GENESIS II 71420572Recalls & reported problems
Smith & Nephew, Inc. · GENESIS II BICONVEX PATELLA 32MM · Class 2 · Implant
All models of GENESIS II (462)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 03596010208286), received between May 5, 2018 and Sep 5, 2023 — classified as 5 injuries.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).
6 other recalls cover devices cleared under the same FDA submission (K951987); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Device Dislodged or Dislocated1
- Loose or Intermittent Connection1
- Migration1
- Unstable1
Patient problems
- Failure of Implant2
- Injury1
- Joint Dislocation1
- Unspecified Infection1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjurySep 5, 2023Migration; Migration
IT WAS REPORTED THAT, AFTER A TKA THAT TOOK PLACE ON (B)(6) 2008. A REVISION SURGERY HAD TO BE PERFORMED ON (B)(6) 2023 DUE TO THE GNS II BICONVEX PAT 32MM BECOMING DISLODGED, RESULTING IN IT FREELY MOVING IN THE KNEE. THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.
Patient problems: Failure of Implant
MAUDE report 17687700 ↗InjuryOct 9, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
US LEGAL. IT WAS REPORTED THAT, AFTER AN TKA SURGERY HAD BEEN PERFORMED ON (B)(6) 2008, THE POST OF A GII PS HI FLEX ISRT SZ 5-6 11 BROKE, MAKING THE KNEE UNSTABLE. ON (B)(6) 2020, AN ARTHROSCOPY WAS PERFORMED TO CONFIRM POST BREAKAGE. THE INSERT WAS EXCHANGED DURING A REVISION SURGERY ON (B)(6) 2020. DURING SURGERY, THE PATELLA SHOWED MINIMAL WEAR AND A PATELLAR TIBIAL TUBERCULE PIN WAS PLACED. THE PATIENT TOLERATED THE PROCEDURE WELL AND WAS THEN TRANSPORTED TO THE RECOVERY ROOM IN GOOD CONDITION.
Patient problems: Failure of Implant
MAUDE report 12602211 ↗InjuryJul 24, 2019Loose or Intermittent Connection; Loose or Intermittent Connection
IT WAS REPORTED THAT THE PATELLA WAS REVISED DUE TO LOOSENING AND POSSIBLE VASCULAR NECROSIS OF THE NATIVE PATELLA. DR. STATED THAT THE PEGS ON THE ORIGINAL 35 MM RESURFACING GEN II PATELLA REF 71420578 WERE NOT IN THE SAME PLACE AS THE CEMENT PLUGS, SO THE COMPONENT SEEMS TO HAVE MOVED INSIDE THE BONE. THE PATELLA WAS REMOVED AND REPLACED WITH A 32 MM BICONVEX PATELLA (PICTURED). NO OTHER COMPONENTS WERE TOUCHED. NO MORE INFORMATION PROVIDED YET.
Patient problems: Injury
MAUDE report 8821442 ↗InjuryJul 22, 2019Unstable; Unstable
IT WAS REPORTED THAT A SECOND STAGE REVISION WAS PERFORMED DUE TO INFECTION AND LOOSENING.
Patient problems: Unspecified Infection
MAUDE report 8813871 ↗InjuryMay 5, 2018Device Dislodged or Dislocated; Device Dislodged or Dislocated
PATIENT PRESENTED 4 YEARS POST TKR WITH PATELLA IMPLANT BROKEN OFF AND LODGED BEHIND FEMORAL PROSTHESIS BETWEEN POSTERIOR CONDYLES. REVISION SURGERY REQUIRED TO REMOVE BROKEN/LOOSE PATELLA, AND TIBIAL INSERT MAY NEED TO BE REMOVED (EXCHANGED) TO FACILITATE REMOVAL.
Patient problems: Joint Dislocation
MAUDE report 7489385 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K951987. One submission often covers many models, so check the product and lot details in each record.
- Z-1418-2022 · Jun 27, 2022 · Terminated(1) Genesis II Non-Porous Tibial Baseplate Size 5 Right, Product #71420186. (2) Genesis II Non-Porous Tibial Baseplate Size 4 Left, Product #71420166.
- Z-2774-2020 · Jul 16, 2020 · Open, ClassifiedTibial knee prosthesis
- Z-2883-2018 · Aug 1, 2017 · TerminatedGENESIS II NON-POROUS CRUCIATE RETAINING FEMORAL COMPONENT, SIZE 6 RIGHT, REF 71420026 Knee prosthesis
- Z-0573-2016 · Nov 9, 2015 · TerminatedSmith & Nephew GENESIS(R) II/LEGION(TM) PATELLAR COMPONENT, RESURFACING ROUND, 32MM STANDARD, FOR CEMENTED USE ONLY WITH JOURNEY PEG, STERILE EO, Rx only, REF Number: 71934490 kn
- Z-0131-2014 · Oct 7, 2013 · TerminatedGENESIS(R) II, NON-POROUS TIBIAL BASE, SIZE 6, RIGHT, QTY: (1), REF 71420188. Product Usage: orthopaedic
- Z-0130-2014 · Oct 7, 2013 · TerminatedGENESIS(R) II, NON-POROUS TIBIAL BASE, SIZE 5, LEFT, QTY: (1), REF 71420168. Product Usage: orthopaedic
Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.