Skip to content

PERSONA® Vivacit-E® 42-5121-007-14Recalls & reported problems

Zimmer, Inc. · Class 2 · Implant

All models of PERSONA Vivacit-E (285)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00889024469259), received between Jan 5, 2022 and Mar 11, 2024 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Difficult to Insert” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Insert1
  • Patient Device Interaction Problem1
  • Positioning Failure1

Patient problems

  • Joint effusion1
  • Loss of Range of Motion1
  • No Clinical Signs, Symptoms or Conditions1
  • Pain1
  • Swelling/ Edema1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 11, 2024Patient Device Interaction Problem; Patient Device Interaction Problem

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A LEFT KNEE ARTHROPLASTY REVISION OF THE ARTICULAR SURFACE TO ADDRESS PAIN, SWELLING AND INSTABILITY THAT WAS INTERFERING WITH RANGE OF MOTION APPROXIMATELY ONE YEAR POST-OPERATIVELY. THE SURGEON REPORTED THAT THE TIBIAL ROTATION DID NOT MATCH THE CORRECT PLACEMENT FOR A MEDIAL CONGRUENT BEARING AS THE SOFT TISSUES WERE TOO TIGHT TO ALLOW COMPONENT ARTICULATION. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION HAVE BEEN MADE; HOWEVER, NO MORE IS AVAILABLE AT THIS TIME.

    Patient problems: Joint effusion; Loss of Range of Motion; Pain; Swelling/ Edema

    MAUDE report 18882009 ↗
  • MalfunctionJan 5, 2022Positioning Failure; Difficult to Insert; Positioning Failure; Difficult to Insert

    IT WAS REPORTED THAT AN ARTICULAR SURFACE WAS NOT SEATING TO THE TIBIAL TRAY. THERE IS NO ADDITIONAL INFORMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13172127 ↗

Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices

All PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page