GENESIS II 71420820Recalls & reported problems
Smith & Nephew, Inc. · GENESIS II POSTERIOR STABILIZED INSERT SIZE 3-4 13MM · Class 2 · Implant
All models of GENESIS II (462)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 03596010209597), received between Nov 14, 2017 and Oct 25, 2023 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Difficult to Insert” (1 report).
6 other recalls cover devices cleared under the same FDA submission (K951987); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Difficult to Insert1
- Mechanical Jam1
Patient problems
- Failure of Implant2
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 25, 2023Difficult to Insert; Difficult to Insert
IT WAS REPORTED THAT, DURING A TKA, THE CORRECT EXPOSURE OF THE TIBIA WAS PERFORMED IN ORDER TO PLACE THE GII PS INSERT SZ 3-4 13MM. THE POSITIONING OF THE INSERT WAS PERFORMED AND THEN WITH THE DEFINITIVE IMPACTOR TO PERFORM THE PROPER BLOCK IN THE DEFINITIVE TIBIA, IT WAS VISUALIZED THAT THERE IS A LIGHT BETWEEN THE INSERT AND THE TIBIA. THEN, IT WAS IMPACTED WITHOUT OBTAINING THE PROPER INSERTION. THIS PROCESS WAS PERFORMED THREE TIMES WITHOUT ANY SUCCESS. SURGERY WAS PERFORMED AFTER A DELAY GREATER THAN 30 MIN, WITH A BACK-UP DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18008245 ↗InjuryMar 30, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT, AFTER 13 YEARS FROM THE PRIMARY PROCEDURE, A REVISION SURGERY TO REPLACED A POLY WAS PERFORMED DUE TO A POST-BROKE IN THE PATIENT. THE GII PS INSERT SZ 3-4 13MM WAS EXPLANTED. PATIENT OUTCOME IS UNKNOWN.
Patient problems: Failure of Implant
MAUDE report 11593552 ↗InjuryNov 14, 2017Mechanical Jam; Mechanical Jam
IT WAS REPORTED THAT DURING A SURGERY THE INSERT DID NOT LOCK INTO THE LEFT TIBIAL BASEPLATE. IT WAS FOUND THAT THE TRAILING EDGE OF THE INSERT AND THE JOINT OF TIBIAL BASEPLATE DID NOT MATCH. THE TOTAL DELAY OF THE SURGERY WAS 70 MINUTES AND ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY.
Patient problems: Failure of Implant
MAUDE report 7028069 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K951987. One submission often covers many models, so check the product and lot details in each record.
- Z-1418-2022 · Jun 27, 2022 · Terminated(1) Genesis II Non-Porous Tibial Baseplate Size 5 Right, Product #71420186. (2) Genesis II Non-Porous Tibial Baseplate Size 4 Left, Product #71420166.
- Z-2774-2020 · Jul 16, 2020 · Open, ClassifiedTibial knee prosthesis
- Z-2883-2018 · Aug 1, 2017 · TerminatedGENESIS II NON-POROUS CRUCIATE RETAINING FEMORAL COMPONENT, SIZE 6 RIGHT, REF 71420026 Knee prosthesis
- Z-0573-2016 · Nov 9, 2015 · TerminatedSmith & Nephew GENESIS(R) II/LEGION(TM) PATELLAR COMPONENT, RESURFACING ROUND, 32MM STANDARD, FOR CEMENTED USE ONLY WITH JOURNEY PEG, STERILE EO, Rx only, REF Number: 71934490 kn
- Z-0130-2014 · Oct 7, 2013 · TerminatedGENESIS(R) II, NON-POROUS TIBIAL BASE, SIZE 5, LEFT, QTY: (1), REF 71420168. Product Usage: orthopaedic
- Z-0131-2014 · Oct 7, 2013 · TerminatedGENESIS(R) II, NON-POROUS TIBIAL BASE, SIZE 6, RIGHT, QTY: (1), REF 71420188. Product Usage: orthopaedic
Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page