1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00889024681965), received between Nov 4, 2025 and Jan 22, 2026 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Fracture” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

4 other recalls cover devices cleared under the same FDA submission (K233148, K241043); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 20, 2026Class IIOpen, Classified

    Five complaints have been received identifying issues where the plastic guides were catching during reaming, leading to breakage.

    Z-2527-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fracture1
  • Mechanical Jam1

Patient problems

  • Bone Fracture(s)1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 22, 2026Mechanical Jam

    IT WAS REPORTED THAT DURING SHOULDER SURGERY THE ANTERIOR PORTION OF THE GLENOID FRACTURED WHEN THE REAMER BLADE INTERACTED WITH A DISPOSABLE OFF-AXIS REAMER GUIDE. DURING INTRA-OPERATIVE REAMING, THE BLADE DUG INTO THE PLASTIC OF THE GUIDE, WHICH TORQUED THE GLENOID AND RESULTED IN AN ANTERIOR GLENOID FRACTURE. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: Bone Fracture(s)

    MAUDE report 24152002 ↗
  • MalfunctionNov 4, 2025Fracture; Fracture

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23469483 ↗

FDA clearance and classification

Cleared through Special 510(k) K241043 on May 15, 2024 (decision: Substantially Equivalent). Submitted by Fix Surgical as “Augment Off-Axis Instrument System”. The FDA review took 28 days. FDA record ↗

Cleared through 510(k) K233148 on Mar 22, 2024 (decision: Substantially Equivalent). Submitted by Fix Surgical as “Augment Off-Axis Instrument System”. The FDA review took 177 days. FDA record ↗

The FDA regulates product code KWT (prosthesis, shoulder, non-constrained, metal/polymer cemented) under 21 CFR 888.3650 as a Class II device. 21 CFR 888.3650 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K233148, K241043. One submission often covers many models, so check the product and lot details in each record.

  • Z-2530-2026 · May 20, 2026 · Open, Classified
    Off-Axis Alliance Glenoid, Left, 4-Peg, Augment, Reamer Guide, Model/Catalog Number: 110040444
  • Z-2531-2026 · May 20, 2026 · Open, Classified
    Off-Axis Alliance Glenoid, Right, 4-Peg, Augment, Reamer Guide, Model/Catalog Number: 110040476
  • Z-2529-2026 · May 20, 2026 · Open, Classified
    Off-Axis Comprehensive Shoulder System, Large, Augment, Off-Axis, Reamer Guide Model/Catalog Number: 110040242
  • Z-2528-2026 · May 20, 2026 · Open, Classified
    Off-Axis Comprehensive Shoulder System, Medium, Augment, Off-Axis, Reamer Guide Model/Catalog Number: 110040241

Common questions about the Comprehensive 110040240

Has the Comprehensive 110040240 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2527-2026 (Class II), began on May 20, 2026: Five complaints have been received identifying issues where the plastic guides were catching during reaming, leading to breakage. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the Comprehensive 110040240?

The FDA's MAUDE database lists 2 adverse event reports for this device, received between Nov 4, 2025 and Jan 22, 2026. In those reports, the most common specific device problems are “Fracture” (1) and “Mechanical Jam” (1); the most common patient problem is “Bone Fracture(s)” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Comprehensive 110040240?

None of the reports are classified as deaths, 1 as an injury and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Comprehensive 110040240 FDA approved?

It was cleared through 510(k) K233148 on Mar 22, 2024; its record lists 2 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the Comprehensive 110040240, and is it still sold?

The FDA's device database lists Zimmer, Inc. as the company responsible for it (brand Comprehensive, model 110040240). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented devices

All Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page