PHALINX 45A01004Recalls & reported problems
Wright Medical Technology, Inc. · Cannulated Straight · Class 2
All models of PHALINX (10)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00889797102162), received between Dec 4, 2023 and Dec 4, 2023 — classified as 2 injuries.
The most frequently reported specific device problem is “Mechanical Jam” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Mechanical Jam2
- Positioning Failure2
- Separation Failure2
- Unstable2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryDec 4, 2023Positioning Failure; Unstable; Separation Failure; Mechanical Jam; Positioning Failure; Unstable; Separation Failure; Mechanical Jam
THE PATIENT HAD HAMMERTOES ON 2ND & 3RD DIGITS THAT NEEDED FIXATION. IT WAS REPORTED THAT THE IMPLANTS DID NOT WANT TO COME OFF OF THE HANDLES AND THE DRILL BITS DID NOT WANT TO DETACH FROM THE HANDLES EITHER TO DRILL USING POWER. IT TOOK APPROXIMATELY 10-15 MINUTES LONGER DUE TO HAVING TO PRY THE IMPLANT AND DRILL FROM THEIR HANDLES. AND THE IMPLANT WAS PUT IN TO THE MIDDLE PHALANX THREE TIMES ON THE 2ND DIGIT AND IT PULLED OUT WITH THE HANDLE EACH TIME. IT WAS LOOSE AND SECURED WITH K WIRE. THE 3RD DIGIT WAS IMPLANTED WITHOUT THE USE OF THE HANDLE IN THE KIT AND WENT SMOOTHLY MINUS ISSUES REMOVING THE IMPLANT FROM THE HANDLE. BONE QUALITY IN PATIENT APPEARED TO BE GOOD AS HE HAD TO USE SOME EFFORT WHEN DRILLING USING POWER. WHEN THE IMPLANT WAS SCREWED INTO THE MIDDLE PHALANX AND FULLY SEATED HE FELT THAT THERE WAS A GOOD AMOUNT OF BITE, BUT WHEN THE HANDLE WAS PULLED BACK IT WOULD NOT RELEASE THE IMPLANT. THIS CAUSED THE IMPLANT TO PULL OUT OF THE MIDDLE PHALANX AND IT REMOVED BONE WITH IT. THIS WAS ATTEMPTED THREE TIMES BEFORE PULLING THE IMPLANT OUT OF THE HANDLE WHICH REQUIRED SOME FORCE AND THEN IT WAS INSERTED USING FORCEPS. DEVICE DID NOT APPEAR TO BE BROKEN AND NO OTHER C
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18263585 ↗InjuryDec 4, 2023Positioning Failure; Unstable; Separation Failure; Mechanical Jam; Positioning Failure; Unstable; Separation Failure; Mechanical Jam
THE PATIENT HAD HAMMERTOES ON 2ND AND 3RD DIGITS THAT NEEDED FIXATION. IT WAS REPORTED THAT THE IMPLANTS DID NOT WANT TO COME OFF OF THE HANDLES AND THE DRILL BITS DID NOT WANT TO DETACH FROM THE HANDLES EITHER TO DRILL USING POWER. IT TOOK APPROXIMATELY 10-15 MINUTES LONGER DUE TO HAVING TO PRY THE IMPLANT AND DRILL FROM THEIR HANDLES. AND THE IMPLANT WAS PUT IN TO THE MIDDLE PHALANX THREE TIMES ON THE 2ND DIGIT AND IT PULLED OUT WITH THE HANDLE EACH TIME. IT WAS LOOSE AND SECURED WITH K WIRE. THE 3RD DIGIT WAS IMPLANTED WITHOUT THE USE OF THE HANDLE IN THE KIT AND WENT SMOOTHLY MINUS ISSUES REMOVING THE IMPLANT FROM THE HANDLE. BONE QUALITY IN PATIENT APPEARED TO BE GOOD AS HE HAD TO USE SOME EFFORT WHEN DRILLING USING POWER. WHEN THE IMPLANT WAS SCREWED INTO THE MIDDLE PHALANX AND FULLY SEATED HE FELT THAT THERE WAS A GOOD AMOUNT OF BITE, BUT WHEN THE HANDLE WAS PULLED BACK IT WOULD NOT RELEASE THE IMPLANT. THIS CAUSED THE IMPLANT TO PULL OUT OF THE MIDDLE PHALANX AND IT REMOVED BONE WITH IT. THIS WAS ATTEMPTED THREE TIMES BEFORE PULLING THE IMPLANT OUT OF THE HANDLE WHICH REQUIRED SOME FORCE AND THEN IT WAS INSERTED USING FORCEPS. DEVICE DID NOT APPEAR TO BE BROKEN AND NO OTHER
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18263498 ↗
FDA clearance and classification
Cleared through 510(k) K212996 on Nov 16, 2021 (decision: Substantially Equivalent). Submitted by Wrightmedicaltechnologyinc as “Sterile PHALINX Hammertoe System”. The FDA review took 57 days. FDA record ↗
The FDA regulates product code HTY (pin, fixation, smooth) under 21 CFR 888.3040 as a Class II device. 21 CFR 888.3040 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the PHALINX 45A01004
Has the PHALINX 45A01004 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00889797102162), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the PHALINX 45A01004?
The FDA's MAUDE database lists 2 adverse event reports for this device, received on Dec 4, 2023. In those reports, the most common specific device problems are “Mechanical Jam” (2), “Positioning Failure” (2) and “Separation Failure” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the PHALINX 45A01004?
None of the reports are classified as deaths, 2 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the PHALINX 45A01004 FDA approved?
It was cleared through 510(k) K212996 on Nov 16, 2021. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the PHALINX 45A01004 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the PHALINX 45A01004, and is it still sold?
The FDA's device database lists Wright Medical Technology, Inc. as the company responsible for it (brand PHALINX, model 45A01004). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other PIN, FIXATION, SMOOTH devices
- RIGIDFIX 210133Medos International Sàrl · PIN, FIXATION, SMOOTH91 reports
- SMART TOE ST0A-16PStryker GmbH · PIN, FIXATION, SMOOTH1 recall5 reports
- SMART TOE ST0-19PStryker GmbH · PIN, FIXATION, SMOOTH1 recall2 reports
- SMART TOE ST0-22PStryker GmbH · PIN, FIXATION, SMOOTH1 recall0 reports
- SMART TOE ST0XS-11Stryker GmbH · PIN, FIXATION, SMOOTH1 recall0 reports
- ARROW-LOK 1Arrowhead De LLC · Pin, Fixation, Smooth1 recall0 reports
- ARROW-LOK 1Arrowhead De LLC · Pin, Fixation, Smooth1 recall0 reports
- ARROW-LOK 1Arrowhead De LLC · Pin, Fixation, Smooth1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page