Skip to content

Matrixx System 54M-30-12-SPRecalls & reported problems

Nexxt Spine, LLC · TLIF Oblique 30D x 10W x 12H · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00889929026229), received between Mar 18, 2019 and Mar 18, 2019 — classified as 1 malfunction.

The most frequently reported specific device problem is “Fracture” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Fracture1

Patient problems

  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 18, 2019Fracture; Fracture

    IT WAS REPORTED ON (B)(6) 2019 THAT A TLIF OBLIQUE IMPLANT FRACTURED DURING THE INSERT AND ROTATE TECHNIQUE. THE PROCEDURE INVOLVED THE IMPLANTATION OF TWO TLIF OBLIQUE CAGES. ONE WAS SUCCESSFUL. THE SURGEON POUNDED A SECOND 30DX10WX12H TLIF OBLIQUE IMPLANT INTO A 4MM DISC SPACE, INCREASING THE DISC SPACE TO 10MM TO ACCOMMODATE THE INTERBODY. WHEN ROTATING FROM 10MM TO 12MM ORIENTATION, THE INTERBODY BROKE. THE SURGEON USED EXCESSIVE FORCE TO ROTATE THE INTERBODY WHICH EXCEEDED THE DESIGN LIMITS FOR ROTATION. THE IMPLANT WAS INVESTIGATED TO BE SECURE, AND NO PIECES BROKE OFF OR FELL INTO THE WOUND, SO IT WAS LEFT IN PLACE TO PERFORM ITS INTENDED FUNCTION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8430343 ↗

Other Spinal Vertebral Body Replacement Device devices

All Spinal Vertebral Body Replacement Device devices →