Skip to content

NexGen® Complete Knee Solution 5962-40-10Recalls & reported problems

Zimmer, Inc. · Class 2 · Implant

All models of NexGen Complete Knee Solution (10)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00889024666092), received between Mar 4, 2025 and Nov 25, 2025 — classified as 1 injury, 6 malfunctions.

The most frequently reported specific device problem is “Difficult to Insert” (3 reports).

4 other recalls cover devices cleared under the same FDA submission (K042271, K173057); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 3 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Insert3
  • Positioning Failure3
  • Fracture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • Insufficient Information1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 25, 2025Fracture; Fracture

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Insufficient Information

    MAUDE report 23644765 ↗
  • MalfunctionNov 5, 2025Positioning Failure; Positioning Failure

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23479250 ↗
  • MalfunctionJun 3, 2025Positioning Failure; Positioning Failure

    IT WAS REPORTED THAT DURING AN INITIAL SURGERY, THE LINER WOULD NOT ASSEMBLE WITH THE TIBIAL TRAY RESULTING IN A 1-HOUR DELAY WAITING FOR ANOTHER LINER. THERE WAS NO CONSEQUENCES OR IMPACT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22135532 ↗
  • MalfunctionMay 16, 2025Positioning Failure; Positioning Failure

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22036397 ↗
  • MalfunctionMay 16, 2025Difficult to Insert; Difficult to Insert

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22034893 ↗
  • MalfunctionMay 6, 2025Difficult to Insert; Difficult to Insert

    IT WAS REPORTED DURING KNEE ARTHROPLASTY THAT THE ARTICULAR SURFACE COULD NOT BE SEATED ON THE TIBIAL TRAY. ANOTHER ARTICULAR SURFACE WAS USED TO COMPLETE THE PROCEDURE. NO ADVERSE EVENTS WERE REPORTED AS A RESULT OF THIS MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21970764 ↗
  • MalfunctionMar 4, 2025Difficult to Insert; Difficult to Insert

    IT IS REPORTED THAT THE ARTICULAR SURFACE FAILED TO MATCH THE PLATFORM PROPERLY. A NEW ARTICULAR SURFACE WAS USED TO COMPLETE THE PROCEDURE. NO HARM TO THE PATIENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21517692 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K042271, K173057. One submission often covers many models, so check the product and lot details in each record.

  • Z-1433-2025 · Feb 27, 2025 · Open, Classified
    NexGen LPS Flex, Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis, Model Number 00-5962-042-10
  • Z-2092-2023 · May 19, 2023 · Terminated
    NexGen Precoat Stemmed Tibial Plate Size 5
  • Z-0321-2020 · Oct 10, 2019 · Terminated
    Various Knee Systems for knee joint arthroplasty, knee prosthesis, Item Nos. 00522000100 00522000200 00522000300 00522000400 00522006304 00522006504 00522006600 00522006800 0052200
  • Z-0609-2011 · Apr 16, 2010 · Terminated
    NexGen Complete Knee Solution Stemmed tibial broach impactor, Catalog # 00-5977-011-00, Zimmer Inc., Warsaw, IN.

Other PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER devices

All PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page