API® 20 STREP 20600Recalls & reported problems
Biomerieux SA · API 20 Strep combines 20 biochemical tests to enable group or species ID of most streptococci and enterococci · Class 1
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 03573026048440), received between May 17, 2017 and May 27, 2021 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings2
- False Positive Result1
- Incorrect Or Inadequate Test Results1
Patient problems
- No Patient Involvement2
- No Clinical Signs, Symptoms or Conditions1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 27, 2021False Positive Result; False Positive Result
INTENDED USE: API® 20 STREP IS A QUALITATIVE STANDARDIZED SYSTEM FOR THE GROUP OR SPECIES IDENTIFICATION OF MOST STREPTOCOCCI AND ENTEROCOCCI, AND THOSE MOST COMMON RELATED ORGANISMS. IT USES MINIATURIZED TESTS AS WELL AS SPECIALLY ADAPTED DATABASE. INOCULATION AND READING OF THE STRIP ARE PERFORMED MANUALLY AND THE IDENTIFICATION IS OBTAINED USING AN IDENTIFICATION SOFTWARE. DESCRIPTION OF THE ISSUE: A CUSTOMER IN UNITED STATES NOTIFIED OF OBTAINING FALSE POSITIVE FOR ESCULIN FERRIC CITRATE (ESC) WHEN TESTING A QUALITY CONTROL STRAIN ; STREPTOCOCCUS EQUI SPP ZOOEPIDEMICUS ATCC® 700400¿ WITH THE API 20 STREP - REFERENCE 20600 - EXPIRATION DATE 21/09/2021. THE REACTION IN THE STRIP DOES NOT MATCH WITH A GOOD PATHOGEN PROFILE. THE TEST HAS BEEN PERFORMED THREE TIMES AND FALSE POSITIVE FOR ESC HAS BEEN OBTAINED EACH TIME. AS IT CONCERNS A QUALITY CONTROL TEST, NO PATIENT IMPACT HAS BEEN REPORTED AS A CONSEQUENCE OF THIS FALSE RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11895173 ↗MalfunctionNov 8, 2019Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
A CUSTOMER IN (B)(6) NOTIFIED BIOMERIEUX OF INCORRECT WELL REACTIONS FOR HIPPURIC ACID (HIP), BETA-GALACTOSIDASE (BGAL), AND PEPTIDOGLYCAN-ASSOCIATED LIPOPROTEIN (PAL) FOR AN EXTERNAL QC STREPTOCOCCUS DYSGALACTIAE STRAIN USING API 20 STREP 25STRIPS+25MEDIA STRIPS (REFERENCE (B)(4)). THE RESULTS THE CUSTOMER OBTAINED WERE CONSISTENT WITH AEROCOCCUS URINAE NOT STREPTOCOCCUS DYSGALACTIAE. BIOMERIEUX CUSTOMER SERVICE REQUESTED ADDITIONAL INFORMATION ABOUT THIS EVENT INCLUDING LOT NUMBER AND SAMPLE PREPARATION TECHNIQUE. ADDITIONALLY, CUSTOMER SERVICE RECOMMENDED TO PERFORM QC TESTED USING ISOLATES THAT ARE LISTED IN THE API 20 STREP 25STRIPS+25MEDIA PACKAGE INSERT. AT THE TIME OF THIS ASSESSMENT NO ADDITIONAL INFORMATION HAS BEEN PROVIDED AS THERE IS NO PATIENT ASSOCIATED WITH THIS QUALITY CONTROL STRAIN, THERE IS NO ADVERSE EVENT RELATED TO ANY PATIENT'S STATE OF HEALTH. BIOMERIEUX WILL INITIATE AN INTERNAL INVESTIGATION.
Patient problems: No Patient Involvement
MAUDE report 9296602 ↗MalfunctionNov 19, 2018Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
A CUSTOMER IN (B)(6) REPORTED A MISIDENTIFICATION OF AN ENTEROCOCCUS GALLINARUM SURVEY SAMPLE (CQE ECBU 2018-3A PROSPECTIVE BIOLOGY 11.1646 ) AS ENTEROCOCCUS FAECIUM IN ASSOCIATION WITH THE API® STREP TEST STRIP. THE CUSTOMER REPEATED THE TEST SEVERAL TIMES WITH SEVERAL LOTS AND FROM DIFFERENT CULTURE MEDIA, AND OBTAINED THE SAME RESULT. THE QUALITY CONTROL WAS CONFORMING. THERE WAS NO PATIENT INVOLVEMENT AS THE EVENT PERTAINED TO A SURVEY SAMPLE. A BIOMÉRIEUX INTERNAL INVESTIGATION WILL BE INITIATED.
Patient problems: No Patient Involvement
MAUDE report 8087036 ↗MalfunctionMay 17, 2017Incorrect Or Inadequate Test Results; Incorrect Or Inadequate Test Results
A CUSTOMER IN FRANCE NOTIFIED BIOMÉRIEUX OF DISCREPANT RESULTS ASSOCIATED WITH API® 20 STREP. THE CUSTOMER REPORTED THAT THE API® 20 STREP TEST GAVE A FALSE IDENTIFICATION. THE FALSE IDENTIFICATION OF AEROCOCCUS VIRIDANS WAS REPORTED TO THE CLINICIAN. THERE WAS NO INJURY TO PATIENT DUE TO THE FALSE IDENTIFICATION, HOWEVER THERE WAS A DELAY GREATER THAN 24 HOURS BECAUSE THE FIRST IDENTIFICATION WAS PERFORMED ON VITEK® MS AND THEN SUBSEQUENT TESTING ON THE API® 20 STREP STRIP. AN INTERNAL BIOMÉRIEUX INVESTIGATION WILL BE INITIATED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6572875 ↗
Other DISCS, STRIPS AND REAGENTS, MICROORGANISM DIFFERENTIATION devices
- API® LISTERIA 10300Biomerieux SA · DISCS, STRIPS AND REAGENTS, MICROORGANISM DIFFERENTIATION1 recall13 reports
- Chemmette Reagents 300145Microbiologics Inc. · Discs, strips and reagents, microorganism differentiation1 recall0 reports
- NIT 1 + NIT 2 70442Biomerieux SA · DISCS, STRIPS AND REAGENTS, MICROORGANISM DIFFERENTIATION1 recall0 reports
- VP 1 + VP 2 70422Biomerieux SA · DISCS, STRIPS AND REAGENTS, MICROORGANISM DIFFERENTIATION1 recall0 reports
- JAMES 70542Biomerieux SA · DISCS, STRIPS AND REAGENTS, MICROORGANISM DIFFERENTIATION1 recall0 reports
- RAPID ID 32 STREPBiomerieux SA · DISCS, STRIPS AND REAGENTS, MICROORGANISM DIFFERENTIATION5 reports
- BD BBL™ DMACA Indole Reagent Droppers 261187Becton, Dickinson And Company · DISCS, STRIPS AND REAGENTS, MICROORGANISM DIFFERENTIATION4 reports
- BD BBL™ DrySlide™ Oxidase 231746Becton, Dickinson And Company · DISCS, STRIPS AND REAGENTS, MICROORGANISM DIFFERENTIATION3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page