NexGen® Complete Knee Solution 5962-40-12Recalls & reported problems
Zimmer, Inc. · Class 2 · Implant
All models of NexGen Complete Knee Solution (10)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00889024666108), received between Apr 17, 2025 and Mar 18, 2026 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Difficult to Insert” (2 reports).
4 other recalls cover devices cleared under the same FDA submission (K042271, K173057); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Difficult to Insert2
- Fracture2
- Noise, Audible1
- Unstable1
Patient problems
- Joint Laxity2
- No Clinical Signs, Symptoms or Conditions2
- Failure of Implant1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryMar 18, 2026Fracture; Unstable; Noise, Audible
IT WAS REPORTED THAT THE PATIENT UNDERWENT A KNEE ARTHROPLASTY REVISION TO ADDRESS INSTABILITY AND AUDIBLE NOISE FROM THE KNEE SECONDARY TO POLYETHYLENE IMPLANT FRACTURE APPROXIMATELY NINE (9) YEARS POST-OPERATIVELY. ATTEMPTS HAVE BEEN MADE, HOWEVER, NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Joint Laxity
MAUDE report 24629743 ↗MalfunctionJul 16, 2025Difficult to Insert; Difficult to Insert
IT WAS REPORTED THE LINER WOULD NOT ASSEMBLE WITH THE MATING TIBIAL TRAY. THERE WAS A 20 MINS DELAY. FINALLY SURGEON USED SIZE 10MM LINER TO COMPLETE THE SURGERY. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION HAVE BEEN MADE; HOWEVER, NO MORE IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22519983 ↗InjuryMay 12, 2025Fracture; Fracture
IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL LEFT KNEE SURGERY. APPROXIMATELY SEVEN YEARS POST-OPERATION, THE PATIENT STOOD UP AND FELT A POP AND EXPERIENCED INSTABILITY. SUBSEQUENTLY, THE PATIENT WAS REVISED AND DURING THE REVISION SURGERY, THE PS POST WAS FOUND BROKEN. A POLY SWAP WAS PERFORMED. ALL AVAILABLE INFORMATION HAS BEEN PROVIDED.
Patient problems: Failure of Implant; Joint Laxity
MAUDE report 22002836 ↗MalfunctionApr 17, 2025Difficult to Insert; Difficult to Insert
IT WAS REPORTED DURING KNEE ARTHROPLASTY THAT THE ARTICULAR SURFACE COULD NOT BE ASSEMBLED WITH THE TIBIAL TRAY AFTER MULTIPLE ATTEMPTS. ANOTHER ARTICULAR SURFACE WAS USED TO COMPLETE THE PROCEDURE. ATTEMPTS HAVE BEEN MADE, HOWEVER, NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21858434 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K042271, K173057. One submission often covers many models, so check the product and lot details in each record.
- Z-1433-2025 · Feb 27, 2025 · Open, ClassifiedNexGen LPS Flex, Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis, Model Number 00-5962-042-10
- Z-2092-2023 · May 19, 2023 · TerminatedNexGen Precoat Stemmed Tibial Plate Size 5
- Z-0321-2020 · Oct 10, 2019 · TerminatedVarious Knee Systems for knee joint arthroplasty, knee prosthesis, Item Nos. 00522000100 00522000200 00522000300 00522000400 00522006304 00522006504 00522006600 00522006800
- Z-0609-2011 · Apr 16, 2010 · TerminatedNexGen Complete Knee Solution Stemmed tibial broach impactor, Catalog # 00-5977-011-00, Zimmer Inc., Warsaw, IN.
Other PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER devices
- ATTUNE 1512-14-050Depuy (Ireland) · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER106 reports
- ATTUNE 1504-40-205Depuy (Ireland) · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 recall53 reports
- ATTUNE 1504-40-105Depuy (Ireland) · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 recall46 reports
- ATTUNE 1547-06-001Depuy (Ireland) · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 recall37 reports
- ATTUNE 1513-12-060Depuy (Ireland) · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER72 reports
- Triathlon 5521-B-300Howmedica Osteonics Corp. · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 recall19 reports
- Triathlon 5531-G-709-EHowmedica Osteonics Corp. · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 recall19 reports
- ATTUNE 1512-16-080Depuy (Ireland) · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 recall16 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page