Skip to content

Wagner SL Revision® 01.00101.916 (UDI 00889024593473)Recalls & reported problems

Zimmer GmbH · Class 2 · Implant

All models of Wagner SL Revision (33)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00889024593473), received between Aug 26, 2026 and Aug 26, 2026 — classified as 1 injury.

The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Loosening of Implant Not Related to Bone-Ingrowth1

Patient problems

  • Aseptic loosening1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 26, 2026Loosening of Implant Not Related to Bone-Ingrowth

    It was reported from a clinical study that a patient underwent a right hip stage I revision arthroplasty with final implants used as a spacer for infection. Subsequently, a stage II revision arthroplasty was performed; during which, the gross aseptic loosening was noted in the cup and stem. All components were revised without complication. The patient entered the current study at the stage II revision with no further concerns or complications reported.Attempts have been made and no further information has been provided.

    Patient problems: Aseptic loosening

    MAUDE report 50453557 ↗

Other Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented devices

All Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page