VIDAS® Lyme IgG II 417401Recalls & reported problems
Biomerieux SA · VIDAS Lyme IgG II assay is a qualitative immunoassay intended for use on VIDAS family instrument in the presumptive detection of human IgG antibodies to Borrelia burgdorferi. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 03573026458270), received between Dec 1, 2017 and Apr 3, 2020 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (5 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Sep 22, 2021Class IIOpen, Classified
bioMerieux has been receiving an increasing number of complaints linked to a VIDAS substrate error. It prevents the test from being run, therefore leads to potential delayed results as the user must run another test.
Z-0437-2022 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Most reported problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings5
- False Negative Result1
Patient problems
- No Known Impact Or Consequence To Patient6
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 3, 2020Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
A CUSTOMER IN THE UNITED STATES NOTIFIED BIOMÉRIEUX OF FALSE NEGATIVE RESULTS FOR THREE PATIENTS WHEN USING VIDAS® LYME IGG II 60 TESTS (REF. 417401, LOT 1007649070). FOR THREE PATIENTS, THE CUSTOMER OBTAINED NEGATIVE RESULTS WHEN USING VIDAS LYME IGG II 60 TESTS, LOT 1007649070 WITH RESULTS RESPECTIVELY AT 0.02, 0.03 AND 0.04. THE SAME SAMPLES TESTED WITH WESTERN BLOT IGG ASSAY SHOWED POSITIVE RESULTS. SAMPLE (B)(6), RFV:28.,TV: 0.02, RESULT:NEGATIVE. SAMPLE (B)(6), RFV:30, TV:0.03, RESULT: NEGATIVE. SAMPLE (B)(6): RFV , TV:0.04, RESULT NEGATVIE. WESTERN BLOT IGG SAMPLE (B)(6) POSITIVE. SAMPLE (B)(6) POSITIVE. SAMPLE (B)(6) POSITIVE. THERE IS NO INDICATION OR REPORT FROM THE LABORATORY THAT THE FALSE NEGATIVE RESULTS LED TO ANY ADVERSE EVENT RELATED TO ANY PATIENT'S STATE OF HEALTH.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9920397 ↗MalfunctionApr 3, 2020Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
A CUSTOMER IN THE UNITED STATES NOTIFIED BIOMÉRIEUX OF FALSE NEGATIVE RESULTS FOR THREE PATIENTS WHEN USING VIDAS® LYME IGG II 60 TESTS (REF. 417401, LOT 1007649070). FOR THREE PATIENTS, THE CUSTOMER OBTAINED NEGATIVE RESULTS WHEN USING VIDAS LYME IGG II 60 TESTS, LOT 1007649070 WITH RESULTS RESPECTIVELY AT 0.02, 0.03 AND 0.04. THE SAME SAMPLES TESTED WITH WESTERN BLOT IGG ASSAY SHOWED POSITIVE RESULTS. SAMPLE (B)(6), RFV:28.,TV: 0.02, RESULT:NEGATIVE. SAMPLE (B)(6), RFV:30, TV:0.03, RESULT: NEGATIVE. SAMPLE (B)(6): RFV , TV:0.04, RESULT NEGATVIE. WESTERN BLOT IGG SAMPLE (B)(6) POSITIVE. SAMPLE (B)(6) POSITIVE. SAMPLE (B)(6) POSITIVE. THERE IS NO INDICATION OR REPORT FROM THE LABORATORY THAT THE FALSE NEGATIVE RESULTS LED TO ANY ADVERSE EVENT RELATED TO ANY PATIENT'S STATE OF HEALTH.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9920382 ↗MalfunctionApr 3, 2020Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
