PERSONA® Natural Tibia® 42-5320-075-01Recalls & reported problems
Zimmer, Inc. · Class 2 · Implant
All models of PERSONA Natural Tibia (25)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Zimmer Persona Natural Tibia knee implant is a cemented partial knee replacement. The 34 reports mostly describe the implant loosening over time and the pain, swelling, stiffness, and instability that followed, with several patients needing revision surgery; 33 reports involve injuries and no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 34
- Deaths reported
- 0
- Injuries reported
- 33
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 34 adverse event reports that list this exact device (DI 00889024471078), received between Mar 13, 2020 and Jul 31, 2026 — classified as 33 injuries, 1 malfunction.
The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (13 reports).
40 other recalls cover devices cleared under the same FDA submission (K113369, K122765, K172524); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem15
- Loosening of Implant Not Related to Bone-Ingrowth13
- Migration3
- Fracture2
- Biocompatibility1
- Loss of or Failure to Bond1
- Manufacturing, Packaging or Shipping Problem1
- Mechanical Jam1
Patient problems
- Failure of Implant13
- Pain13
- Swelling/ Edema6
- Loss of Range of Motion5
- Joint Laxity4
- Scar Tissue4
- Insufficient Information3
- Unspecified Infection3
Recent adverse event reports
The 25 most recent of 34 reports, in the reporter's own words as published by the FDA.
InjuryJul 31, 2026Migration
It was reported that the patient underwent a knee arthroplasty revision to address tibial tray migration resulting in instability. Upon removal of the components, femoral osteolysis around the metaphyseal region of the femur was identified that extended to the posterior condyles both medially and laterally. All components were revised and replaced. Attempts have been made, however, no additional information is available.
Patient problems: Joint Laxity; Osteolysis
MAUDE report 50100361 ↗InjuryNov 13, 2025Biocompatibility; Biocompatibility
IT WAS REPORTED THAT A PATIENT IS EXPERIENCING SWELLING AND POSSIBLE ALLERGIC REACTION APPROXIMATELY TWO YEARS SEVEN MONTHS POST IMPLANTATION. THE PATIENT CALLED STATING HE NEEDS TO KNOW IF HIS IMPLANT CONTAINS NICKEL AS HE HAS BEEN FOUND TO BE ALLERGIC TO NICKEL. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION HAVE BEEN MADE; HOWEVER, NO MORE INFORMATION IS AVAILABLE.
Patient problems: Hypersensitivity/Allergic reaction; Swelling/ Edema
MAUDE report 23544405 ↗InjuryJul 30, 2025Adverse Event Without Identified Device or Use Problem; Noise, Audible; Adverse Event Without Identified Device or Use Problem; Noise, Audible
IT WAS REPORTED A PATIENT HAD AN INITIAL TOTAL KNEE ARTHROPLASTY. SUBSEQUENTLY, THE PATIENT BEGAN EXPERIENCING PAIN, ONGOING SWELLING, JOINT EFFUSIONS, PERSISTENT CLICKING, AND GIVING AWAY. RADIOGRAPHIC IMAGING, SERUM INFECTION MARKERS, AND AN ASPIRATION WERE PERFORMED 8 MONTHS POST IMPLANTATION, AND THE TESTING RENDERED NO SIGNIFICANT FINDINGS. JOINT LAXITY MEDIALLY AND LATERALLY FOUND UPON ASSESSMENT. SUBSEQUENTLY, ONE YEAR POST-IMPLANTATION, THE PATIENT UNDERWENT A REVISION PROCEDURE. THE ARTICULAR SURFACE WAS EXCHANGED FOR A LARGER SIZE WITHOUT COMPLICATION AND THE PATIENT'S STABILITY ISSUES DURING RANGE OF MOTION HAVE BEEN REPORTED AS RESOLVED. THE FEMORAL AND TIBIAL COMPONENTS REMAIN IMPLANTED.
Patient problems: Joint Laxity; Joint effusion; Pain; Swelling/ Edema
MAUDE report 22669005 ↗InjuryJan 30, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED PATIENT WAS REVISED ONE YEAR, NINE MONTHS POST PROCEDURE DUE TO PAIN AND STIFFNESS (ASEPTIC). FEMORAL, TIBIAL AND BEARING WERE REVISED. THE PATELLA WAS NOT REVISED.
