Skip to content

GENESIS II 71440322Recalls & reported problems

Smith & Nephew, Inc. · GENESIS II PATELLAR CEMENT CLAMP · Class 2

Adverse event reports
7
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 03596010214980), received between Feb 21, 2021 and Dec 29, 2025 — classified as 1 injury, 6 malfunctions.

The most frequently reported specific device problem is “Break” (4 reports).

4 other recalls cover devices cleared under the same FDA submission (K121393); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break4
  • Fracture2
  • Detachment of Device or Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 29, 2025Fracture; Fracture

    IT WAS REPORTED THAT, DURING A TKA SURGERY, THE BLACK BUMPER OF ONE (1) GII PATELLAR CEMENT CLAMP BROKE OFF. NO PIECES FELL INSIDE PATIENT. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING AN EQUIVALENT S+N BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23907978 ↗
  • MalfunctionDec 31, 2023Fracture; Fracture

    IT WAS REPORTED THAT DURING A TKA SURGERY WHILE IMPACTING THE GII PATELLAR CEMENT CLAMP THE HEAD BROKE OFF. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18423035 ↗
  • MalfunctionMar 16, 2022Break; Break

    IT WAS REPORTED THAT DURING INSPECTION OF THE RETURNED TRAY, THE GII PATELLAR CEMENT CLAMP WAS BROKEN. IT IS UNKNOWN WHEN THE DEVICE BROKE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13778246 ↗
  • InjuryFeb 27, 2022Break; Break

    IT WAS REPORTED THAT, DURING TOTAL KNEE ARTHROPLASTY, WHILE REMOVING A GENESIS II PATELLAR CEMENT CLAMP FROM THE PATELLA ONCE THE CEMENT HAD CURED, THE SURGEON SPUN THE GOLD KNOB AND IT ALL CAME OFF, WHILE IT WAS INSIDE THE PATIENT. THE SMALL STOP ON THE END OF THE CLAMP TO STOP THIS FROM HAPPENING HAS COME OFF. SURGERY WAS RESUMED, WITHOUT ANY DELAY, WITH THE SAME DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM. THE PATIENT WAS X-RAYED AFTER CLOSURE TO ENSURE THE BROKEN PIECE WAS NOT LEFT INSIDE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13615116 ↗
  • MalfunctionOct 8, 2021Break; Break

    IT WAS REPORTED THAT, DURING A TKA SURGERY, A GII PATELLAR CEMENT CLAMP BROKE. IT IS UNKNOWN WHETHER THE EVENT HAPPENED DURING USE IN PATIENT AND IF THERE WAS PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12597990 ↗
  • MalfunctionSep 2, 2021Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT, AFTER A TKA PROCEDURE, THE BLACK PLASTIC TIP OF A GII PATELLAR CEMENT CLAMP CAME OUT AFTER TRYING TO REMOVE THE LARGER ROUND PLASTIC PIECE. AS THIS WAS NOTICED AFTER SURGERY, NO SURGICAL DELAY OCCURRED AND THE PROCEDURE WAS CONCLUDED WITH THE SAME DEVICE. NO PATIENT HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12418378 ↗
  • MalfunctionFeb 21, 2021Break; Break

    IT WAS REPORTED THAT DURING INSPECTION, CLAMP WAS FOUND TO BE BROKEN WHEN IT WAS OPENED. A BACKUP DEVICE FROM SMITH AND NEPHEW WAS USED TO COMPLETE THE PROCEDURE AND NO DELAY OCCURRED DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11356743 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K121393. One submission often covers many models, so check the product and lot details in each record.

  • Z-0878-2020 · Dec 16, 2019 · Open, Classified
    LEGION AP Cutting Block Size 4, REF 71434409; LEGION AP Cutting Block Size 6, REF 71434411; LEGION AP Cutting Block Size 7, REF 71434412; LEGION AP Cutting Block Size 8, REF 714344
  • Z-0200-2019 · Sep 18, 2018 · Terminated
    Smith+Nephew Orthopaedic Division LEGION A/P FEMORAL CUTTING BLOCK, SIZE 6 REF 71434411, QTY: (1) The LEGION AP Femoral Cutting Block is a reusable surgical instrument used to mak
  • Z-2163-2018 · May 4, 2018 · Terminated
    smith&nephew GENESIS(R) II TIBIAL Punch, 13 MM, REF 71440408, QTY: (1)
  • Z-1902-2018 · Apr 9, 2018 · Terminated
    smith&nephew 5o (5 degree) GENESIS(R) II VALGUS BUSHING, REF CATALOG NUMBER 71440014 The GENESIS II Knee System Components are indicated for rheumatoid arthritis; post-traumatic a

Other PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER devices

All PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.