Skip to content

S100 Sink Spout 70-0280Recalls & reported problems

Nephros, Inc. · The S100 Point of Use Filter is a disposable point-of-use micro-filter used to filter EPA quality drinking water. The filter retains bacteria. By retaining bacteria in water for washing and drinking, the filter may aid in infection control. The filter produces water that is suitable for cleaning of equipment used in medical procedures and washing of surgeon’s hands. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00896241002541), received between Aug 26, 2022 and Aug 26, 2022 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Break2
  • Filtration Problem2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 26, 2022Break; Filtration Problem; Break; Filtration Problem

    ON (B)(6) 2022 IT WAS REPORTED THAT A HEALTHCARE FACILITY RECEIVED A POSITIVE CULTURE DOWNSTREAM OF AN S100 POINT OF USE (70-0280) FILTER ON ONE OF THEIR SINKS. THERE WAS NO REPORTED ADVERSE EVENT. NEPHROS S100 POINT OF USE IS A SINK WATER FILTER, CLASS II MEDICAL DEVICE THAT IS PLACED ON SINK FAUCETS IN HEALTH CARE FACILITIES IN ORDER TO RETAIN BACTERIA. THE USER FACILITY ROUTINELY TESTS THEIR WATER AT THEIR SINKS AND FOUND THAT THERE WAS POSITIVE BACTERIA DETECTED. A POSITIVE CULTURE CAN BE CAUSED BY IMPROPER INSTALLATION, CROSS CONTAMINATION AND/OR A BREACH OF THE FILTER MEMBRANE. COMPREHENSIVE REVIEW OF THE SPECIFIC FILTER DEVICE HISTORY RECORD AND MANUFACTURING INSPECTIONS DID NOT IDENTIFY ANY PRODUCT ISSUES AND THE INVESTIGATION IS UNDERWAY TO DETERMINE THE ROOT CAUSE. ONCE THE INVESTIGATION IS COMPLETED, A SUPPLEMENTAL/FOLLOW-UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15307026 ↗
  • MalfunctionAug 26, 2022Break; Filtration Problem; Break; Filtration Problem

    ON (B)(6) 2022 IT WAS REPORTED THAT A HEALTHCARE FACILITY RECEIVED A POSITIVE CULTURE DOWNSTREAM OF AN S100 POINT OF USE (70-0280) FILTER ON ONE OF THEIR SINKS. THERE WAS NO REPORTED ADVERSE EVENT. NEPHROS S100 POINT OF USE IS A SINK WATER FILTER, CLASS II MEDICAL DEVICE THAT IS PLACED ON SINK FAUCETS IN HEALTH CARE FACILITIES IN ORDER TO RETAIN BACTERIA. THE USER FACILITY ROUTINELY TESTS THEIR WATER AT THEIR SINKS AND FOUND THAT THERE WAS POSITIVE BACTERIA DETECTED. A POSITIVE CULTURE CAN BE CAUSED BY IMPROPER INSTALLATION, CROSS CONTAMINATION AND/OR A BREACH OF THE FILTER MEMBRANE. COMPREHENSIVE REVIEW OF THE SPECIFIC FILTER DEVICE HISTORY RECORD AND MANUFACTURING INSPECTIONS DID NOT IDENTIFY ANY PRODUCT ISSUES AND THE INVESTIGATION IS UNDERWAY TO DETERMINE THE ROOT CAUSE. ONCE THE INVESTIGATION IS COMPLETED, A SUPPLEMENTAL/FOLLOW-UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15307007 ↗

Other System, Water Purification, General Medical Use devices

All System, Water Purification, General Medical Use devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page