Patient problems: Pain
MAUDE report 21277400 ↗InjuryNov 12, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL LEFT TOTAL KNEE ARTHROPLASTY. SUBSEQUENTLY, 7 DAYS POST IMPLANTATION, THE PATIENT BEGAN TO EXPERIENCE CHEST TIGHTNESS. THE PATIENT WAS ULTIMATELY DIAGNOSED WITH MYOCARDIAL INFARCTION REQUIRING PLACEMENT OF 2 STENTS APPROXIMATELY 15 DAYS POST IMPLANTATION. THE EVENT HAS NOW BEEN DOCUMENTED WITHIN THE STUDY AS RESOLVED, NO DEVICE DEFICIENCIES OR ABNORMALITIES HAVE BEEN REPORTED AT THE 3-MONTH FOLLOW-UP AND ALL INITIAL COMPONENTS REMAIN IMPLANTED.
Patient problems: Myocardial Infarction
MAUDE report 20664588 ↗InjuryOct 29, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Loss of Range of Motion
MAUDE report 20560397 ↗InjuryOct 24, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THE PATIENT UNDERWENT A LEFT KNEE REVISION APPROXIMATELY 10 MONTHS POST IMPLANTATION DUE TO PAIN, SWELLING, AND INSTABILITY. DURING THE REVISION, SCAR TISSUE WAS DEBRIDED AND A MANIPULATION UNDER ANESTHESIA WAS PERFORMED, AND THE COMPONENTS WERE NOTED TO BE WELL FIXED. ALL IMPLANTS WERE REPLACED WITHOUT COMPLICATIONS.
Patient problems: Joint Laxity; Loss of Range of Motion; Pain; Scar Tissue; Swelling/ Edema
MAUDE report 20521526 ↗InjuryOct 23, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
IT WAS REPORTED THAT A PATIENT UNDERWENT AN INITIAL TOTAL KNEE ARTHROPLASTY. APPROXIMATELY TWO (2) YEARS AND SIX (6) WEEKS POST-IMPLANTATION, THE PATIENT UNDERWENT REVISION SURGERY DUE TO LOOSENING OF THE TIBIAL COMPONENT. THE LOOSENED DEVICE WAS REPLACED WITHOUT COMPLICATION. IT WAS REPORTED THAT NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Failure of Implant
MAUDE report 20517384 ↗InjuryJul 17, 2024Loss of or Failure to Bond; Loosening of Implant Not Related to Bone-Ingrowth; Loss of or Failure to Bond; Loosening of Implant Not Related to Bone-Ingrowth
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Failure of Implant
MAUDE report 19761745 ↗InjuryJun 21, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Failure of Implant; Joint Laxity
MAUDE report 19583494 ↗InjuryJun 21, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Synovitis
MAUDE report 19583166 ↗InjuryJun 12, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Failure of Implant; Pain
MAUDE report 19518848 ↗InjuryApr 29, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT WAS REVISED APPROXIMATELY ONE YEAR THREE WEEKS POST IMPLANTATION DUE TO PAIN. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION HAVE BEEN MADE; HOWEVER, NO MORE INFORMATION IS AVAILABLE.
Patient problems: Pain
MAUDE report 19208490 ↗InjuryApr 3, 2024Fracture; Fracture
IT WAS REPORTED, THAT THE PATIENT UNDERWENT A LEFT KNEE REVISION. APPROXIMATELY 5.5 YEARS POST IMPLANTATION, DUE TO A FRACTURED TIBIAL PLATE. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: Failure of Implant; Insufficient Information; Metal Related Pathology; Pain; Scar Tissue; Swelling/ Edema
MAUDE report 19038176 ↗InjuryMar 11, 2024Mechanical Jam; Migration; Mechanical Jam; Migration
IT WAS REPORTED THAT THE PATIENT UNDERWENT A LEFT KNEE ARTHROPLASTY REVISION OF THE ARTICULAR SURFACE TO ADDRESS PAIN, SWELLING AND INSTABILITY THAT WAS INTERFERING WITH RANGE OF MOTION APPROXIMATELY ONE YEAR POST-OPERATIVELY. THE SURGEON REPORTED THAT THE TIBIAL ROTATION DID NOT MATCH THE CORRECT PLACEMENT FOR A MEDIAL CONGRUENT BEARING AS THE SOFT TISSUES WERE TOO TIGHT TO ALLOW COMPONENT ARTICULATION. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION HAVE BEEN MADE; HOWEVER, NO MORE IS AVAILABLE AT THIS TIME.
