 24, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED VIA STUDY A PATIENT UNDERWENT A RIGHT KNEE ARTHROPLASTY. SUBSEQUENTLY, THE PATIENT DEVELOPED PAIN, SWELLING, DECREASED RANGE OF MOTION, DIFFICULTY WITH MOBILITY, AND A BAKER¿S CYST ALSO FORMED. SUBSEQUENTLY, APPROXIMATELY FIVE MONTHS LATER, THE PATIENT UNDERWENT AN ARTHROSCOPIC PROCEDURE FOR LYSIS OF SCAR TISSUE AND ADHESIONS AS WELL AS DRAINAGE OF THE CYST. AT THE ONE YEAR CHECK-UP ALL THE KNEE COMPONENTS REMAIN IN PLACE AND THE PATIENT IS REPORTEDLY NOW DOING WELL WITH NO FURTHER COMPLICATIONS KNOWN.

    Patient problems: Adhesion(s); Ambulation or Postural Difficulties; Cyst(s); Joint effusion; Loss of Range of Motion; Pain; Scar Tissue

    MAUDE report 17383685 ↗
  • InjuryOct 5, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED A PATIENT UNDERWENT A RIGHT TOTAL KNEE ARTHROPLASTY. APPROXIMATELY 4 MONTHS POST IMPLANTATION, THE PATIENT UNDERWENT MANIPULATION UNDER ANESTHESIA DUE TO PAIN, STIFFNESS, AND LIMITED FLEXION. AFTER THE SURGERY, THE PATIENT CONTINUED TO IMPROVE AND HAS GAINED FULL RANGE OF MOTION. ALL IMPLANTS REMAIN IN PLACE.

    Patient problems: Loss of Range of Motion; Pain

    MAUDE report 15543645 ↗

Other Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer + additive/metal/polymer + additive devices

All Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer + additive/metal/polymer + additive devices →