SPECTRON 71312102Recalls & reported problems
Smith & Nephew, Inc. · SPECTRON EF PRIMARY STANDARD OFFSET 12/14 SIZE 2 · Class 3 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 7
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 03596010195524), received between Nov 11, 2020 and Sep 11, 2025 — classified as 7 injuries.
The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Loosening of Implant Not Related to Bone-Ingrowth5
- Adverse Event Without Identified Device or Use Problem2
Patient problems
- Inadequate Osseointegration5
- Osteolysis3
- Bone Fracture(s)1
- Injury1
- Metal Related Pathology1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjurySep 11, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT, AFTER A THA SURGERY PERFORMED ON AN UNSPECIFIED DATE, THE PATIENT EXPERIENCED A FEMUR FRACTURE. A REVISION SURGERY WAS PERFORMED ON AN UNSPECIFIED DATE TO REVISE THE SPECTRON EF PRIMARY STANDARD OFFSET 12/14 SZ 2 AND DURALOC LINER. THERE WAS SIGNIFICANT BONE LOSS AND VERY LITTLE CORTICAL BONE REMAINING IN THE LEFT HIP (BOTH HIPS HAD PREVIOUSLY HAD HIP REPLACEMENTS BUT ONLY THE LEFT HAD SIGNIFICANT BONE LOSS). A ZIMMER NCB PLATE WAS USED TO REPAIR THE ORIF AND THEN A CORAIL CEMENTED KHO SIZE 14 STEM WAS USED (BASED ON DIMENSIONS, THE PREVIOUS STEM THAT WAS INSITU WAS BETWEEN A 11-12 CORAIL STEM). HOWEVER, DUE TO SIGNIFICANT BONE LOSS THIS WAS NO LONGER LARGE ENOUGH. THE DURALOC LINER WAS NOT WORN, BUT IT WAS ALSO EXPLANTED DUE TO ALREADY BEING UNDERWAY WITH A REVISION AND IT HAD BEEN INSITU FOR ~25 YEARS, SO IT WAS DETERMINED TO BE WORTHWHILE TO REPLACE WHILE ALREADY COMPLETING SURGERY. THE 52MM LOCKING RING WAS REPLACED AND THE SAME LINER, A 52+10 DEGREE LIPPED LINER, WAS PUT IN SITU. THE ONLY DIFFERENCE IN THE NEW LINER IMPLANTED WAS THAT IT ACCEPTS A 32MM HEAD RATHER THAN THE PREVIOUS 28MM HEAD. THE CURRENT HEALTH STATUS OF THE PATIENT IS UNKNOWN.
Patient problems: Bone Fracture(s)
MAUDE report 23028528 ↗InjuryJan 25, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
IT WAS REPORTED THAT ON LITERATURE REVIEW "INFLAMMATORY TISSUE REACTIONS AROUND ASEPTICALLY LOOSE CEMENTED HIP PROSTHESES: A RETRIEVAL STUDY OF THE SPECTRON EF STEM WITH REFLECTION ALL-POLY ACETABULAR CUP" 136 MONTHS AFTER A THA SURGERY, WHERE AN SPECTRON EF STEM AND A REFLECTION NON-CROSSLINKED ALL-POLLY CUP WERE IMPLANTED, THE PATIENT UNDERWENT A REVISION SURGERY DUE TO LOOSING OF THE CUP AND THE STEM. THE STUDY DETERMINED THAT BOTH THE STEM AND THE CUP WERE WORN, ADDITIONALLY PE, CR, CO AND ZR PARTICLES WERE PRESENT IN THE PERIPROSTHETIC TISSUE. CURRENT HEALTH STATUS OF THE PATIENT IS UNKNOWN. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: Inadequate Osseointegration
MAUDE report 18577303 ↗InjuryJan 24, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
IT WAS REPORTED THAT ON LITERATURE REVIEW "INFLAMMATORY TISSUE REACTIONS AROUND ASEPTICALLY LOOSE CEMENTED HIP PROSTHESES: A RETRIEVAL STUDY OF THE SPECTRON EF STEM WITH REFLECTION ALL-POLY ACETABULAR CUP" 114 MONTHS AFTER A THA SURGERY, WHERE AN SPECTRON EF STEM AND A REFLECTION NON-CROSSLINKED ALL-POLLY CUP WERE IMPLANTED, THE PATIENT UNDERWENT A REVISION SURGERY DUE TO LOOSINNG OF THE CUP AND THE STEM. THE STUDY DETERMINED THAT BOTH THE STEM AND THE CUP WERE WORN, ADDITIONALLY PE, CR, CO AND ZR PARTICLES WERE PRESENT IN THE PERIPROSTHETIC TISSUE. CURRENT HEALTH STATUS OF THE PATIENT IS UNKNOWN. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: Inadequate Osseointegration
MAUDE report 18577218 ↗InjuryJan 19, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