Patient problems: Joint effusion; Loss of Range of Motion; Pain; Swelling/ Edema
MAUDE report 18882008 ↗InjuryJan 16, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Unspecified Infection
MAUDE report 18519980 ↗InjuryDec 18, 2023Use of Device Problem; Manufacturing, Packaging or Shipping Problem; Use of Device Problem; Manufacturing, Packaging or Shipping Problem
NO ADDITIONAL INFORMATION TO REPORT.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 18344396 ↗MalfunctionNov 13, 2023Fracture; Fracture
IT WAS REPORTED THAT DURING A KNEE PROCEDURE, AS THE SURGEON PLACED THE TIBIAL INSERT INTO THE TIBIAL TRAY USING AN INSERTER, THE TIBIAL TRAY CRACKED AND SMALL CHIP BROKE FROM THE TRAY. THE TIBIAL TRAY AND INSERT WERE REMOVED AND ADDITIONAL IMPLANTS WERE USED TO COMPLETE THE PROCEDURE. NO ADVERSE EVENTS HAVE BEEN REPORTED AS A RESULT OF THE MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18124248 ↗InjurySep 13, 2023Loosening of Implant Not Related to Bone-Ingrowth; Migration; Loosening of Implant Not Related to Bone-Ingrowth; Migration
IT WAS REPORTED THE PATIENT A LEFT KNEE REVISION APPROXIMATELY TWO AND A HALF YEARS POST-IMPLANTATION DUE TO PAIN AND TIBIAL SUBSIDENCE. DURING REMOVAL OF THE TIBIAL TRAY, IT WAS NOTED THAT THE CEMENT REMAINED ATTACHED TO THE MEDIAL SIDE OF THE TIBIA (WHICH HAD SUBSIDED) AND CAME DETACHED FROM THE LATERAL SIDE. NO ADDITIONAL PATIENT CONSEQUENCES WERE REPORTED.
Patient problems: Failure of Implant; Pain; Scar Tissue
MAUDE report 17741195 ↗InjuryAug 7, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED PATIENT IS EXPERIENCING PAIN IN KNEE AND UNABLE TO STAND OR WALK FOR JUST A FEW MINUTES THREE YEARS POST IMPLANTATION. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION HAVE BEEN MADE; HOWEVER, NO MORE IS AVAILABLE.
Patient problems: Ambulation or Postural Difficulties; Pain
MAUDE report 17469711 ↗InjuryJun 28, 2023Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Failure of Implant
MAUDE report 17216235 ↗InjuryJun 13, 2023Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Failure of Implant; Unspecified Infection
MAUDE report 17119272 ↗InjuryJan 10, 2023Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
IT WAS REPORTED A PATIENT UNDERWENT A LEFT KNEE REVISION APPROXIMATELY NINE MONTHS POST-IMPLANTATION DUE TO STIFFNESS AND TIBIAL LOOSENING.
Patient problems: Failure of Implant
MAUDE report 16133906 ↗InjuryDec 9, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED BY PATIENT¿S LEGAL COUNSEL THAT THE PATIENT UNDERWENT A KNEE ARTHROPLASTY. LEGAL COUNSEL IS REPORTING PATIENT HAS FILED AN UNKNOWN ALLEGATION. PATIENT HAS NOT BEEN REVISED TO DATE. NO ADDITIONAL INFORMATION AVAILABLE.
Patient problems: Insufficient Information
MAUDE report 15950471 ↗InjuryJul 7, 2022Adverse Event Without Identified Device or Use Problem; Loosening of Implant Not Related to Bone-Ingrowth; Adverse Event Without Identified Device or Use Problem; Loosening of Implant Not Related to Bone-Ingrowth
IT WAS REPORTED IN A CLINICAL STUDY THAT THE PATIENT UNDERWENT A LEFT KNEE PROCEDURE. SUBSEQUENTLY, 2 YEARS 6 MONTHS POST PROCEDURE, THE PATIENT HAS PAIN IN THE LEFT KNEE. ATTEMPTS TO GATHER ADDITIONAL INFORMATION ARE UNDERWAY. HOWEVER, NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.
