Persona® Trabecular Metal® 42-5022-066-01Recalls & reported problems
Zimmer, Inc. · Class 2 · Implant
All models of Persona Trabecular Metal (57)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00889024470118), received between Feb 6, 2023 and Dec 30, 2024 — classified as 3 injuries.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem3
- Insufficient Device Problem Information1
Patient problems
- Unspecified Infection2
- Discomfort1
- Failure of Implant1
- Inadequate Osseointegration1
- Insufficient Information1
- Loss of Range of Motion1
- Pain1
- Scar Tissue1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryDec 30, 2024Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
IT WAS REPORTED THAT A PATIENT UNDERWENT AN INITIAL TOTAL KNEE ARTHROPLASTY. SUBSEQUENTLY, ONE (1) YEAR POST-IMPLANTATION, THE PATIENT UNDERWENT AN ADDITIONAL SURGERY TO RESURFACE THE NATIVE PATELLA DUE TO INFECTION. A PATELLAR COMPONENT WAS IMPLANTED AND ALL INITIAL DEVICES REMAINED INTACT.
Patient problems: Insufficient Information; Unspecified Infection
MAUDE report 21036821 ↗InjuryMay 3, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED A PATIENT UNDERWENT A LEFT KNEE ARTHROPLASTY. AT 3 MONTHS POST OP, THE PATIENT WAS PAIN FREE WITH 120 DEGREES FLEXION. APPROXIMATELY 4 MONTHS POST IMPLANTATION, THE PATIENT PRESENTED WITH KNEE PAIN. THE ORIGINAL REASON FOR REVISION WAS FOR SUSPECTED INFECTION, HOWEVER, THE SURGEON MENTIONED THE KNEE DID NOT LOOK INFECTED WHEN REMOVING THE PROSTHESIS. THE SURGEON TOOK 6 SWABS, 3 ON DISTAL FEMUR AND 3 ON PROXIMAL TIBIA. WHEN REMOVING THE TIBIA, IT WAS NOTED THAT THE MEDIAL TM PEG HAD SNAPPED PRIOR TO ATTEMPTS TO REMOVE. THE BONE INGROWTH ON THE UNDERSIDE OF TM TIBIA WAS SMALL. THE PEG THAT WAS LEFT IN PATIENT HAD BONE INGROWTH. IT WAS NOTED THAT THE MEDIAL TM PEG IS NOT PERPENDICULAR TO THE BASEPLATE IN X-RAYS. ADDITIONALLY, METAL DEBRIS SEEN IN X-RAYS. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: Failure of Implant; Inadequate Osseointegration; Pain; Unspecified Infection
MAUDE report 16860494 ↗InjuryFeb 6, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED INITIAL RIGHT TOTAL KNEE ARTHROPLASTY. APPROXIMATELY 2.5 MONTHS IMPLANTATION, THE PATIENT UNDERWENT A MANIPULATION UNDER ANESTHESIA DUE TO STIFFNESS. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: Discomfort; Loss of Range of Motion; Scar Tissue
MAUDE report 16308612 ↗
Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page