Skip to content

VITEK® MS-DSRecalls & reported problems

Biomerieux SA · The VITEK MS-DS target slides enable the identification of microorganisms according to the MALDI-TOF mass spectrometry method. · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 03573026554224), received between Feb 16, 2021 and Feb 16, 2021 — classified as 1 malfunction.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Nov 9, 2022Open, Classified

    As the temperature and time out of range was exceeded, product performances cannot be guaranteed

    Z-0814-2023 · Root cause: Storage · FDA record ↗

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings1

Patient problems

  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 16, 2021Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    A CUSTOMER FROM (B)(6) NOTIFIED BIOMÉRIEUX OF A MISIDENTIFICATION OF ESCHERICHIA COLI AS SALMONELLA WHEN TESTING A SAMPLE WITH VITEK® MS-DS (REF. 410893) ¿ LOT 11111256BM USED IN ASSOCIATION WITH VITEK® MS INSTRUMENT (REF. 410895) ¿ SERIAL NUMBER (B)(4). THE CUSTOMER REPORTED HE OBTAINED SALMONELLA, 99.4% DURING INITIAL TESTING. THE CUSTOMER PERFORMED AN ANTIBIOTIC SUSCEPTIBILITY TEST AND REPORTED THAT THE RESULTS DID NOT CORRELATE TO A SALMONELLA IDENTIFICATION. HE RETESTED THE SAMPLE VIA VITEK® MS AND OBTAINED ESCHERICHIA COLI. THE RESULT OF THE FIRST IDENTIFICATION (SALMONELLA) WAS REPORTED TO THE CLINICIAN. HOWEVER, THERE IS NO INDICATION OR REPORT FROM THE LABORATORY THAT THE DISCREPANT RESULT LED TO ANY ADVERSE EVENT RELATED TO ANY PATIENT'S STATE OF HEALTH.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 11331697 ↗

Other System, Mass Spectrometry, Maldi Tof, Microorganism Identification, Cultured Isolates devices

All System, Mass Spectrometry, Maldi Tof, Microorganism Identification, Cultured Isolates devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.