VIDAS® Progesterone 30409-01Recalls & reported problems
Biomerieux SA · VIDAS ® Progesterone (PRG) assay is a quantitative IA intended for use on VIDAS family instrument in the determination of progesterone concentration in human serum or plasma. · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 03573026156954), received between Nov 10, 2021 and Nov 10, 2021 — classified as 1 malfunction.
The most frequently reported specific device problem is “Unable to Obtain Readings” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Nov 9, 2022Class IIOpen, Classified
As the temperature and time out of range was exceeded, product performances cannot be guaranteed
Z-0866-2023 · Root cause: Storage · Recall details · FDA record ↗
Most reported problems
Device problems
- Unable to Obtain Readings1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 10, 2021Unable to Obtain Readings; Unable to Obtain Readings
ON (B)(6) 2021, A CUSTOMER IN UNITED STATES NOTIFIED BIOMÉRIEUX OF OBTAINING A SUBSTRATE ERROR WHILE USING THE VIDAS® PROGESTERONE 60 TESTS KIT (REFERENCE # 30409-01, LOT # 1008775630). IN THE CASE OF A SUBSTRATE ERROR, AN ERROR MESSAGE IS DISPLAYED ON THE EQUIPMENT VISIBLE BY THE USER, AND IT IS IMPOSSIBLE TO RUN TESTS AND GENERATE RESULTS. AT THE TIME OF THIS ASSESSMENT, THERE IS NO INFORMATION REGARDING DELAYS FOR ANY PATIENT SAMPLES. HOWEVER, IF THE CUSTOMER DOES NOT HAVE AN ALTERNATIVE TESTING METHOD, THIS COULD POTENTIALLY LEAD TO DELAYED PATIENT TEST RESULTS. THERE IS NO INDICATION OR REPORT FROM THE LABORATORY THAT THE ISSUE LED TO ANY ADVERSE EVENT RELATED TO ANY PATIENT'S STATE OF HEALTH. THE LOT IMPACTED BY THIS COMPLAINT IS PART OF THE SCOPE OF FSCA (B)(4) FOR SUBSTRATE ERROR ON MULTIPLE VIDAS® PRODUCT REFERENCES THAT WAS RELEASED ON (B)(6) 2021. A BIOMÉRIEUX INTERNAL INVESTIGATION HAS BEEN INITIATED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12789720 ↗
Other RADIOIMMUNOASSAY, PROGESTERONE devices
- IMMULITE® 2000 Systems PRG Progesterone 10381181Siemens Healthcare Diagnostics Products Ltd · RADIOIMMUNOASSAY, PROGESTERONE1 recall4 reports
- IMMULITE® 2000 Systems PRG Progesterone 10381170Siemens Healthcare Diagnostics Products Ltd · RADIOIMMUNOASSAY, PROGESTERONE1 recall3 reports
- ARCHITECT 7K77-25Abbott Ireland · RADIOIMMUNOASSAY, PROGESTERONE1 recall1 report
- Alinity 08P3620Abbott Ireland · RADIOIMMUNOASSAY, PROGESTERONE1 recall1 report
- IMMULITE 2000 Progesterone Kit Domestic 200T 10901873Siemens Healthcare Diagnostics Products Ltd · RADIOIMMUNOASSAY, PROGESTERONE1 recall0 reports
- IMMULITE 2000 Progesterone Kit Domestic 600T 10901874Siemens Healthcare Diagnostics Products Ltd · RADIOIMMUNOASSAY, PROGESTERONE1 recall0 reports
- IMMULITE®/IMMULITE® 1000 Systems PRG Progesterone 10381128Siemens Healthcare Diagnostics Products Ltd · RADIOIMMUNOASSAY, PROGESTERONE1 recall0 reports
- Atellica IM PRGE 10995659Siemens Healthcare Diagnostics Inc. · RADIOIMMUNOASSAY, PROGESTERONE16 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page