Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 03573026588649), received between Aug 21, 2020 and Aug 21, 2020 — classified as 1 malfunction.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings1
Patient problems
- No Patient Involvement1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 21, 2020Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
AN INDUSTRY CUSTOMER IN (B)(6)NOTIFIED BIOMÉRIUX OF A MISIDENTIFICATION OF CRONOBACTER STRAINS AS ENTEROBACTER CLOACAE SPECIES IN ASSOCIATION WITH THE API® RAPID ID 32 E 25STRIPS (REF. (B)(4), LOT. 1007739570) WHEN TESTING COLONIES ORIGINALLY SOURCED FROM MILK POWDER. THE SAMPLE WAS TESTED USING GENE-UP WHICH OBTAINED A CRO POSITIVE RESULT. THE ORGANISM WAS THEN SUBCULTURED ONTO ESIA AGAR BLUE-GREEN COLONIES WERE OBSERVED INDICATING CRONOBACTER. THEN COLONIES WERE SUBCULTURED ON A NUTRITIVE AGAR AND THE CONFIRMATION OF THE IDENTIFICATION PERFORMED WITH THE IMPACTED LOT OF RAPID ID 32 E STRIPS THAT GAVE ENTEROBACTER CLOACAE SPECIES INSTEAD OF CRONOBACTER AS EXPECTED. AS THERE IS NO PATIENT ASSOCIATED WITH THIS INDUSTRY STRAIN, THERE IS NO ADVERSE EVENT RELATED TO ANY PATIENT'S STATE OF HEALTH.
Patient problems: No Patient Involvement
MAUDE report 10437425 ↗
Other KIT, IDENTIFICATION, ENTEROBACTERIACEAE devices
- API® 20 EBiomerieux SA · KIT, IDENTIFICATION, ENTEROBACTERIACEAE1 recall14 reports
- API® 20 NE 20050Biomerieux SA · KIT, IDENTIFICATION, ENTEROBACTERIACEAE7 reports
- RAPID 20 E™Biomerieux SA · KIT, IDENTIFICATION, ENTEROBACTERIACEAE2 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page