Skip to content

CHROMID® Salmonella EliteRecalls & reported problems

Biomerieux SA · chromID™ Salmonella Eliteagar is a chromogenic medium for the selective isolation and identification of Salmonellain human specimens (stools) · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 03573026597313), received between Apr 24, 2024 and Apr 24, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “False Negative Result” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • False Negative Result1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 24, 2024False Negative Result; False Negative Result

    INTENDED USE: CHROMID® SALMONELLA ELITE AGAR IS A CHROMOGENIC MEDIUM FOR THE SELECTIVE ISOLATION AND IDENTIFICATION OF SALMONELLA IN HUMAN SPECIMENS (STOOLS). ISSUE DESCRIPTION: A CUSTOMER IN SPAIN NOTIFIED BIOMÉRIEUX OF NO GROWTH WITH A SALMONELLA STRAIN IN ASSOCIATION WITH CHROMID SALMONELLA ELIT 20 PLT (REF. 412108, LOT UNKNOWN). THE CUSTOMER STATED THAT WITH SALMONELLA CHROMOGENIC PLATES THEY DO NOT DETECT SOME STRAINS BECAUSE THEY DO NOT CHANGE THE COLOR. THEY HAVE CHECKED WITH MALDI-TOF AND THE STRAINS ARE SALMONELLA BUT THE CHROMID MEDIUM HAS NOT DETECTED THEM. THERE IS NO DELAY IN DIAGNOSIS BECAUSE THEY USE OTHER MEANS IN PARALLEL. AT THE TIME OF ASSESSMENT, THERE IS NO INDICATION OR REPORT FROM THE CUSTOMER THAT THIS EVENT LED TO ANY ADVERSE EVENT RELATED TO ANY PATIENT'S STATE OF HEALTH. AN INVESTIGATION WILL BE INITIATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19172914 ↗

Other CULTURE MEDIA, SELECTIVE AND DIFFERENTIAL devices

All CULTURE MEDIA, SELECTIVE AND DIFFERENTIAL devices →