Adherus NUS-106Recalls & reported problems
Stryker Corporation · Adherus AutoSpray Dural Sealant · Class 3 · Implant
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00897146002018), received between Aug 15, 2022 and May 5, 2026 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Physical Resistance/Sticking” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Physical Resistance/Sticking1
Patient problems
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 5, 2026
PT UNDERWENT L4-L5 LUMBAR DISCECTOMY FOR DISC HERNIATION WITH RADICULOPATHY. PER THE OPERATIVE NOTE, A DURA RENT WAS NOTED UNDERNEATH THE BONE AND COULD NOT BE REASONABLY SEWN THROUGH THE TUBE. THEREFORE, DURAL SEALANT WAS USED. THE SURGEON ASKED FOR ADHERUS DURAL SEALANT. THE OR TEAM STARTED WITH THE REGULAR, SHORT TIP SEALANT. THE SURGICAL TECH FOLLOWED THE INSTRUCTIONS ON THE PACKAGING AND WHEN THE SURGEON WENT TO USE IT, THE PRODUCT LEAKED OUT OF THE AREA OUT OF THE BOTTLES ATTACHED INSTEAD OF THE SPOUT. A SECOND DEVICE WAS OBTAINED, THIS TIME WITH THE LONGER APPLICATION TIP. AGAIN, THE INSTRUCTIONS WERE FOLLOWED STEP BY STEP WITH MULTIPLE STAFF IN THE OR FOLLOWING ALONG. THE PRODUCT AGAIN LEAKED OUT OF THE AREA WHERE THE BOTTLES ATTACHED. A THIRD DEVICE WAS OBTAINED, DIRECTIONS FOLLOWED EXACTLY AS BEFORE, AND THE THIRD TIME THE THIRD DEVICE WORKED WITH LEAKING AND WITHOUT ISSUE. TO NOTE THE STAFF IN THE OR THAT DAY WERE EXPERIENCED STAFF AND ARE MEMBERS OF THE NEURO/SPINE TEAM SO ARE FAMILIAR WITH ASSEMBLING THIS DEVICE. THERE WAS NO DEVIATION FROM THE IFU [INSTRUCTIONS FOR USE]. THIS APPEARS TO BE A PRODUCT FAILURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25080979 ↗MalfunctionAug 15, 2022Physical Resistance/Sticking
ADHERES AUTOSPRAY DURAL SEALANT MALFUNCTIONED AND BECAME STUCK WHEN TRYING TO MIX PRODUCT RESULTING IN ONLY PARTIAL MIXTURE AND INABILITY TO USE. FDA SAFETY REPORT ID# (B)(4).
Patient problems: Insufficient Information
MAUDE report 15233557 ↗
Other Sealant, Dural devices
- DuraSeal® 202050Integra Lifesciences Corporation · Sealant, Dural20 reports
- DuraSeal Exact Spinal Sealant System 206520Integra Lifesciences Corporation · Sealant, Dural15 reports
- DuraSeal® 202050Integra Lifesciences Corporation · Sealant, dural4 reports
- CraniSeal Dural Sealant CS-2055Pramand LLC · Sealant, Dural4 reports
- DuraSeal® Exact 206520Integra Lifesciences Corporation · Sealant, dural2 reports
- Duraseal 206520Covidien LP · Sealant, dural1 report
- DuraSeal® 202010DSIntegra Lifesciences Corporation · Sealant, Dural1 report
- SpineSeal Spine Sealant SS-2055Pramand LLC · Sealant, Dural1 report
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.