IT WAS REPORTED THAT ON LITERATURE REVIEW "INFLAMMATORY TISSUE REACTIONS AROUND ASEPTICALLY LOOSE CEMENTED HIP PROSTHESES: A RETRIEVAL STUDY OF THE SPECTRON EF STEM WITH REFLECTION ALL-POLY ACETABULAR CUP" 138 MONTHS AFTER A THA SURGERY, WHERE AN SPECTRON EF STEM AND A REFLECTION NON-CROSSLINKED ALL-POLLY CUP WERE IMPLANTED, THE PATIENT UNDERWENT A REVISION SURGERY DUE TO LOOSENING OF THE CUP AND THE STEM, AND OSTEOLYSIS. THE STUDY DETERMINED THAT BOTH THE STEM AND THE CUP WERE WORN, ADDITIONALLY PE, CR, CO AND ZN PARTICLES WERE PRESENT IN THE PERIPROSTHETIC TISSUE. CURRENT HEALTH STATUS OF THE PATIENT IS UNKNOWN. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: Inadequate Osseointegration; Osteolysis
MAUDE report 18543223 ↗InjuryJan 17, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
IT WAS REPORTED THAT ON LITERATURE REVIEW "INFLAMMATORY TISSUE REACTIONS AROUND ASEPTICALLY LOOSE CEMENTED HIP PROSTHESES: A RETRIEVAL STUDY OF THE SPECTRON EF STEM WITH REFLECTION ALL-POLY ACETABULAR CUP" 188 MONTHS AFTER A THA SURGERY, WHERE AN SPECTRON EF STEM AND A REFLECTION NON-CROSSLINKED ALL-POLLY CUP WERE IMPLANTED, THE PATIENT UNDERWENT A REVISION SURGERY DUE TO LOOSINNG OF THE CUP AND THE STEM. THE STUDY DETERMINED THAT BOTH THE STEM AND THE CUP WERE WORN, ADDITIONALLY PE, CR, CO AND ZR PARTICLES WERE PRESENT IN THE PERIPROSTHETIC TISSUE. CURRENT HEALTH STATUS OF THE PATIENT IS UNKNOWN. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: Inadequate Osseointegration
MAUDE report 18531935 ↗InjuryJan 17, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
IT WAS REPORTED THAT ON LITERATURE REVIEW "INFLAMMATORY TISSUE REACTIONS AROUND ASEPTICALLY LOOSE CEMENTED HIP PROSTHESES: A RETRIEVAL STUDY OF THE SPECTRON EF STEM WITH REFLECTION ALL-POLY ACETABULAR CUP" 144 MONTHS AFTER A THA SURGERY, WHERE AN SPECTRON EF STEM AND A REFLECTION NON-CROSSLINKED ALL-POLLY CUP WERE IMPLANTED, THE PATIENT UNDERWENT A REVISION SURGERY DUE TO LOOSINNG OF THE CUP AND THE STEM, AND OSTEOLYSIS. THE STUDY DETERMINED THAT BOTH THE STEM AND THE CUP WERE WORN, ADDITIONALLY PE, CR, CO AND ZR PARTICLES WERE PRESENT IN THE PERIPROSTHETIC TISSUE. CURRENT HEALTH STATUS OF THE PATIENT IS UNKNOWN. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: Inadequate Osseointegration; Osteolysis
MAUDE report 18531899 ↗InjuryNov 11, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A REVISION SURGERY WAS PERFORMED ON A TOTAL HIP REPLACEMENT AFTER 5 YEARS OF THE PRIMARY SURGERY DUE TO METALLOSIS OF SOME PART OF THE PROSTHESIS. THE PATIENT SUFFERED FROM SOFT TISSUE DAMAGE AND BONE LOSS. IT WAS A CHANGED IN THE SURGICAL TECHNIQUE. IT IS UNKNOWN IF THERE WAS A DELAY IN THE SURGERY OR THE PATIENT HEALTH.
Patient problems: Injury; Metal Related Pathology; Osteolysis
MAUDE report 10825973 ↗
Other Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Ceramic/Metal, Cemented Or Uncemented devices
- BIOLOX DELTA 1365-36-310Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented839 reports
- BIOLOX DELTA 1365-36-320Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented643 reports
- PINNACLE POROCOAT 1217-22-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented335 reports
- PINNACLE POROCOAT 1217-22-054Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented263 reports
- BIOLOX DELTA 1365-36-330Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented255 reports
- PINNACLE POROCOAT 1217-22-050Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented253 reports
- PINNACLE POROCOAT 1217-22-056Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented220 reports
- SUMMIT POROCOAT 1570-11-120Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/ceramic/metal, cemented or uncemented1 death204 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page