VIDAS® NEPHROCHECK® 421172-01Recalls & reported problems

Biomerieux SA · VIDAS NEPHROCHECK 60 TESTS is an automated quantitative test to be used on VIDAS3. It is intended to be used as an aid in the risk assessment for moderate or severe acute kidney injury · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 03573026616038), received between Apr 12, 2024 and Apr 12, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Low Test Results” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Low Test Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 12, 2024Low Test Results; Low Test Results

    PRODUCT DESCRIPTION. VIDAS® NEPHROCHECK® IS AN AUTOMATED TEST FOR USE ON THE VIDAS® 3 INSTRUMENT FOR THE IMMUNOENZYMATIC QUANTITATIVE DETERMINATION OF TIMP-2 (TISSUE INHIBITOR OF METALLOPROTEINASE-2) AND IGFBP-7 (INSULIN-LIKE GROWTH FACTOR-BINDING PROTEIN 7) PROTEINS IN HUMAN URINE USING THE ELFA TECHNIQUE (ENZYME LINKED FLUORESCENT ASSAY) FOR CALCULATION OF THE AKIRISK¿ SCORE. THE VIDAS® NEPHROCHECK® ASSAY IS INTENDED TO BE USED IN CONJUNCTION WITH CLINICAL EVALUATION IN PATIENTS WHO CURRENTLY HAVE OR HAVE HAD WITHIN THE PAST 24 HOURS ACUTE CARDIOVASCULAR AND OR RESPIRATORY COMPROMISE AND ARE ICU PATIENTS AS AN AID IN THE RISK ASSESSMENT FOR MODERATE OR SEVERE ACUTE KIDNEY INJURY (AKI) WITHIN 12 HOURS OF PATIENT ASSESSMENT. THE VIDAS® NEPHROCHECK® TEST IS INTENDED TO BE USED IN PATIENTS 21 YEARS OF AGE OR OLDER. ISSUE DESCRIPTION . ON 29-MAR-2024, A CUSTOMER FROM THE UNITED STATES NOTIFIED BIOMÉRIEUX OF OBTAINING AN OUT-OF-RANGE RESULTS WHEN TESTING AN EXTERNAL QUALITY CONTROL WITH VIDAS NEPHROCHECK 60 T - US (REF. 421172-01, LOT: 1010057250, EXPIRY DATE: 24-SEP-2024). THE CUSTOMER TESTED THE EXTERNAL QUALITY CONTROL LQC FROM CLINIQA (LOT 2312117) WITH VIDAS NEPHROCHECK 60 T - US

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19097501 ↗

Other Acute kidney injury test system devices

All Acute kidney injury test system devices →

Sources and updates