ALTERA 1124.1232Recalls & reported problems
Globus Medical, Inc. · ALTERA Spacer, 10 x 36, 9-13mm, 15° · Class 2 · Implant
All models of ALTERA (23)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
1 FDA recall record name this device's identifier.
21 other recalls cover devices cleared under the same FDA submission (K140411); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Recalls naming this device
- Jul 22, 2020Class IITerminated
Specific lots of ALTERA implants have internal components that may have been manufactured using stainless steel rather than the specified cobalt chromium molybdenum alloy.
Z-2869-2020 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
FDA clearance and classification
Cleared through 510(k) K140411 on Jul 1, 2014 (decision: Substantially Equivalent). Submitted by Globus Medical, Inc. as “Altera Spacer”. The FDA review took 133 days. FDA record ↗
The FDA regulates product code MAX (intervertebral fusion device with bone graft, lumbar) under 21 CFR 888.3080 as a Class II device. 21 CFR 888.3080 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
FDA definition of this device type: Intended to stabilize spinal segment to promote fusion in order to restrict motion and decrease pain using bone graft.
Recalls under the same FDA submission
These recalls involve devices cleared under K140411. One submission often covers many models, so check the product and lot details in each record.
- Z-2860-2020 · Jul 22, 2020 · TerminatedALTERA Spacer ALTERA Spacer, 10x31, 10-14mm, 8
- Z-2868-2020 · Jul 22, 2020 · TerminatedALTERA Spacer ALTERA Spacer, 10x36, 12-16mm, 8
- Z-2853-2020 · Jul 22, 2020 · TerminatedALTERA Spacer ALTERA Spacer, 10x26, 10-14mm, 8
- Z-2862-2020 · Jul 22, 2020 · TerminatedALTERA Spacer ALTERA Spacer, 10x31, 9-13mm, 15
- Z-2854-2020 · Jul 22, 2020 · TerminatedALTERA Spacer ALTERA Spacer, 10x26, 12-16mm, 8
- Z-2850-2020 · Jul 22, 2020 · TerminatedALTERA Spacer ALTERA Spacer, 10x31, 8-12mm, 8 (sterile)
- Z-2864-2020 · Jul 22, 2020 · TerminatedALTERA Spacer ALTERA Spacer, 10x31, 12-16mm, 15
- Z-2871-2020 · Jul 22, 2020 · TerminatedALTERA Spacer ALTERA Spacer, 10x36, 12-16mm, 15
- Z-2865-2020 · Jul 22, 2020 · TerminatedALTERA Spacer ALTERA Spacer, 10x36, 8-12mm, 8
- Z-2863-2020 · Jul 22, 2020 · TerminatedALTERA Spacer ALTERA Spacer, 10x31, 10-14mm, 15
- Z-2867-2020 · Jul 22, 2020 · TerminatedALTERA Spacer ALTERA Spacer, 10x36, 10-14mm, 8
- Z-2861-2020 · Jul 22, 2020 · TerminatedALTERA Spacer ALTERA Spacer, 10x31, 12-16mm, 8
- Z-2851-2020 · Jul 22, 2020 · TerminatedALTERA Spacer ALTERA Spacer, 10x26, 8-12mm, 8
- Z-2858-2020 · Jul 22, 2020 · TerminatedALTERA Spacer ALTERA Spacer, 10x31, 8-12mm, 8
- Z-2866-2020 · Jul 22, 2020 · TerminatedALTERA Spacer ALTERA Spacer, 10x36, 9-13mm, 8
…and 6 more.
Common questions about the ALTERA 1124.1232
Has the ALTERA 1124.1232 been recalled?
Yes. 1 FDA recall record names this device's identifier, and the FDA has closed (terminated) it. The recall, Z-2869-2020 (Class II), began on Jul 22, 2020: Specific lots of ALTERA implants have internal components that may have been manufactured using stainless steel rather than the specified cobalt chromium molybdenum alloy. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
Is the ALTERA 1124.1232 FDA approved?
It was cleared through 510(k) K140411 on Jul 1, 2014. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the ALTERA 1124.1232 MRI safe?
According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.
Who makes the ALTERA 1124.1232, and is it still sold?
The FDA's device database lists Globus Medical, Inc. as the company responsible for it (brand ALTERA, model 1124.1232). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Intervertebral fusion device with bone graft, lumbar devices
- T-PAL 03.812.003Synthes GmbH · Intervertebral fusion device with bone graft, lumbar152 reports
- T-PAL 03.812.004Synthes GmbH · Intervertebral fusion device with bone graft, lumbar120 reports
- Catalyft PL Expandable Interbody System 6068073Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall66 reports
- Catalyft PL Expandable Interbody System 6069076Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall51 reports
- Catalyft PL Expandable Interbody System 6068076Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall46 reports
- EIT TLIF Inserter TFT30101EIT Emerging Implant Technologies GmbH · Intervertebral Fusion Device With Bone Graft, Lumbar88 reports
- T-PAL 03.812.001Synthes GmbH · Intervertebral fusion device with bone graft, lumbar76 reports
- Catalyft PL Expandable Interbody System 6069096Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall25 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page