Alere Triage® 97021HSRecalls & reported problems
Alere San Diego, Inc. · Alere Triage® Cardiac Panel, TnI · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00899722002788), received between Sep 20, 2017 and Mar 6, 2019 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Low Test Results” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Low Test Results2
- High Test Results1
Patient problems
- No Consequences Or Impact To Patient3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 6, 2019High Test Results; High Test Results
CUSTOMER REPORTED THE FOLLOWING OCCURRING FOR ONE PATIENT. ON (B)(6) 2019 A MALE PATIENT PRESENTED TO THE ED WITH CONGESTION. AT 9:47 THE PATIENT WAS TESTED ON TRIAGE AND YIELDED A TNI OF 7.26. THE DOCTOR QUESTIONED THE RESULT BECAUSE IT WAS NOT CONSISTENT WITH PATIENT PRESENTATION. NEW SAMPLES WERE DRAWN FROM THE PATIENT AT 12:44 AND 16:37; BOTH DRAWS RESULTED IN TRIAGE TNI <0.05. CUSTOMER REPEATED THE 9:47 SAMPLE DRAW AT 18:32 AND RECEIVED A RESULT OF <0.05 TNI ON TRIAGE. CUSTOMER STATED THAT THE PATIENT WAS TREATED BASED ON SITE'S PROTOCOL FOR ELEVATED TNI WHEN THE 7.26 TRIAGE TNI RESULT WAS RECEIVED. THIS INCLUDED A SECOND EKG, A CHEST X-RAY, LABS OF CBC, CMP, CK, PT, AND PTT. ALL TESTS WERE NORMAL EXCEPT FOR ELEVATED GLU, AST AND ALT AND DECREASED CO2. THE EKG RESULT WAS STATED AS SINUS TACHYCARDIA. CUSTOMER LISTED HEPARIN AND PLAVIX AS THE MEDICATIONS PROVIDED TO THE PATIENT; IT IS UNCLEAR WHEN THESE MEDICATIONS WERE ADMINISTERED TO THE PATIENT. THE PATIENT WAS DIAGNOSED WITH PNEUMONIA. THE CUSTOMER STATED THE ELEVATED TRIAGE TNI RESULT INCREASED THE PATIENTS' LENGTH OF STAY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8396911 ↗MalfunctionJun 19, 2018Low Test Results; Low Test Results
CUSTOMER REPORTED THAT A DISCORDANT TROPONINI (TNI) RESULT WAS OBTAINED WITH A PATIENT SAMPLE ON THE TRIAGE SYSTEM. PATIENT PRESENTED TO THE ER WITH CHEST PAIN. PATIENT SAMPLE TESTED ON TRIAGE YIELDED <0.05 FOR TNI AND THE LAB RESULTED A FEW MINUTES LATER WITH 0.11NG/ML. LAB RESULT WAS ON THE BECKMAN ANALYZER, CUT-OFF 0.04NG/ML. TRIAGE CUT-OFF IS 0.05NG/ML. PATIENT'S EKG WAS ABNORMAL, WITH LATERAL ST/T WAVE, INDICATING ISCHEMIA. CUSTOMERS LAB MED DIRECTOR REVIEWED THE CHART FOR THE PATIENT, STATED THE EVENT WAS COMPLICATED BECAUSE THE PATIENT WAS IN AN ACCIDENT AND THE TROPONIN ELEVATION COULD BE HEART CONTUSION. NO FURTHER INFORMATION PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7617508 ↗MalfunctionSep 20, 2017Low Test Results; Low Test Results
THE CUSTOMER WAS PERFORMING A CORRELATION STUDY BETWEEN THE TRIAGE SYSTEM AND THE DIMENSIONS EXL ANALYZER. THE FOLLOWING EVENTS WERE REPORTED OCCURRING ON ONE PATIENT: BOTH AN EDTA PLASMA AND A HEPARIN PLASMA SAMPLE WAS COLLECTED FROM THE PATIENT. THE PLASMA SAMPLES WERE COLLECTED AND STORED FROZEN OVER THE WEEKEND. ON (B)(6) 2017, THE EDTA PLASMA SAMPLE WAS TESTED USING THE TRIAGE TNI PANEL AND PRODUCED A NORMAL TNI RESULT OF 0.19 NG/ML. THE HEPARIN PLASMA SAMPLE WAS TESTED ON THE DIMENSIONS EXL ANALYZER AND PRODUCED AN ABNORMAL TNI RESULT OF 0.63 NG/ML. THERE WAS NO TREATMENT OR INTERVENTION PROVIDED BASED ON THE TRIAGE TNI RESULT. THE FACILITY'S NORMAL TRIAGE TNI CUTOFF WAS REPORTED TO BE 0.40NG/ML THE FACILITY'S NORMAL DIMENSIONS EXL ANALYZER TNI CUTOFF WAS REPORTED TO BE 0.56NG/ML.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6881606 ↗
Other Immunoassay Method, Troponin Subunit devices
- i-STAT 03P90-25Abbott Point Of Care Inc. · IMMUNOASSAY METHOD, TROPONIN SUBUNIT474 reports
- Atellica IM TNIH 10997841Siemens Healthcare Diagnostics Inc. · IMMUNOASSAY METHOD, TROPONIN SUBUNIT463 reports
- Access hsTnI B52669Beckman Coulter, Inc. · Immunoassay Method, Troponin Subunit2 recalls170 reports
- i-STAT 06P23-25Abbott Point Of Care Inc. · IMMUNOASSAY METHOD, TROPONIN SUBUNIT195 reports
- VITROS 6802301Ortho Clinical Diagnostics · IMMUNOASSAY METHOD, TROPONIN SUBUNIT1 recall139 reports
- Atellica IM TNIH 10997840Siemens Healthcare Diagnostics Inc. · IMMUNOASSAY METHOD, TROPONIN SUBUNIT166 reports
- Dimension EXL LOCI Troponin I High Sensitive 10471068Siemens Healthcare Diagnostics Inc. · IMMUNOASSAY METHOD, TROPONIN SUBUNIT1 recall76 reports
- Access AccuTnI+3 A98143Beckman Coulter, Inc. · Immunoassay Method, Troponin Subunit1 death106 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page