ILIZAROV SYSTEMS 102911Recalls & reported problems
Smith & Nephew, Inc. · FIXATION BOLT WRENCH · Class 1
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 03596010031839), received between Feb 5, 2019 and Jul 18, 2022 — classified as 1 injury, 11 malfunctions.
The most frequently reported specific device problem is “Break” (10 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Break10
- Crack2
Patient problems
- No Clinical Signs, Symptoms or Conditions9
- No Consequences Or Impact To Patient3
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 18, 2022Crack; Crack
IT WAS REPORTED THAT, DURING SURGERY, THE SIDE PART OF SPANNER OF THE FIXATION BOLT WRENCH CRACKED AND COME AWAY FROM MAIN BODY WHILST TIGHTENING A BOLT. DEVICE OUTSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITHOUT DELAY USING A S+N BACK-UP DEVICE. PATIENT WAS NOT HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15038800 ↗MalfunctionMar 31, 2022Break; Break
IT WAS REPORTED THAT, DURING SURGERY, THE FIXATION BOLT WRENCH BROKE, OUTSIDE OF THE PATIENT. PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, WITH A BACK-UP DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13961431 ↗MalfunctionOct 29, 2021Crack; Crack
IT WAS REPORTED THAT, DURING AN EXTERNAL FIXATION SURGERY, A FIXATION BOLT WRENCH BROKE OFF. NO SURGICAL DELAY WAS REPORTED TO OCCUR DUE TO THIS INCIDENT, AS A SMITH AND NEPHEW BACKUP DEVICE WAS AVAILABLE TO CONCLUDE THE PROCEDURE. NO PATIENT HARM WAS REPORTED TO OCCUR DUE TO THIS MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12719633 ↗MalfunctionSep 28, 2021Break; Break
IT WAS REPORTED THAT, DURING AN EXTERNAL FIXATION PROCEDURE, A FIXATION BOLT WRENCH BROKE WHILE TIGHTENING, INSTRUMENT WERE OUTSIDE THE PATIENT. PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, WITH A BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12539785 ↗MalfunctionAug 5, 2021Break; Break
IT WAS REPORTED THAT, DURING AN EXTERNAL FIXATION ONE SIDE OF A FIXATION BOLT WRENCH WERE SHEARED OFF DURING THE TIGHTENING OF A NUT. SURGERY WAS FINISHED WITH A SMITH AND NEPHEW BACK UP DEVICE, WITHOUT ANY SURGICAL DELAY. PATIENT WAS NOT HARMED AS A CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12285685 ↗MalfunctionApr 11, 2021Break; Break
IT WAS REPORTED THAT DURING SURGERY, THE FIXATION BOLT WRENCH BROKE WHILE USING IT OUTSIDE THE PATIENT. THE PROCEDURE WAS FINISHED USING A SMITH AND NEPHEW BACK UP DEVICE, WITHOUT SURGICAL DELAY AND NO INJURY TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11646834 ↗InjuryMar 23, 2021Break; Break
IT WAS REPORTED THAT, DURING PROCEDURE WHEN TIGHTENING FIXATION BOLT WRENCH, THIS BROKE. INSTRUMENTS OUTSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITHOUT DELAY USING A SMITH-NEPHEW BACK-UP DEVICE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11548663 ↗MalfunctionFeb 21, 2021Break; Break
IT WAS REPORTED THAT THE FIXATION BOLT WRENCH IS BROKEN. UNKNOWN WHEN THIS EVENT HAPPENED AND IF THERE IS A PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11356760 ↗MalfunctionJan 18, 2021Break; Break
IT WAS REPORTED THAT, THE END BIT OF THE WRENCH BROKE OFF DURING USE OUTSIDE OF THE PATIENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT DELAY USING A SMITH+NEPHEW BACK-UP DEVICE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11189774 ↗MalfunctionJul 24, 2019Break; Break
IT WAS REPORTED THAT DURING PROCEDURE, SURGEON WAS TIGHTENING COMBI BOLT AND ONE SIDE OF THE WRENCH TEETH BROKE OFF. NO MORE INFORMATION PROVIDED YET.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8824199 ↗MalfunctionMay 26, 2019Break; Break
IT WAS REPORTED THAT DURING A SURGERY, WIRE FIXATION WRENCH EDGES HAVE STARTED TO BREAK AWAY ABOVE THE INCISION. NO PIECES FELL INTO THE WOUND AND NO INJURY. THERE WAS NO DELAY. PROCEDURE WAS CONCLUDED WITH A BACKUP DEVICE FROM SMITH AND NEPHEW.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8643147 ↗MalfunctionFeb 5, 2019Break; Break
IT WAS REPORTED THAT DURING SURGERY, FRAGMENT OF THE 90-DEGREE WRENCH BROKE OFF WHEN TIGHTENING A BOLT DURING NORMAL USE. NO DELAY OR INJURY REPORTED. A BACKUP DEVICE WAS NOT AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8307908 ↗
Other WRENCH devices
- MONARCH 277040510Medos International Sàrl · WRENCH199 reports
- PINNACLE 2217-50-008Depuy (Ireland) · WRENCH108 reports
- SIGMA 96-1673Depuy (Ireland) · WRENCH102 reports
- P.F.C. 86-5189Depuy (Ireland) · WRENCH92 reports
- Veptr 03.641.004Synthes GmbH · WRENCH86 reports
- RECLAIM 2975-00-300Depuy (Ireland) · WRENCH51 reports
- PINNACLE 2217-50-006Depuy (Ireland) · WRENCH34 reports
- NA 388.261Synthes GmbH · WRENCH32 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.