Skip to content

PERSONA® Vivacit-E® 42-5121-009-14Recalls & reported problems

Zimmer, Inc. · Class 2 · Implant

All models of PERSONA Vivacit-E (285)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00889024469419), received between Apr 22, 2024 and Aug 24, 2026 — classified as 3 injuries, 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Detachment of Device or Device Component1
  • Device Contaminated During Manufacture or Shipping1
  • Failure to Osseointegrate1
  • Naturally Worn1

Patient problems

  • Pain2
  • Aseptic loosening1
  • Joint effusion1
  • No Clinical Signs, Symptoms or Conditions1
  • Synovitis1
  • Thrombosis/Thrombus1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 24, 2026Device Contaminated During Manufacture or Shipping

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50415823 ↗
  • InjuryMar 20, 2026Failure to Osseointegrate; Detachment of Device or Device Component

    IT WAS REPORTED PATIENT UNDERWENT A REVISION PROCEDURE APPROXIMATELY 8 MONTHS POST-IMPLANTATION DUE TO TIBIAL COMPONENT LOOSENING AND PAIN. POSTOPERATIVELY, THE PATIENT DEVELOPED KNEE PAIN, AND AT FOLLOW-UP VISITS THE SURGEON IDENTIFIED PROGRESSIVE TIBIAL COMPONENT LOOSENING ON RADIOGRAPHS WITH MINIMAL TO NO BONE INGROWTH AT THE BONE¿IMPLANT INTERFACE. INTRA-OPERATIVELY DURING THE REVISION, THE TIBIAL COMPONENT WAS CONFIRMED TO BE LOOSE WITH MINIMAL BONE INGROWTH, WAS REMOVED, AND REPLACED WITH A CEMENTED TIBIAL COMPONENT WITH A STEM AND A POLYETHYLENE INSERT. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION HAVE BEEN MADE; HOWEVER, NO MORE IS AVAILABLE AT THIS TIME.

    Patient problems: Aseptic loosening; Pain

    MAUDE report 24647524 ↗
  • InjuryMay 7, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED A PATIENT HAD AN INITIAL LEFT TOTAL KNEE ARTHROPLASTY FOLLOWED BY AN I&D WITH POLY EXCHANGE ONE MONTH POST PROCEDURE. SUBSEQUENTLY, WITHIN A ONE MONTH POST-OPERATIVE TIMEFRAME, THE PATIENT DEVELOPED A DEEP VEIN THROMBUS AND WAS PLACED ON ELIQUIS. DUE TO A REACTION TO ELIQUIS, AN IVC FILTER WAS THEN PLACED AND THE PATIENT WAS REFERRED TO A HEMATOLOGIST FOR CONCERN FOR A BLOOD CLOTTING DISORDER

    Patient problems: Thrombosis/Thrombus

    MAUDE report 19260917 ↗
  • InjuryApr 22, 2024Naturally Worn; Naturally Worn

    IT WAS REPORTED A PATIENT HAD AN INITIAL LEFT TOTAL KNEE ARTHROPLASTY FOLLOWED BY A I&D WITH POLY EXCHANGE ABOUT ONE MONTH POST IMPLANTATION FOR INFECTION. SUBSEQUENTLY, THE PATIENT BEGAN TO EXPERIENCE PERSISTENT PAIN AND SWELLING AND UPON ASSESSMENT WAS DIAGNOSED WITH MECHANICAL LOOSENING. TWO YEARS, FIVE MONTHS AFTER THE I&D REVISION THE PATIENT UNDERWENT A REVISION DUE TO ASEPTIC FAILURE. DURING THE REVISION, ENCOUNTERED SYNOVITIS, PITTING TO THE POLY FROM THIRD BODY WEAR, AND DEFECTS NOTED ON THE FEMORAL AND TIBIAL COMPONENTS. ALL COMPONENTS WERE REVISED WITHOUT COMPLICATION.

    Patient problems: Joint effusion; Pain; Synovitis

    MAUDE report 19159425 ↗

Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices

All PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices →

Sources and updates