A CUSTOMER IN THE UNITED STATES NOTIFIED BIOMÉRIEUX OF FALSE NEGATIVE RESULTS FOR THREE PATIENTS WHEN USING VIDAS® LYME IGG II 60 TESTS (REF. 417401, LOT 1007649070). FOR THREE PATIENTS, THE CUSTOMER OBTAINED NEGATIVE RESULTS WHEN USING VIDAS LYME IGG II 60 TESTS, LOT 1007649070 WITH RESULTS RESPECTIVELY AT 0.02, 0.03 AND 0.04. THE SAME SAMPLES TESTED WITH WESTERN BLOT IGG ASSAY SHOWED POSITIVE RESULTS. SAMPLE (B)(6), RFV:28.,TV: 0.02, RESULT:NEGATIVE. SAMPLE (B)(6), RFV:30, TV:0.03, RESULT: NEGATIVE. SAMPLE (B)(6): RFV , TV:0.04, RESULT NEGATVIE. WESTERN BLOT IGG: SAMPLE (B)(6) POSITIVE. SAMPLE (B)(6) POSITIVE. SAMPLE (B)(6) POSITIVE. THERE IS NO INDICATION OR REPORT FROM THE LABORATORY THAT THE FALSE NEGATIVE RESULTS LED TO ANY ADVERSE EVENT RELATED TO ANY PATIENT'S STATE OF HEALTH.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9920372 ↗MalfunctionJan 30, 2020Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
A CUSTOMER IN THE UNITED STATES NOTIFIED BIOMÉRIEUX OF OBTAINING A FALSE NEGATIVE RESULT FOR A PATIENT SAMPLE IN ASSOCIATION WITH THE VIDAS® LYME IGG 60 TESTS (REF 417401, LOT 1007326720). THE CUSTOMER TESTED THE PATIENT SAMPLE USING VIDAS® LYME IGG 60 TESTS WHICH OBTAINED A RESULT OF NEGATIVE. THE CUSTOMER ALSO TESTED THE SAMPLE USING LINE BLOT AND SOFIA TEST METHOD, BOTH ASSAYS YIELDED A RESULT OF POSITIVE FOR BOTH SAMPLES. THERE IS NO INDICATION OR REPORT FROM THE LABORATORY THAT THE DISCREPANT RESULTS LED TO ANY ADVERSE EVENT RELATED TO THE PATIENT'S STATE OF HEALTH. BIOMÉRIEUX WILL INITIATE AN INTERNAL INVESTIGATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9649682 ↗MalfunctionJan 30, 2020Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
A CUSTOMER IN THE UNITED STATES NOTIFIED BIOMÉRIEUX OF OBTAINING A FALSE NEGATIVE RESULT FOR A PATIENT SAMPLE IN ASSOCIATION WITH THE VIDAS® LYME IGG 60 TESTS (REF 417401, LOT 1007326720). THE CUSTOMER TESTED THE PATIENT SAMPLE USING VIDAS® LYME IGG 60 TESTS WHICH OBTAINED A RESULT OF NEGATIVE. THE CUSTOMER ALSO TESTED THE SAMPLE USING LINE BLOT AND SOFIA TEST METHOD, BOTH ASSAYS YIELDED A RESULT OF POSITIVE FOR BOTH SAMPLES. THERE IS NO INDICATION OR REPORT FROM THE LABORATORY THAT THE DISCREPANT RESULTS LED TO ANY ADVERSE EVENT RELATED TO THE PATIENT'S STATE OF HEALTH. BIOMÉRIEUX WILL INITIATE AN INTERNAL INVESTIGATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9649657 ↗MalfunctionDec 1, 2017False Negative Result; False Negative Result
A CUSTOMER FROM THE UNITED STATES NOTIFIED BIOMÉRIEUX OF FALSE NEGATIVE RESULTS ASSOCIATED WITH THE VIDAS® LYME IGG II INVOLVING A PATIENT SAMPLE. THE CUSTOMER INDICATED RUNNING THE PATIENT SAMPLE AND GETTING NEGATIVE RESULT; HOWEVER, THE SAMPLE WAS SENT TO A REFERENCE LAB WHERE THE WESTERN BLOT WAS POSITIVE FOR MULTIPLE BANDS. THE CUSTOMER REPORTED THE SPECIMEN THEY TESTED AND THE SPECIMEN THE REFERENCE LAB TESTED WERE OBTAINED FROM THE PATIENT AT THE SAME TIME BUT WERE IN DIFFERENT TUBES. SERUM WAS USED FOR BOTH TESTS. THE CUSTOMER INDICATED NEITHER SPECIMEN IS AVAILABLE FOR SUBMISSION TO BIOMÉRIEUX. A BIOMÉRIEUX INVESTIGATION WILL BE INITIATED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7076042 ↗
Other Reagent, Borrelia Serological Reagent devices
- VIDAS® Lyme IgM II 416436Biomerieux SA · Reagent, Borrelia Serological Reagent1 recall2 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page