Skip to content

GENESIS II 71420874Recalls & reported problems

Smith & Nephew, Inc. · GENESIS II POROUS POSTERIOR STABILIZED FEMORAL SIZE 7 LEFT · Class 2 · Implant

All models of GENESIS II (462)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 03596010209849), received between Mar 16, 2020 and Mar 16, 2020 — classified as 1 injury.

The most frequently reported specific device problem is “Break” (1 report).

6 other recalls cover devices cleared under the same FDA submission (K951987); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Break1

Patient problems

  • Injury1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 16, 2020Break; Break

    IT WAS REPORTED TWELVE YEARS AFTER IMPLANTATION REVISION SURGERY WAS PERFORMED DUE TO INSTABILITY. UPON INSPECTION, THE FEMORAL COMPONENT APPEARED PRISTINE LIKE NEW. BUT, POST ON THE PS POLY WAS BROKEN. ALL PIECES WERE RECOVERED. CUSTOMER IS REQUIRING AN ANALYSIS OF BOTH STRUCTURAL INTEGRITY AND VOLUMETRIC WEAR.

    Patient problems: Injury

    MAUDE report 9833779 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K951987. One submission often covers many models, so check the product and lot details in each record.

  • Z-1418-2022 · Jun 27, 2022 · Terminated
    (1) Genesis II Non-Porous Tibial Baseplate Size 5 Right, Product #71420186. (2) Genesis II Non-Porous Tibial Baseplate Size 4 Left, Product #71420166.
  • Z-2774-2020 · Jul 16, 2020 · Open, Classified
    Tibial knee prosthesis
  • Z-2883-2018 · Aug 1, 2017 · Terminated
    GENESIS II NON-POROUS CRUCIATE RETAINING FEMORAL COMPONENT, SIZE 6 RIGHT, REF 71420026 Knee prosthesis
  • Z-0573-2016 · Nov 9, 2015 · Terminated
    Smith & Nephew GENESIS(R) II/LEGION(TM) PATELLAR COMPONENT, RESURFACING ROUND, 32MM STANDARD, FOR CEMENTED USE ONLY WITH JOURNEY PEG, STERILE EO, Rx only, REF Number: 71934490 kn
  • Z-0130-2014 · Oct 7, 2013 · Terminated
    GENESIS(R) II, NON-POROUS TIBIAL BASE, SIZE 5, LEFT, QTY: (1), REF 71420168. Product Usage: orthopaedic
  • Z-0131-2014 · Oct 7, 2013 · Terminated
    GENESIS(R) II, NON-POROUS TIBIAL BASE, SIZE 6, RIGHT, QTY: (1), REF 71420188. Product Usage: orthopaedic

Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices

All PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page