Patient problems: Failure of Implant; Pain; Unspecified Infection
MAUDE report 14944992 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K113369, K122765, K172524. One submission often covers many models, so check the product and lot details in each record.
- Z-1224-2018 · Dec 12, 2017 · TerminatedPERSONA THE PERSONALIZED KNEE SYSTEM "PSN TIB STM 5 DEG SZ D R" indicated for patients with severe knee pain and disability due to: - Rheumatoid arthritis, osteoarthritis, traumati
- Z-1807-2018 · Aug 7, 2017 · TerminatedPersona Cemented Tibial SZ PLT Size H Left, Item Number 42539908301 For use during orthopaedic surgery for implantation of a prosthesis.
- Z-1806-2018 · Aug 7, 2017 · TerminatedPersona Stemmed Tibial Provisional Size H Right, Item Number 42532108302 For use during orthopaedic surgery for implantation of a prosthesis.
- Z-1805-2018 · Aug 7, 2017 · TerminatedPersona Stemmed Tibial Provisional Size H Left, Item Number 42532108301 For use during orthopaedic surgery for implantation of a prosthesis.
- Z-1808-2018 · Aug 7, 2017 · TerminatedPersona Cemented Tibial SZ PLT Size H Right, Item Number 42539908302 For use during orthopaedic surgery for implantation of a prosthesis.
- Z-1341-2016 · Mar 17, 2016 · TerminatedPersona The Personalized Knee System Articular Surface Posterior Stabilized (PS) Left Height 18mm Sterile For use in total knee arthroplasty. REF 42-5114-008-18
- Z-1340-2016 · Mar 17, 2016 · TerminatedPersona The Personalized Knee System Articular Surface Posterior Stabilized (PS) Left Height 14mm Sterile. For use in total knee arthroplasty. REF 42-5114-005-14
- Z-1157-2016 · Jan 11, 2016 · TerminatedProduct 59 consists of all product under product code: JWH and same usage: Item no: 42540000026 PSN ALL POLY PAT PLY 26MM 42540000029 PSN ALL POLY PAT PLY 29MM 42540000032 PSN ALL
- Z-1360-2015 · Mar 10, 2015 · TerminatedBrand Name: Persona Anterior Referencing Sizer with Locking Boom The Persona Anterior Referencing Sizer is a sterilizable instrument intended for use in multiple total knee arthrop
- Z-0853-2015 · Dec 4, 2014 · TerminatedPersona Cemented Tibial Broach Inserter/Extractor. The Persona Cemented Tibial Broach Inserter/Extractor Handle is a Use sterilizable instrument intended for use In multiple TKA p
- Z-1042-2015 · Nov 25, 2014 · TerminatedZimmer PERSONA Tibial Articular Surface Inserter; Item # 42-5299-001-00. Manual surgical instrument for knee arthroplasty (orthopedic).
- Z-0198-2015 · Oct 9, 2014 · TerminatedPersona Stemmed 5 Degree Cemented Tibia Product Usage: This device is indicated for patients with severe knee pain and disability due to: Rheumatoid arthritis, osteoarthritis, t
- Z-0123-2015 · Oct 7, 2014 · TerminatedThe Persona Tibial Articular Surface Inserter is a sterilizable instrument intended for use in multiple total knee arthroplasty procedures. Per the Persona Surgical technique, the
- Z-2576-2014 · Aug 7, 2014 · TerminatedPERSONA The Personalized Knee System Constrained Posterior Stabilized Tibial Articular Surface Provisional TOP with the following: L 3-5 CD; L 6-9 CD; L 3-5 EF; L 6-9 EF; L 10-11 E
- Z-2579-2014 · Aug 7, 2014 · TerminatedPERSONA The Personalized Knee System PS Tibial Articular Surface Provisional Left with Size 3-5 CD Top; Size 6-9 CD Top; Size 3-5 EF Top; Size 6-9 EF Top; Size 10-11 EF Top; Size 6
…and 25 more.
Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page