ETEST® Tobramycin (TM)Recalls & reported problems
Biomerieux SA · Etest is a quantitative technique for determining susceptibility of clinically significant bacteria. It provides a gradient of 15 doubling dilutions for MIC determination. · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 03573026550301), received between May 21, 2021 and May 21, 2021 — classified as 1 malfunction.
The most frequently reported specific device problem is “Low Test Results” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (K102668); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Recalls naming this device
- Nov 9, 2022Open, Classified
As the temperature and time out of range was exceeded, product performances cannot be guaranteed
Z-0830-2023 · Root cause: Storage · Recall details · FDA record ↗
Most reported problems
Device problems
- Low Test Results1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 21, 2021Low Test Results; Low Test Results
INTENDED USE: ETEST® IS A QUANTITATIVE TECHNIQUE FOR DETERMINING THE ANTIMICROBIAL SUSCEPTIBILITY OF GRAM-NEGATIVE AND GRAM-POSITIVE AEROBIC BACTERIA SUCH AS ENTEROBACTERIACAEA, PSEUDOMONAS, STAPHYLOCOCCUS AND ENTEROCOCCUS SPECIES AND FASTIDIOUS BACTERIA, SUCH AS ANAEROBES, N. GONORRHOEAE, S. PNEUMONIAE, STEPTOCOCCUS AND HAEMOPHILUS SPECIES. THE SYSTEM COMPRISES A PREDEFINED ANTIBIOTIC GRADIENT WHICH IS USED TO DETERMINE THE MINIMUM INHIBITORY CONCENTRATION (MIC), IN ¿G/ML, OF DIFFERENT ANTIMICROBIAL AGENTS AGAINST MICROORGANISMS AS TESTED ON AGAR MEDIUM USING OVERNIGHT INCUBATION. DESCRIPTION OF THE ISSUE: A CUSTOMER IN UNITED STATES NOTIFIED BIOMÉRIEUX OF OBTAINING OUT-OF-RANGE-LOW RESULTS ON QUALITY STRAIN ENTEROCOCCUS FAECALIS ATCC® 29212¿ WHEN USING THE ETEST® TOBRAMYCIN TM 256 US S30 - REFERENCE 412478 - LOT #1007859110 - EXPIRATION DATE: 30/01/2022. THE CUSTOMER OBTAINED TOBRAMYCIN MIC = 4 UG/ML WHILE THE EXPECTED RANGE FOR THIS QC STRAIN IS FROM 8 UG/ML TO 32UG/ML. AS THERE IS NO PATIENT ASSOCIATED WITH THIS QUALITY CONTROL STRAIN, THERE IS NO ADVERSE EVENT RELATED TO ANY PATIENT'S STATE OF HEALTH. A BIOMÉRIEUX INTERNAL INVESTIGATION WILL BE INITIATED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11866572 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K102668. One submission often covers many models, so check the product and lot details in each record.
- Z-1531-2017 · Dec 15, 2016 · TerminatedETEST Tobramycin TM256 FOAM packaging, Product Name: ETEST Tobramycin TM 0.016-256(low). ETEST is a quantitative technique for determining the antimicrobial susceptibility of Gram-
Other Manual Antimicrobial Susceptibility Test Systems devices
- ETEST® Imipenem/Relebactam (IPR)Biomerieux SA · Manual antimicrobial susceptibility test systems1 recall4 reports
- ETEST® Amikacin (AK)Biomerieux SA · Manual Antimicrobial Susceptibility Test Systems1 recall2 reports
- ETEST® Erythromycin (EM)Biomerieux SA · Manual Antimicrobial Susceptibility Test Systems1 recall0 reports
- ETEST® Rifampicin (RI)Biomerieux SA · Manual Antimicrobial Susceptibility Test Systems1 recall0 reports
- ETEST® Eravacycline (ERV)Biomerieux SA · Manual antimicrobial susceptibility test systems1 recall0 reports
- ETEST® Benzylpenicillin (PG)Biomerieux SA · Manual Antimicrobial Susceptibility Test Systems1 recall0 reports
- ETEST® Ampicillin (AM)Biomerieux SA · Manual Antimicrobial Susceptibility Test Systems1 recall0 reports
- ETEST® Vancomycin (VA)Biomerieux SA · Manual antimicrobial susceptibility test systems1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page